Three dimensional dose distribution comparison of simple and complex acquisition trajectories in dedicated breast CT.

Jainil P Shah1, Steve D Mann2, Randolph L McKinley3, Martin P Tornai4.   

Abstract

PURPOSE: A novel breast CT system capable of arbitrary 3D trajectories has been developed to address cone beam sampling insufficiency as well as to image further into the patient's chest wall. The purpose of this study was to characterize any trajectory-related differences in 3D x-ray dose distribution in a pendant target when imaged with different orbits.
METHODS: Two acquisition trajectories were evaluated: circular azimuthal (no-tilt) and sinusoidal (saddle) orbit with ±15° tilts around a pendant breast, using Monte Carlo simulations as well as physical measurements. Simulations were performed with tungsten (W) filtration of a W-anode source; the simulated source flux was normalized to the measured exposure of a W-anode source. A water-filled cylindrical phantom was divided into 1 cm(3) voxels, and the cumulative energy deposited was tracked in each voxel. Energy deposited per voxel was converted to dose, yielding the 3D distributed dose volumes. Additionally, three cylindrical phantoms of different diameters (10, 12.5, and 15 cm) and an anthropomorphic breast phantom, initially filled with water (mimicking pure fibroglandular tissue) and then with a 75% methanol-25% water mixture (mimicking 50-50 fibroglandular-adipose tissues), were used to simulate the pendant breast geometry and scanned on the physical system. Ionization chamber calibrated radiochromic film was used to determine the dose delivered in a 2D plane through the center of the volume for a fully 3D CT scan using the different orbits.
RESULTS: Measured experimental results for the same exposure indicated that the mean dose measured throughout the central slice for different diameters ranged from 3.93 to 5.28 mGy, with the lowest average dose measured on the largest cylinder with water mimicking a homogeneously fibroglandular breast. These results align well with the cylinder phantom Monte Carlo studies which also showed a marginal difference in dose delivered by a saddle trajectory in the central slice. Regardless of phantom material or filled fluid density, dose delivered by the saddle scan was negligibly different than the simple circular, no-tilt scans. The average dose measured in the breast phantom was marginally higher for saddle than the circular no tilt scan at 3.82 and 3.87 mGy, respectively.
CONCLUSIONS: Not only does nontraditional 3D-trajectory CT scanning yield more complete sampling of the breast volume but also has comparable dose deposition throughout the breast and anterior chest volume, as verified by Monte Carlo simulation and physical measurements.

Entities:  

Mesh:

Substances:

Year:  2015        PMID: 26233179      PMCID: PMC4499047          DOI: 10.1118/1.4923169

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  25 in total

1.  Glandular breast dose for monoenergetic and high-energy X-ray beams: Monte Carlo assessment.

Authors:  J M Boone
Journal:  Radiology       Date:  1999-10       Impact factor: 11.105

2.  A comprehensive analysis of DgN(CT) coefficients for pendant-geometry cone-beam breast computed tomography.

Authors:  J M Boone; N Shah; T R Nelson
Journal:  Med Phys       Date:  2004-02       Impact factor: 4.071

3.  Dedicated breast CT: radiation dose for circle-plus-line trajectory.

Authors:  Srinivasan Vedantham; Linxi Shi; Andrew Karellas; Frederic Noo
Journal:  Med Phys       Date:  2012-03       Impact factor: 4.071

4.  Cone-beam CT for breast imaging: Radiation dose, breast coverage, and image quality.

Authors:  Avice O'Connell; David L Conover; Yan Zhang; Posy Seifert; Wende Logan-Young; Chuen-Fu Linda Lin; Lawrence Sahler; Ruola Ning
Journal:  AJR Am J Roentgenol       Date:  2010-08       Impact factor: 3.959

5.  Breast dose from magnification films in mammography.

Authors:  J Law
Journal:  Br J Radiol       Date:  2005-09       Impact factor: 3.039

6.  A high-resolution voxel phantom of the breast for dose calculations in mammography.

Authors:  Christoph Hoeschen; Ute Fill; Maria Zankl; Werner Panzer; Dieter Regulla; Wilfried Döhring
Journal:  Radiat Prot Dosimetry       Date:  2005       Impact factor: 0.972

7.  Experimental spectral measurements of heavy K-edge filtered beams for x-ray computed mammotomography.

Authors:  D J Crotty; R L McKinley; M P Tornai
Journal:  Phys Med Biol       Date:  2007-01-10       Impact factor: 3.609

8.  Evaluation of the absorbed dose to the breast using radiochromic film in a dedicated CT mammotomography system employing a quasi-monochromatic x-ray beam.

