Arnaud Savouré1, Alexis Mechulan2, Marc Burban3, Audrey Olivier4, Arnaud Lazarus5.
Abstract
BACKGROUND: The impact of magnetic resonance imaging (MRI) on pacemakers is potentially hazardous. We present clinical results from a novel MRI conditional pacing system with the capability to switch automatically to asynchronous mode in the presence of a strong magnetic field. AIMS: The IKONE (Assessment of the MRI solution: KORA 100™ and Beflex™ pacing leads system) study is an open-label, prospective, multicenter study aimed at confirming the safety and effectiveness of the system, when used in patients undergoing MRI of anatomical regions excluding the chest.Entities:
Keywords: 1.5-T MRI; bradycardia lead; pacemaker
Year: 2015 PMID: 26327785 PMCID: PMC4536771 DOI: 10.4137/CMC.S24976
Source DB: PubMed Journal: Clin Med Insights Cardiol ISSN: 1179-5468
Demographic characteristics.
| VARIABLES | ALL PATIENTS ( |
|---|---|
| Age, mean ± SD | 72.8 ± 11.4 |
| Male gender, | 23 (69.7%) |
| Atrio-ventricular block | 14 (42.4%) |
| Bundle branch block | 2 (6.1%) |
| Sinus node disease | 10 (30.3%) |
| Syncope | 7 (21.2%) |
| Atrial arrhythmias | 13 (39.4%) |
| Ventricular extrasystole | 1 (3.0%) |
| Coronary artery disease | 10 (30.3%) |
| Cardiomyopathy | 8 (24.2%) |
| Valvular heart disease | 12 (36.4%) |
| Arterial hypertension | 22 (66.7%) |
| Bypass | 7 (21.2%) |
| Valve replacement | 8 (24.2%) |
| Angioplasty | 6 (18.2%) |
| Diabetes mellitus | 10 (30.3%) |
Notes:
Includes fibrillation/flutter/tachycardia.
Abbreviation: SD, standard deviation.
Stability of mean PCT, sensing amplitude, and impedance after the MRI examination.
| MEAN ± SD (MIN–MAX) | PRE-MRI | POST-MRI | 1-MONTH POST-MRI | MEAN ABSOLUTE VARIATION AT 1 MONTH POST-MRI | MEAN ABSOLUTE VARIATION POST-MRI |
|---|---|---|---|---|---|
| RA | 0.76 ± 0.33 V | 0.79 ± 0.31 V | 0.81 ± 0.35 V | 0.18 ± 0.16 | 0.11 ± 0.19 V |
| RV | 0.76 ± 0.32 V | 0.88 ± 0.27 V | 0.85 ± 0.22 V | 0.18 ± 0.22 | 0.10 ± 0.20 V |
| RA | 4.48 ± 1.82 mV | 4.53 ± 1.64 mV | 4.55 ± 1.61 mV | 0.77 ± 1.02 mV | 0.62 ± 1.00 mV |
| RV | 11.75 ± 3.62 mV | 11.89 ± 3.60 mV | 11.78 ± 3.90 mV | 0.57 ± 0.72 mV | 0.30 ± 0.57 mV |
| RA | 477 ± 72 Ω | 536 ± 87 Ω | 482 ± 81 Ω | 29 ± 32 Ω | 64 ± 83 Ω |
| RV | 578 ± 151 Ω | 627 ± 173 Ω | 602 ± 183 Ω | 39 ± 44 Ω | 55 ± 71 Ω |
Note:
P < 0.001, comparison with the prespecified value 0.75 V.
Abbreviations: RA, right atrium; RV, right ventricle; SD, standard deviation; V, volt.
Characteristics of commercially available 1.5-T MRI conditional pacemakers.
| PACEMAKER MODELS | ADVISA MRI, ENSURA-MRI | ACCENT RF MRI ACCENT MRI | ELUNA, EPYRA, ETRINSA, EVIA, ESTELLA, ENTOVIS, ECURO | INGENIO, FORMIO VITALIO | KORA 100™ | KORA 250 | ||||
|---|---|---|---|---|---|---|---|---|---|---|
| Manufacturer | Medtronic | St Jude | Biotronik | Boston Scientific | Sorin | |||||
| Maximum SAR depending on leads | ≤2 W kg−1 | ≤2 W kg−1 | ≤4 W kg−1 | ≤2 W kg−1 | ≤4 W kg−1 | ≤2 W kg−1 | ≤4 W kg−1 | ≤2 W kg−1 | ||
| MRI conditional leads | CapSureFix | Trendril STS | Tendril | NC | Safio | Solia | Fineline II | Ingevity | Beflex | |
| Scanning zone | Full body | Out-of-chest | Full body | Out-of-chest | Full body | Full body | Out-of-chest | Full body | ||
| MRI mode | Manual | Activator™ hand-held device | Manual | Manual (timer) | AutoMRI | |||||
| Size/weight | 12.7 cc/22 g | 13.1 cc/24 g | Evia: 12 cc/25 g | Ingenio: 14 cc/32 g | 8 cc/20 g | |||||