Magnetic resonance imaging of the metal clip in a breast: safety and its availability as a negative marker.

Kazunori Kubota1, Naoya Gomi, Toshihiko Wakita, Hitoshi Shibuya, Masaki Kakimoto, Takayuki Osanai.   

Abstract

PURPOSE: This study assesses magnetic resonance (MR) safety of the stainless-steel clip inserted after stereotactic-guided directional vacuum-assisted biopsy (DVAB) of the breast, and evaluates its imaging value.
METHODS: We used a sausage as a substitute breast and inserted the clip into it. The MR images of the substitute were scanned using a breast coil, and it was then dissected. After the substitute experimentation, MR scanning of the breast was performed using a dynamic contrast enhanced technique, in which a clip was placed after DVAB for suspicion of ductal carcinoma was seen as grouped amorphous calcifications on mammography.
RESULTS: On every magnetic resonance image of the substitute, the clip was seen as a spotty signal void, with no surrounding artifact. There was no movement and no evidence of increased clip temperature on dissected of the substitute, confirming the safety of breast MR with a clip in place. There was no patient complaint of feeling heat or pain during the MRI examination and there were only biopsy scars on the surgically excised breast specimen material. On the breast MR images, a spreading region of the tumor adjoining the position of the signal void was identified as an early enhancing lesion.
CONCLUSIONS: The safety and reliability of breast MR examination using a mammotome clip was demonstrated by both the mock examination and the breast examination. It is possible to localize of tumor spread regions based on the marker position using the clip as a negative signal marker.

Entities:  

Mesh:

Year:  2004        PMID: 14718794     DOI: 10.1007/bf02968003

Source DB:  PubMed          Journal:  Breast Cancer        ISSN: 1340-6868            Impact factor:   4.239


  3 in total

1.  Application of ultrasound-guided placement of markers for locating axillary lymph nodes of breast cancer.

Authors:  Xiaohui Ji; Mengying Wei; Liuyuan Wang; Juanjuan Li; Dongxia Gao; Cuizhi Geng
Journal:  Gland Surg       Date:  2021-11

2.  Assessment of MRI Issues at 3 Tesla for a New Metallic Tissue Marker.

Authors:  Charlotte M Cronenweth; Frank G Shellock
Journal:  Int J Breast Cancer       Date:  2015-07-22

3.  Tracking nonpalpable breast cancer for breast-conserving surgery with carbon nanoparticles: implication in tumor location and lymph node dissection.

Authors:  Yanyan Jiang; Nan Lin; Sheng Huang; Chongping Lin; Na Jin; Zaizhong Zhang; Jun Ke; Yinghao Yu; Jianping Zhu; Yu Wang
Journal:  Medicine (Baltimore)       Date:  2015-03       Impact factor: 1.889

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.