| Literature DB >> 26278979 |
Timothy Grant1, Nikolaus Grigorieff2.
Abstract
We demonstrate a significant anisotropic magnification distortion, found on an FEI Titan Krios microscope and affecting magnifications commonly used for data acquisition on a Gatan K2 Summit detector. We describe a program (mag_distortion_estimate) to automatically estimate anisotropic magnification distortion from a set of images of a standard gold shadowed diffraction grating. We also describe a program (mag_distortion_correct) to correct for the estimated distortion in collected images. We demonstrate that the distortion present on the Titan Krios microscope limits the resolution of a set of rotavirus VP6 images to ∼7 Å, which increases to ∼3 Å following estimation and correction of the distortion. We also use a 70S ribosome sample to demonstrate that in addition to affecting resolution, magnification distortion can also interfere with the classification of heterogeneous data.Entities:
Keywords: Electron cryo-microscopy; Image correction; Magnification anisotropy; Resolution
Mesh:
Substances:
Year: 2015 PMID: 26278979 PMCID: PMC6760661 DOI: 10.1016/j.jsb.2015.08.006
Source DB: PubMed Journal: J Struct Biol ISSN: 1047-8477 Impact factor: 2.867