Abstract

BACKGROUND: Poor vessel wall apposition of flow diverter (FD) stents poses risks for stroke-related complications when treating intracranial aneurysms, necessitating long-term surveillance imaging. To facilitate quantitative evaluation of deployed devices, a novel algorithm is presented that generates intuitive two-dimensional representations of wall apposition from either high-resolution contrast-enhanced cone-beam CT (VasoCT) or intravascular optical coherence tomography (OCT) images.
METHODS: VasoCT and OCT images were obtained after FD implant (n=8 aneurysms) in an experimental sidewall aneurysm model in canines. Surface models of the vessel wall and FD device were extracted, and the distance between them was presented on a two-dimensional flattened map. Maps and cross-sections at potential locations of malapposition detected on VasoCT-based maps were compared. The performance of OCT-based apposition detection was evaluated on manually labeled cross-sections using logistic regression against a thresholded (≥0.25 mm) apposition measure.
RESULTS: VasoCT and OCT acquisitions yielded similar Grading of Regional Apposition after Flow-Diverter Treatment (GRAFT) apposition maps. GRAFT maps from VasoCT highlighted 16 potential locations of malapposition, of which two were found to represent malapposed device struts. Logistic regression showed that OCT could detect malapposition with a sensitivity of 98% and a specificity of 81%.
CONCLUSIONS: GRAFT delivered quantitative and visually convenient representations of potential FD malapposition and occasional acute thrombus formation. A powerful combination for future neuroendovascular applications is foreseen with the superior resolution delivered by intravascular OCT. Published by the BMJ Publishing Group Limited. For permission to use (where not already granted under a licence) please go to http://www.bmj.com/company/products-services/rights-and-licensing/

Entities:  

Keywords:  Aneurysm; Angiography; Endoscopy; Flow Diverter; Technology

Mesh:

Year:  2015        PMID: 26220411     DOI: 10.1136/neurintsurg-2015-011843

Source DB:  PubMed          Journal:  J Neurointerv Surg        ISSN: 1759-8478            Impact factor:   5.836


  14 in total

1.  Wall Apposition Is a Key Factor for Aneurysm Occlusion after Flow Diversion: A Histologic Evaluation in 41 Rabbits.

Authors:  A Rouchaud; C Ramana; W Brinjikji; Y-H Ding; D Dai; T Gunderson; J Cebral; D F Kallmes; R Kadirvel
Journal:  AJNR Am J Neuroradiol       Date:  2016-07-07       Impact factor: 3.825

2.  Risk Factors for Ischemic Complications following Pipeline Embolization Device Treatment of Intracranial Aneurysms: Results from the IntrePED Study.

Authors:  W Brinjikji; G Lanzino; H J Cloft; A H Siddiqui; E Boccardi; S Cekirge; D Fiorella; R Hanel; P Jabbour; E Levy; D Lopes; P Lylyk; I Szikora; D F Kallmes
Journal:  AJNR Am J Neuroradiol       Date:  2016-04-21       Impact factor: 3.825

3.  Filament inadequate wall apposition of the different ends of flow diverters in the abdominal aorta of rabbits.

Authors:  Kuizhong Wang; Peng Jin; Peigang Lu; Qiang Liu; Bo Li; Zhenqiang Hao
Journal:  Neuroradiol J       Date:  2019-11-27

4.  Intravascular Optical Coherence Tomography for Neurointerventional Surgery.

Authors:  Matthew J Gounis; Giovanni J Ughi; Miklos Marosfoi; Demetrius K Lopes; David Fiorella; Hiram G Bezerra; Conrad W Liang; Ajit S Puri
Journal:  Stroke       Date:  2018-11-29       Impact factor: 7.914

5.  Aneurysm Remnants after Flow Diversion: Clinical and Angiographic Outcomes.

Authors:  T P Madaelil; J A Grossberg; B M Howard; C M Cawley; J Dion; R G Nogueira; D C Haussen; F C Tong
Journal:  AJNR Am J Neuroradiol       Date:  2019-03-07       Impact factor: 3.825

6.  Open-cell stent and use of cone-beam CT enables a safe and effective coil embolization of true ophthalmic artery and anterior choroidal artery aneurysms with preservation of parent vessel: Clinical and angiographic results.

Authors:  Samuel Y Hou; Anna Luisa Kühn; Ajit S Puri; Ajay K Wakhloo
Journal:  Interv Neuroradiol       Date:  2017-12-14       Impact factor: 1.610

7.  Treatment of an internal carotid artery aneurysm with a flow diverter through a double lumen balloon catheter.

Authors:  José M Amorim; Santiago Rosati; Ronit Agid; Vítor Mendes Pereira; Timo Krings
Journal:  Interv Neuroradiol       Date:  2017-01-01       Impact factor: 1.610

8.  Collar Sign in Incompletely Occluded Aneurysms after Pipeline Embolization: Evaluation with Angiography and Optical Coherence Tomography.

Authors:  C J Griessenauer; R Gupta; S Shi; A Alturki; R Motiei-Langroudi; N Adeeb; C S Ogilvy; A J Thomas
Journal:  AJNR Am J Neuroradiol       Date:  2016-11-10       Impact factor: 3.825

Review 9.  Optical Coherence Tomography in Cerebrovascular Disease: Open up New Horizons.

Authors:  Ran Xu; Qing Zhao; Tao Wang; Yutong Yang; Jichang Luo; Xiao Zhang; Yao Feng; Yan Ma; Adam A Dmytriw; Ge Yang; Shengpan Chen; Bin Yang; Liqun Jiao
Journal:  Transl Stroke Res       Date:  2022-04-21       Impact factor: 6.829

10.  Acute thrombus formation on phosphorilcholine surface modified flow diverters.

Authors:  Miklos Marosfoi; Frederic Clarencon; Erin T Langan; Robert M King; Olivia W Brooks; Takamisu Tamura; John M Wainwright; Matthew J Gounis; Srinivasan Vedantham; Ajit S Puri
Journal:  J Neurointerv Surg       Date:  2017-07-08       Impact factor: 5.836

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.