| Literature DB >> 26300614 |
Heather P Whitley1, Ee Vonn Yong2, Casey Rasinen2.
Abstract
A1C point-of-care (POC) instruments benefit patients with diabetes by facilitating clinician decision making that results in significant glycemic improvements. Three National Glycohemoglobin Standardization Program (NGSP)-certified POC products are available in the United States: the handheld A1CNow (formerly manufactured by Bayer Diabetes Care but now made by Chek Diagnostics) and two bench-top models called the Axis-Shield Afinion Analyzer and the Siemens DCA Vantage. This article compares the three available NGSP-certified POC products in terms of accuracy, precision, ease of use, cost, and additional features. Its goal is to aid health care facilities in conveniently identifying the A1C POC product that best meets their needs. It additionally reviews evidence that supports the continued use of A1C POC instruments in the clinical arena.Entities:
Year: 2015 PMID: 26300614 PMCID: PMC4536639 DOI: 10.2337/diaspect.28.3.201
Source DB: PubMed Journal: Diabetes Spectr ISSN: 1040-9165
Performance Data for NGSP-Certified A1C POC Devices
| Assay Method | Labs ( | NGSP A1C Reference Value (95% CI) | ||||||||
| 5.30 (5.26–5.34) | 6.14 (6.10–6.18) | 8.05 (8.01–8.09) | ||||||||
| Mean A1C (%) | Mean bias | CV (%) | Mean A1C (%) | Mean bias | CV (%) | Mean A1C (%) | Mean bias | CV (%) | ||
| Bayer A1CNow | 13 | 4.95 | –0.35 | 4.6 | 5.70 | –0.44 | 5.5 | 7.43 | –0.62 | 5.7 |
| Axis-Shield Afinion | 34 | 5.25 | –0.05 | 2.8 | 6.04 | –0.10 | 3.1 | 7.64 | –0.41 | 2.4 |
| Siemens DCA Vantage | 258 | 5.21 | –0.09 | 2.3 | 6.03 | –0.11 | 2.5 | 7.84 | –0.21 | 2.7 |
Adapted from ref. 18.
EDTA in the CAP sample has been shown by the manufacturer of A1CNow to cause artificially low results by this method. Routine samples for this method are from fingerstick and do not include EDTA. The manufacturer recommends the use of heparin anticoagulant instead of EDTA when testing venous samples.
Comparison of NGSP-Certified A1C POC Devices
| Characteristic | Bayer A1CNow ( | Axis-Shield Afinion ( | Siemens DCA Vantage ( |
| Physical size | ● Disposable, portable, handheld | ● Bench-top unit | ● Bench-top unit |
| ● Dimensions: | ● Dimensions: | ● Dimensions: | |
| ❍ Depth: 6.35 cm (2.5 in) | ❍ Depth: 34 cm (13.4 in) | ❍ Depth: 27.7 cm (10.5 in) | |
| ❍ Height: 1.0 cm (0.4 in) | ❍ Height: 17 cm (6.7 in) | ❍ Height: 25.4 cm (9.0 in) | |
| ❍ Width: 5.1 cm (2.0 in) | ❍ Width: 19 cm (7.4 in) | ❍ Width: 28.7 cm (11.5 in) | |
| ❍ 0.18 kg (0.4 lb) | ❍ Weight: 5 kg (11 lb) | ❍ Weight: 3.88 kg (9 lb) | |
| Assay methodology | Immunoassay | Boronate affinity separation | Immunoassay |
| Blood sample size (μL) | 5 | 1.5 | 1 |
| Analysis time (min) | 5 | 3 | 6 |
| Reporting A1C range (%) | 4.0–13.0 | 4.0–15.0 | 2.5–14.0 |
| Sources of interference ( | HbF (when level is 10–15%) | HbF (when level is 10–15%) | HbF (when level is >20%) |
| HbC | |||
| HbS | |||
| Other quantitative tests (moderate complexity requiring quarterly proficiency testing) | None | ● Albumin:creatinine ratio | ● Albumin:creatinine ratio |
| ● Creatinine | ● Creatinine | ||
| ● C-reactive protein and cholesterol (not available in the United States) | ● Microalbumin | ||
| Storage | ● Store the test kit refrigerated (2–8°C) until the expiration date or at room temperature (15–25°C) for up to 4 months. | ● Store the test kit refrigerated (2–8°C) until the expiration date or at room temperature (15–25°C) for up to 3 months. | ● Store the test kit refrigerated (2–8°C) until the expiration date or at room temperature (15–25°C) for up to 3 months. |
| ● Use the test cartridge within 2 minutes after opening the foil pouch. | ● Use the test cartridge within 10 minutes after opening the foil pouch. | ● Use the test cartridge within 5 minutes after opening the foil pouch. | |
| Features | ● Memory capacity: none | ● Memory capacity: 500 patient results and 500 control results | ● Memory capacity: 4,000 patient and control records |
| ● Display: black and white, nontouch | ● Display: color touch | ● Color touch display | |
| ● Power supply: battery | ● Power supply: AC/DC adapter | ● Power supply: AC/DC adapter | |
| ● Calibration: none | ● Calibration: none | ● Calibration: lot specific calibration card | |
| ● Separate test cartridge and sampling device | ● Integrated test cartridge and sampling device | ● Separate test cartridge and sampling device | |
| ● Data export: none | ● Data export: | ● Data export | |
| ● Accessories: one | ❍ USB flash drive | ❍ USB flash drive | |
| ❍ Ethernet port | ❍ Ethernet port | ||
| ❍ RS232 | ❍ Accessories: barcode scanner, onboard printer | ||
| ● Accessories: barcode scanner, purchased separately | |||
| Approximate cost per unit (prices vary by distributor) | $40 for unit with 2 cartridges (for the home-use version) | $3,500 for base station | $2,100–3,600 for base station |
| $170 for unit with 20 cartridges | $120 for 15 single-use testing kits | $75 for 10 single-use testing kits | |
| $60 for 2 control sets | $60 for 2 control sets | ||
| Company Internet address |