| Literature DB >> 26210316 |
Kuo-Wei Wang1,2, Hsiao-Huang Chang3,4, Chuan-Chih Hsu5,6, Kuang-Chao Chen7, Jen-Chuen Hsieh8, Lieber Po-Hung Li9, Po-Lei Lee10,11, An-Suey Shiao12.
Abstract
BACKGROUND: Auditory steady-state response (ASSR) induced by repetitive auditory stimulus is commonly used for audiometric testing. ASSR can be measured using electro-encephalography (EEG) and magnetoencephalography (MEG), referred to as steady-state auditory evoked potential (SSAEP) and steady-state auditory evoked field (SSAEF), respectively. However, the signal level of SSAEP and SSAEF are weak so that signal processing technique is required to increase its signal-to-noise ratio. In this study, a complementary ensemble empirical mode decomposition (CEEMD)-based approach is proposed in MEG study and the extraction of SSAEF has been demonstrated in normal subjects and tinnitus patients.Entities:
Mesh:
Year: 2015 PMID: 26210316 PMCID: PMC4514968 DOI: 10.1186/s12938-015-0062-0
Source DB: PubMed Journal: Biomed Eng Online ISSN: 1475-925X Impact factor: 2.819
Demographic data of tinnitus patient
| Patient # | Gender | Age (years) | Duration (years) | Perceived tinnitus location |
|---|---|---|---|---|
| P1 | M | 50 | 5 | Right ear |
| P2 | M | 48 | 6 | Left ear |
| P3 | M | 64 | 5 | Left ear |
| P4 | F | 69 | 3 | Both ears |
| P5 | M | 59 | 3 | Right ear |
Figure 1Spatial template creation using N100m. a The channel plot of induced AEF in subject I. b The magnetic fields of N100m in all MEG channels. c The normalized map of absolute value of b. c is separated into right (unshaded) and left (shaded) hemispheric parts for selecting SSAEF-related IMFs.
Figure 2The schematic diagram for the signal processing of the proposed CEEMD-based approach.
Figure 3Demonstration of signal extraction in a simulated 37 Hz SSAEF using EEMD, CEEMD and band-pass filtering. a The simulation of a modulated 37 Hz SSAEF with −1 dB SNR. b, c The SSAEFs reconstructed by the summation of IMF2 and IMF3 using EEMD and CEEMD with ensemble number equal to 1,000 times, respectively. d The SSAEF extracted by 35–39 Hz band-pass filtering.
Figure 4The RMSE values of extracted SSAEF in Figure 3 with different ensemble numbers.
Figure 5The spatial maps, IMF waveforms and Fourier spectra of the first four IMFs decomposed from right- and left-hemisphere COIs. The IMF2s decomposed from right and left COIs (correlation coefficient = 0.63 and 0.64 in right and left COI decompositions, respectively) were chosen for signal reconstruction.
Figure 6The comparison of SSAEFs extracted by our CEEMD-based approach and conventional filtering method in subject I. a One example of one-second SSAEF waveforms obtained from the two methods. b The Fourier spectra of a. c The spatial map and current sources of CEEMD processed SSAEF estimated at 218 ms in a. d The spatial map and current sources of SSAEF processed by 35–39 Hz bandpass filtering at 218 ms in a.
