The effect of slice thickness on target and organs at risk volumes, dosimetric coverage and radiobiological impact in IMRT planning.

S P Srivastava1,2,3, C-W Cheng4,5,6, I J Das4,5.   

Abstract

PURPOSE: Dose-volume histogram (DVH) has become an important tool for evaluation of radiation outcome as reflected from many clinical protocols. While dosimetric accuracy in treatment planning system (TPS) is well quantified, the variability in volume estimation is uncertain due to reconstruction algorithm that is investigated in this study. In addition, the impact of dose distribution and tumor control probability (TCP) were also investigated with CT slice thickness for IMRT planning.
MATERIALS AND METHODS: A water phantom containing various objects with accurately known volume ranging from 1 to 100 cm(3) was scanned with 1, 2, 3, 5, and 10 mm slice thickness. The CT data sets were sent to Eclipse TPS for contour delineation and volume estimation. The data were compared with known volume for the estimation of error in the volume of each structure. IMRT Plans were generated on phantom containing four objects with different slice thickness (1-5 mm) to calculate TCP. ICRU-83-recommended dose points such as D 2%, D 50%, D 98%, as well as homogeneity and conformity index were also calculated.
RESULTS: The variability of volumes with CT slice thickness was significant especially for small volume structures. A maximum error of 92% was noticed for 1 cm(3) volume of object with 10 mm slice thickness, whereas it was ~19% for 1 mm slice thickness. For 2 and 3 cm(3) objects, the maximum error of 99% was noticed with 10 mm slice thickness and ~60% with 5 mm. The differences are smaller for larger volumes with a cutoff at about 20 cm(3). The calculated volume of the objects is a function of reconstruction algorithm and slice thickness. The PTV mean dose and TCP decreased with increasing slice thickness. Maximum variation of ~5% was noticed in mean dose and ~2% in TCP with change in slice thickness from 1 to 5 mm. The relative decrease in target volume receiving 95% of the prescribed dose is ~5% with change in slice thickness from 1 to 5 mm. The homogeneity index increases up to 163% and conformity index decreases by 4% between 1 and 5 mm slice thickness, producing highly inhomogeneous and least conformal treatment plan.
CONCLUSIONS: Estimation of a volume is dependent on CT slice thickness and the contouring algorithm in a TPS. During commissioning of TPS and for all clinical protocols, evaluation of volume should be included to provide the limit of accuracy in DVH from TPS, especially for small objects. A smaller slice thickness provides superior dosimetry with improved TCP. Thus, the smallest possible slice thickness should be used for IMRT planning, especially when smaller structures are present.

Entities:  

Keywords:  CT slice thickness; IMRT dosimetry; NTCP; TCP; Volume estimation

Mesh:

Year:  2015        PMID: 26311077     DOI: 10.1007/s12094-015-1390-z

Source DB:  PubMed          Journal:  Clin Transl Oncol        ISSN: 1699-048X            Impact factor:   3.405


  25 in total

1.  Dosimetric verification of a commercial collapsed cone algorithm in simulated clinical situations.

Authors:  Andrew Nisbet; Ian Beange; Hans-Stephan Vollmar; Catherine Irvine; Andrew Morgan; David I Thwaites
Journal:  Radiother Oncol       Date:  2004-10       Impact factor: 6.280

2.  The dosimetric verification of a pencil beam based treatment planning system.

Authors:  T Knöös; C Ceberg; L Weber; P Nilsson
Journal:  Phys Med Biol       Date:  1994-10       Impact factor: 3.609

3.  Beam modeling and verification of a photon beam multisource model.

Authors:  Anders Ahnesjö; Lars Weber; Anders Murman; Mikael Saxner; Ingvar Thorslund; Erik Traneus
Journal:  Med Phys       Date:  2005-06       Impact factor: 4.071

4.  Impact of different CT slice thickness on clinical target volume for 3D conformal radiation therapy.

Authors:  Ramachandran Prabhakar; Tharmar Ganesh; Goura K Rath; Pramod K Julka; Pappiah S Sridhar; Rakesh C Joshi; Sanjay Thulkar
Journal:  Med Dosim       Date:  2008-04-01       Impact factor: 1.482

Review 5.  American Association of Physicists in Medicine Radiation Therapy Committee Task Group 53: quality assurance for clinical radiotherapy treatment planning.

Authors:  B Fraass; K Doppke; M Hunt; G Kutcher; G Starkschall; R Stern; J Van Dyke
Journal:  Med Phys       Date:  1998-10       Impact factor: 4.071

6.  Commissioning and quality assurance of treatment planning computers.

Authors:  J Van Dyk; R B Barnett; J E Cygler; P C Shragge
Journal:  Int J Radiat Oncol Biol Phys       Date:  1993-05-20       Impact factor: 7.038

7.  Clinical target volume definition for glioblastoma radiotherapy planning: magnetic resonance imaging and computed tomography.

Authors:  A Fiorentino; R Caivano; P Pedicini; V Fusco
Journal:  Clin Transl Oncol       Date:  2013-01-29       Impact factor: 3.405

Review 8.  Intensity-modulated radiation therapy for head and neck cancer: emphasis on the selection and delineation of the targets.

Authors:  Avraham Eisbruch; Robert L Foote; Brian O'Sullivan; Jonathan J Beitler; Bhadrasain Vikram
Journal:  Semin Radiat Oncol       Date:  2002-07       Impact factor: 5.934

9.  Volume assessment accuracy in computed tomography: a phantom study.

Authors:  Nicolas D Prionas; Shonket Ray; John M Boone
Journal:  J Appl Clin Med Phys       Date:  2010-04-16       Impact factor: 2.102

Review 10.  Quality assurance for clinical trials.

Authors:  Geoffrey S Ibbott; Annette Haworth; David S Followill
Journal:  Front Oncol       Date:  2013-12-19       Impact factor: 6.244

View more
  3 in total

1.  Is it possible to estimate volume of bone defects formed on dry sheep mandibles more practically by secondarily reconstructing section thickness of cone beam computed tomography images?

Authors:  Alaettin Koç; Sema Kaya
Journal:  Dentomaxillofac Radiol       Date:  2020-10-15       Impact factor: 2.419

2.  Prevalence and Clinical Impact of Incidental Findings on Preoperative 3D Planning Computed Tomography for Total Shoulder Arthroplasty.

Authors:  Yuqing Chen; Sarav S Shah; Alexander M Roche; Lambert T Li; Matthew Chilton; Benjamin Saks; Meghan Macaskill; Glen Ross
Journal:  J Am Acad Orthop Surg Glob Res Rev       Date:  2022-08-05

3.  Impact of CT slice thickness on volume and dose evaluation during thoracic cancer radiotherapy.

Authors:  Huanli Luo; Yanan He; Fu Jin; Dingyi Yang; Xianfeng Liu; Xueqi Ran; Ying Wang
Journal:  Cancer Manag Res       Date:  2018-09-20       Impact factor: 3.989

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.