Correlation of Serial Scleral and Corneal Pneumatonometry.

Debbie S Kuo1, Yvonne Ou1, Bennie H Jeng2, Robert Bhisitkul1, Jay M Stewart1, Jacque L Duncan1, Ying Han3.   

Abstract

PURPOSE: To evaluate the usefulness of scleral pneumatonometry as an alternative for corneal measurements of intraocular pressure (IOP) over a broad range of IOPs.
DESIGN: Prospective, observational cohort study. PARTICIPANTS: The study was conducted in the University of California, San Francisco, Retina Clinic between August and November 2013 in 33 adult patients (age range, 34-94 years; mean ± standard deviation, 74.1±13.4 years) receiving anti-vascular endothelial growth factor intravitreal injections, which transiently increase IOP.
METHODS: Corneal pachymetry and serial corneal and temporal scleral pneumatonometry (baseline, immediately after, and 10, 20, and 30 minutes after injection) were collected. One-time baseline corneal and scleral pneumatonometry readings were obtained in the noninjected eye. MAIN OUTCOME MEASURES: Correlation analysis and a Bland-Altman plot were used to evaluate reliability and agreement between scleral and corneal measurements of IOP. A linear mixed model was used to determine the relationship between measurements and to perform covariate analyses.
RESULTS: Scleral and corneal pneumatonometry showed nearly 1:1 linear correlation, although scleral pneumatonometry was biased toward higher values (r = 0.94; P < 0.001). Scleral pneumatonometry averaged 9.0 mmHg higher than corneal pneumatonometry (95% limits of agreement, -1.5 to 19.5 mmHg). A linear mixed model resulted in the following equation: corneal IOP = 1.04 × scleral IOP - 10.37. Age, central corneal thickness, laterality, and glaucoma and lens status did not impact this relationship. The difference between corneal and scleral pneumotonometry was correlated between the two eyes of individual patients (r = 0.75; P < 0.001).
CONCLUSIONS: Differences between serial scleral measurements reflect differences between serial corneal measurements. Scleral pneumatonometry should be considered as an alternative to corneal pneumatonometry for following patients in whom corneal measurements are unreliable or unobtainable.
Copyright © 2015 American Academy of Ophthalmology. Published by Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Substances:

Year:  2015        PMID: 26165473      PMCID: PMC4549181          DOI: 10.1016/j.ophtha.2015.05.033

Source DB:  PubMed          Journal:  Ophthalmology        ISSN: 0161-6420            Impact factor:   12.079


  13 in total

1.  Glaucoma associated with Boston type I keratoprosthesis.

Authors:  Roheena Kamyar; Jennifer S Weizer; Fernando Heitor de Paula; Joshua D Stein; Sayoko E Moroi; Denise John; David C Musch; Shahzad I Mian
Journal:  Cornea       Date:  2012-02       Impact factor: 2.651

2.  Agreement between methods of measurement with multiple observations per individual.

Authors:  J Martin Bland; Douglas G Altman
Journal:  J Biopharm Stat       Date:  2007       Impact factor: 1.051

3.  Correlation between corneal and scleral pneumatonometry: an alternative method for intraocular pressure measurement.

Authors:  Michael A Kapamajian; Jose de la Cruz; Joelle A Hallak; Thasarat S Vajaranant
Journal:  Am J Ophthalmol       Date:  2013-08-12       Impact factor: 5.258

4.  Prevalence, progression, and impact of glaucoma on vision after Boston type 1 keratoprosthesis surgery.

Authors:  Julia C Talajic; Younes Agoumi; Sébastien Gagné; Krystel Moussally; Mona Harissi-Dagher
Journal:  Am J Ophthalmol       Date:  2011-10-07       Impact factor: 5.258

5.  Scleral intraocular pressure measurement in cadaver eyes pre- and postkeratoprosthesis implantation.

Authors:  Charles C Lin; Aiyin Chen; Bennie H Jeng; Travis C Porco; Yvonne Ou; Ying Han
Journal:  Invest Ophthalmol Vis Sci       Date:  2014-04-09       Impact factor: 4.799

6.  Calculating correlation coefficients with repeated observations: Part 1--Correlation within subjects.

Authors:  J M Bland; D G Altman
Journal:  BMJ       Date:  1995-02-18

7.  Ocular rigidity in living human eyes.

Authors:  Ioannis G Pallikaris; George D Kymionis; Harilaos S Ginis; George A Kounis; Miltiadis K Tsilimbaris
Journal:  Invest Ophthalmol Vis Sci       Date:  2005-02       Impact factor: 4.799

8.  In vivo measurement of regional variation in anterior scleral resistance to Schiotz indentation.

Authors:  Hetal Patel; Bernard Gilmartin; Robert P Cubbidge; Nicola S Logan
Journal:  Ophthalmic Physiol Opt       Date:  2011-04-19       Impact factor: 3.117

9.  Glaucoma progression and role of glaucoma surgery in patients with Boston keratoprosthesis.

Authors:  Alja Crnej; Eleftherios I Paschalis; Borja Salvador-Culla; Allyson Tauber; Brigita Drnovsek-Olup; Lucy Q Shen; Claes H Dohlman
Journal:  Cornea       Date:  2014-04       Impact factor: 2.651

10.  Tonometry in adults and children. A manometric evaluation of pneumatonometry, applanation, and TonoPen in vitro and in vivo.

Authors:  D L Eisenberg; B G Sherman; C A McKeown; J S Schuman
Journal:  Ophthalmology       Date:  1998-07       Impact factor: 12.079

View more
  3 in total

1.  Assessment of Glaucomatous Damage After Boston Keratoprosthesis Implantation Based on Digital Planimetric Quantification of Visual Fields and Optic Nerve Head Imaging.

Authors:  Mohsin H Ali; Mark S Dikopf; Anthony G Finder; Ahmad A Aref; Thasarat Vajaranant; Jose de la Cruz; Maria Soledad Cortina
Journal:  Cornea       Date:  2018-05       Impact factor: 2.651

2.  Scleral Tonometry Precision During Scleral Lens Wear: A Pilot Study.

Authors:  Allen Y Ganjei; Gabriel G L Shlager; Daniel Brocks
Journal:  Clin Optom (Auckl)       Date:  2021-12-31

3.  Comparison of Pneumatonometry and Transpalpebral Tonometry Measurements of Intraocular Pressure during Scleral Lens Wear.

Authors:  Jennifer Swingle Fogt; Cherie B Nau; Muriel Schornack; Ellen Shorter; Amy Nau; Jennifer S Harthan
Journal:  Optom Vis Sci       Date:  2020-09       Impact factor: 2.106

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.