Accuracy of linear drilling in temporal bone using drill press system for minimally invasive cochlear implantation.

Neal P Dillon1, Ramya Balachandran2, Robert F Labadie2.   

Abstract

PURPOSE: A minimally invasive approach for cochlear implantation involves drilling a narrow linear path through the temporal bone from the skull surface directly to the cochlea for insertion of the electrode array without the need for an invasive mastoidectomy. Potential drill positioning errors must be accounted for to predict the effectiveness and safety of the procedure. The drilling accuracy of a system used for this procedure was evaluated in bone surrogate material under a range of clinically relevant parameters. Additional experiments were performed to isolate the error at various points along the path to better understand why deflections occur.
METHODS: An experimental setup to precisely position the drill press over a target was used. Custom bone surrogate test blocks were manufactured to resemble the mastoid region of the temporal bone. The drilling error was measured by creating divots in plastic sheets before and after drilling and using a microscope to localize the divots.
RESULTS: The drilling error was within the tolerance needed to avoid vital structures and ensure accurate placement of the electrode; however, some parameter sets yielded errors that may impact the effectiveness of the procedure when combined with other error sources. The error increases when the lateral stage of the path terminates in an air cell and when the guide bushings are positioned further from the skull surface. At contact points due to air cells along the trajectory, higher errors were found for impact angles of [Formula: see text] and higher as well as longer cantilevered drill lengths.
CONCLUSION: The results of these experiments can be used to define more accurate and safe drill trajectories for this minimally invasive surgical procedure.

Entities:  

Keywords:  Cochlear implantation; Drilling accuracy; Minimally invasive surgery; Temporal bone

Mesh:

Year:  2015        PMID: 26183149      PMCID: PMC4716002          DOI: 10.1007/s11548-015-1261-7

Source DB:  PubMed          Journal:  Int J Comput Assist Radiol Surg        ISSN: 1861-6410            Impact factor:   2.924


  28 in total

1.  Suprameatal approach: new surgical approach for cochlear implantation.

Authors:  J Kronenberg; L Migirov; T Dagan
Journal:  J Laryngol Otol       Date:  2001-04       Impact factor: 1.469

2.  The suprameatal approach: an alternative surgical approach to cochlear implantation.

Authors:  Jona Kronenberg; Wolfgang Baumgartner; Lela Migirov; Tal Dagan; Minka Hildesheimer
Journal:  Otol Neurotol       Date:  2004-01       Impact factor: 2.311

3.  A self-developed and constructed robot for minimally invasive cochlear implantation.

Authors:  Brett Bell; Christof Stieger; Nicolas Gerber; Andreas Arnold; Claude Nauer; Volkmar Hamacher; Martin Kompis; Lutz Nolte; Marco Caversaccio; Stefan Weber
Journal:  Acta Otolaryngol       Date:  2012-03-04       Impact factor: 1.494

4.  Percutaneous cochlear implant drilling via customized frames: an in vitro study.

Authors:  Ramya Balachandran; Jason E Mitchell; Grégoire Blachon; Jack H Noble; Benoit M Dawant; J Michael Fitzpatrick; Robert F Labadie
Journal:  Otolaryngol Head Neck Surg       Date:  2010-03       Impact factor: 3.497

5.  Local anesthesia for middle ear surgery.

Authors:  Gül Caner; Levent Olgun; Gürol Gültekin; Levent Aydar
Journal:  Otolaryngol Head Neck Surg       Date:  2005-08       Impact factor: 3.497

6.  Cochlear implantation without mastoidectomy: the pericanal electrode insertion technique.

Authors:  R Häusler
Journal:  Acta Otolaryngol       Date:  2002-10       Impact factor: 1.494

7.  Surgical considerations in cochlear implantation.

Authors:  W F House
Journal:  Ann Otol Rhinol Laryngol Suppl       Date:  1982 Mar-Apr

8.  Clinical validation study of percutaneous cochlear access using patient-customized microstereotactic frames.

Authors:  Robert F Labadie; Ramya Balachandran; Jason E Mitchell; Jack H Noble; Omid Majdani; David S Haynes; Marc L Bennett; Benoit M Dawant; J Michael Fitzpatrick
Journal:  Otol Neurotol       Date:  2010-01       Impact factor: 2.311

9.  Forces and trauma associated with minimally invasive image-guided cochlear implantation.

Authors:  Pooyan Rohani; Jason Pile; Lueder A Kahrs; Ramya Balachandran; Grégoire S Blachon; Nabil Simaan; Robert F Labadie
Journal:  Otolaryngol Head Neck Surg       Date:  2014-01-27       Impact factor: 3.497

10.  Cochlear implantation: minimal hair removal technique.

Authors:  J B Roberson; K R Stidham; K M Scott; L Tonokawa
Journal:  Otolaryngol Head Neck Surg       Date:  2000-05       Impact factor: 5.591

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.