Comparison of estimated current strengths in normal subjects
| Normal subject | The proposed CEEMD approach | Traditional filtering method | ||||||
|---|---|---|---|---|---|---|---|---|
| Current strength of left ear stimulation | Current strength of right ear stimulation | Current strength of left ear stimulation | Current strength of right ear stimulation | |||||
| Left auditory cortex (nAm) | Right auditory cortex (nAm) | Left auditory cortex (nAm) | Right auditory cortex (nAm) | Left auditory cortex (nAm) | Right auditory cortex (nAm) | Left auditory cortex (nAm) | Right auditory cortex (nAm) | |
| N1 | 2.64 ± 0.93** | 5.32 ± 1.40* | 2.16 ± 0.96* | 2.79 ± 1.36 | 2.16 ± 0.73 | 5.14 ± 1.26 | 2.05 ± 0.57 | 2.70 ± 1.26 |
| N2 | 2.59 ± 0.80** | 3.34 ± 1.02** | 2.00 ± 0.82** | 2.86 ± 1.06** | 1.47 ± 0.70 | 2.11 ± 0.64 | 1.54 ± 0.78 | 1.74 ± 0.65 |
| N3 | 2.82 ± 1.00** | 5.44 ± 1.70* | 4.58 ± 1.34** | 4.15 ± 1.66** | 2.61 ± 1.01 | 5.04 ± 1.57 | 4.12 ± 1.15 | 3.36 ± 1.34 |
| N4 | 3.78 ± 1.33** | 8.66 ± 1.64** | 4.97 ± 1.66** | 6.29 ± 1.41** | 2.37 ± 0.99 | 5.24 ± 1.18 | 4.01 ± 1.23 | 5.29 ± 1.19 |
| N5 | 3.09 ± 1.15** | 5.77 ± 1.17** | 2.87 ± 0.80** | 6.02 ± 1.75** | 2.22 ± 0.76 | 4.33 ± 0.88 | 2.14 ± 0.62 | 4.27 ± 1.24 |
| Mean ± std | 2.98 ± 1.11 | 5.68 ± 2.22 | 3.28 ± 1.66 | 4.39 ± 2.14 | 2.19 ± 1.07 | 4.33 ± 1.64 | 2.76 ± 1.40 | 3.39 ± 1.65 |
The current strengths of CEEMD processing greater than traditional filtering method.
* p < 0.05 by paired t test.
** p<0.01 by paired t test.
Comparison of estimated current strengths in tinnitus patients
| Tinnitus patient | The proposed CEEMD approach | Traditional filtering method | ||||||
|---|---|---|---|---|---|---|---|---|
| Current strength of left ear stimulation | Current strength of right ear stimulation | Current strength of left ear stimulation | Current strength of right ear stimulation | |||||
| Left auditory cortex (nAm) | Right auditory cortex (nAm) | Left auditory cortex (nAm) | Right auditory cortex (nAm) | Left auditory cortex (nAm) | Right auditory cortex (nAm) | Left auditory cortex (nAm) | Right auditory cortex (nAm) | |
| P1 | 3.12 ± 0.98** | 6.15 ± 1.62** | 3.45 ± 1.15 | 4.21 ± 1.60** | 2.5 ± 0.78 | 4.98 ± 1.31 | 3.68 ± 1.95 | 3.63 ± 1.38 |
| P2 | 3.87 ± 1.60* | 6.25 ± 1.70** | 3.65 ± 1.48** | 8.83 ± 2.12** | 3.52 ± 1.15 | 4.93 ± 1.34 | 3.01 ± 1.10 | 6.00 ± 1.44 |
| P3 | 3.38 ± 1.36** | 4.66 ± 1.47** | 5.07 ± 2.10** | 10.28 ± 3.88** | 2.03 ± 0.69 | 3.23 ± 1.02 | 2.75 ± 1.66 | 7.07 ± 2.64 |
| P4 | 3.93 ± 1.09** | 8.87 ± 1.41** | 4.68 ± 1.77** | 2.86 ± 1.16** | 3.16 ± 0.63 | 6.54 ± 1.04 | 3.06 ± 1.28 | 2.11 ± 0.86 |
| P5 | 4.28 ± 1.51** | 5.70 ± 1.52** | 4.05 ± 1.49** | 7.90 ± 2.53** | 3.09 ± 1.03 | 3.61 ± 0.96 | 2.57 ± 1.05 | 4.37 ± 1.40 |
| Mean ± std | 3.76 ± 1.39 | 6.30 ± 2.06 | 4.14 ± 1.74 | 6.92 ± 3.68 | 2.86 ± 0.86 | 4.66 ± 1.13 | 2.98 ± 1.53 | 4.67 ± 2.33 |
The current strengths of CEEMD processing greater than traditional filtering method.
* p < 0.05 by paired t test.
** p < 0.01 by paired t test.
Figure 7The LI values during right and left auditory stimulations in normal subjects and tinnitus patients. a The LI values calculated from right-ear (red bars) and left-ear (blue bars) stimulations in normal subjects. b The LI values calculated from right-ear (red bars) and left-ear (blue bars) stimulations in tinnitus patients.