Authors:  Dominic J Crotty; Samuel L Brady; D'Vone C Jackson; Greta I Toncheva; Colin E Anderson; Terry T Yoshizumi; Martin P Tornai
Journal:  Med Phys       Date:  2011-06       Impact factor: 4.071

9.  Computed tomography for imaging the breast.

Authors:  John M Boone; Alex L C Kwan; Kai Yang; George W Burkett; Karen K Lindfors; Thomas R Nelson
Journal:  J Mammary Gland Biol Neoplasia       Date:  2006-04       Impact factor: 2.673

10.  Dedicated breast CT: fibroglandular volume measurements in a diagnostic population.

Authors:  Srinivasan Vedantham; Linxi Shi; Andrew Karellas; Avice M O'Connell
Journal:  Med Phys       Date:  2012-12       Impact factor: 4.071

View more
  9 in total

1.  Task-driven source-detector trajectories in cone-beam computed tomography: I. Theory and methods.

Authors:  J Webster Stayman; Sarah Capostagno; Grace J Gang; Jeffrey H Siewerdsen
Journal:  J Med Imaging (Bellingham)       Date:  2019-05-02

2.  Implementation and CT sampling characterization of a third-generation SPECT-CT system for dedicated breast imaging.

Authors:  Jainil P Shah; Steve D Mann; Randolph L McKinley; Martin P Tornai
Journal:  J Med Imaging (Bellingham)       Date:  2017-07-31

3.  Characterization of CT Hounsfield Units for 3D acquisition trajectories on a dedicated breast CT system.

Authors:  Jainil P Shah; Steve D Mann; Randolph L McKinley; Martin P Tornai
Journal:  J Xray Sci Technol       Date:  2018       Impact factor: 1.535

4.  Cone-beam breast CT using an offset detector: effect of detector offset and image reconstruction algorithm.

Authors:  Hsin Wu Tseng; Andrew Karellas; Srinivasan Vedantham
Journal:  Phys Med Biol       Date:  2022-04-07       Impact factor: 4.174

5.  A dedicated breast-PET/CT scanner: Evaluation of basic performance characteristics.

Authors:  Raymond R Raylman; Will Van Kampen; Alexander V Stolin; Wenbo Gong; Gangadhar Jaliparthi; Peter F Martone; Mark F Smith; David Sarment; Neal H Clinthorne; Mark Perna
Journal:  Med Phys       Date:  2018-02-23       Impact factor: 4.071

6.  Cone beam CT multisource configurations: evaluating image quality, scatter, and dose using phantom imaging and Monte Carlo simulations.

Authors:  Amy E Becker; Andrew M Hernandez; Paul R Schwoebel; John M Boone
Journal:  Phys Med Biol       Date:  2020-12-18       Impact factor: 3.609

7.  Radiation dosimetry of a clinical prototype dedicated cone-beam breast CT system with offset detector.

Authors:  Hsin Wu Tseng; Andrew Karellas; Srinivasan Vedantham
Journal:  Med Phys       Date:  2021-01-26       Impact factor: 4.506

8.  Shading artifact correction in breast CT using an interleaved deep learning segmentation and maximum-likelihood polynomial fitting approach.

Authors:  Peymon Ghazi; Andrew M Hernandez; Craig Abbey; Kai Yang; John M Boone
Journal:  Med Phys       Date:  2019-06-23       Impact factor: 4.071

9.  Motion artifacts assessment and correction using optical tracking in synchrotron radiation breast CT.

Authors:  Luca Brombal; Lucia Mariel Arana Peña; Fulvia Arfelli; Renata Longo; Francesco Brun; Adriano Contillo; Francesca Di Lillo; Giuliana Tromba; Vittorio Di Trapani; Sandro Donato; Ralf Hendrik Menk; Luigi Rigon
Journal:  Med Phys       Date:  2021-07-29       Impact factor: 4.506

  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.