| Literature DB >> 26266051 |
Charlotte M Cronenweth1, Frank G Shellock2.
Abstract
Purpose. To assess the MRI issues at 3 Tesla for a metallic tissue marker used to localize removal areas of tissue abnormalities. Materials and Methods. A newly designed, metallic tissue marker (Achieve Marker, CareFusion, Vernon Hills, IL) used to mark biopsy sites, particularly in breasts, was assessed for MRI issues which included standardized tests to determine magnetic field interactions (i.e., translational attraction and torque), MRI-related heating, and artifacts at 3 Tesla. Temperature changes were determined for the marker using a gelled-saline-filled phantom. MRI was performed at a relatively high specific absorption rate (whole body averaged SAR, 2.9-W/kg). MRI artifacts were evaluated using T1-weighted, spin echo and gradient echo pulse sequences. Results. The marker displayed minimal magnetic field interactions (2-degree deflection angle and no torque). MRI-related heating was only 0.1°C above background heating (i.e., the heating without the tissue marker present). Artifacts seen as localized signal loss were relatively small in relation to the size and shape of the marker. Conclusions. Based on the findings, the new metallic tissue marker is acceptable or "MR Conditional" (using current labeling terminology) for a patient undergoing an MRI procedure at 3 Tesla or less.Entities:
Year: 2015 PMID: 26266051 PMCID: PMC4525757 DOI: 10.1155/2015/823759
Source DB: PubMed Journal: Int J Breast Cancer ISSN: 2090-3189
Figure 1The metallic tissue marker (Achieve Marker, 3-mm length, CareFusion, Vernon Hills, IL) that underwent testing for MRI issues at 3 Tesla. Note the irregular surface, allowing for growth of tissue in and around this implant to reduce migration once it is deployed.
Summary of MRI artifacts at 3 Tesla for the tissue marker.
| Pulse sequence | T1-SE | T1-SE | GRE | GRE |
|
| ||||
| Signal void size (mm2) | 38 | 16 | 82 | 61 |
|
| ||||
| Imaging plane | Parallel | Perpendicular | Parallel | Perpendicular |
(T1-SE: T1-weighted, spin echo; GRE: gradient).
Figure 2(a) MRI artifact associated with the tissue marker on a gradient echo pulse sequence (TR/TE, 100-msec/15-msec; flip angle, 30 degrees; long axis view; artifact area, 82-mm2). The artifact appeared as a low signal intensity signal void that was relatively small in relation to the size and shape of the marker. (b) MRI artifact associated with the tissue marker on a spin echo pulse sequence (TR/TE, 500-msec/20-msec; long axis view; artifact area, 38-mm2). The artifact appeared as a low signal intensity signal void that was relatively small in relation to the size and shape of the marker. Note that the artifact size was larger for the gradient echo pulse sequence (a) compared to that observed for the spin echo pulse sequence.