Automated tube voltage adaptation in combination with advanced modeled iterative reconstruction in thoracoabdominal third-generation 192-slice dual-source computed tomography: effects on image quality and radiation dose.

Jan-Erik Scholtz1, Julian L Wichmann2, Kristina Hüsers2, Martin Beeres2, Nour-Eldin Abdelrehim Nour-Eldin3, Claudia Frellesen2, Thomas J Vogl2, Thomas Lehnert2.   

Abstract

RATIONALE AND
OBJECTIVES: To evaluate image quality and radiation exposure of portal venous-phase thoracoabdominal third-generation 192-slice dual-source computed tomography (DSCT) with automated tube voltage adaptation (TVA) in combination with advanced modeled iterative reconstruction (ADMIRE).
MATERIALS AND METHODS: Fifty-one patients underwent oncologic portal venous-phase thoracoabdominal follow-up CT twice within 7 months. The initial examination was performed on second-generation 128-slice DSCT with fixed tube voltage of 120 kV in combination with filtered back projection reconstruction. The second examination was performed on a third-generation 192-slice DSCT using automated TVA in combination with ADMIRE. Attenuation and image noise of liver, spleen, renal cortex, aorta, vena cava inferior, portal vein, psoas muscle, and perinephric fat were measured. Signal-to-noise ratio (SNR) and contrast-to-noise ratio (CNR) were calculated. Radiation dose was assessed as size-specific dose estimates (SSDE). Subjective image quality was assessed by two observers using five-point Likert scales. Interobserver agreement was calculated using intraclass correlation coefficients (ICC).
RESULTS: Automated TVA set tube voltage to 90 kV (n = 8), 100 kV (n = 31), 110 kV (n = 11), or 120 kV (n = 1). Average SSDE was decreased by 34.9% using 192-slice DSCT compared to 128-slice 120-kV DSCT (7.8 ± 2.4 vs. 12.1 ± 3.2 mGy; P < .001). Image noise was substantially lower; SNR and CNR were significantly increased in 192-slice DSCT compared to 128-slice DSCT (all P < .005). Image quality was voted excellent for both acquisition techniques (5.00 vs. 4.93; P = .083).
CONCLUSIONS: Automated TVA in combination with ADMIRE on third-generation 192-slice DSCT in portal venous-phase thoracoabdominal CT provides excellent image quality with reduced image noise and increased SNR and CNR, whereas average radiation dose is reduced by 34.9% compared to 128-slice DSCT.
Copyright © 2015 AUR. Published by Elsevier Inc. All rights reserved.

Entities:  

Keywords:  Computed tomography; automated tube voltage adaptation; image quality; radiation dose

Mesh:

Substances:

Year:  2015        PMID: 26163706     DOI: 10.1016/j.acra.2015.05.010

Source DB:  PubMed          Journal:  Acad Radiol        ISSN: 1076-6332            Impact factor:   3.173


  11 in total

1.  90-kVp low-tube-voltage CT pulmonary angiography in combination with advanced modeled iterative reconstruction algorithm: effects on radiation dose, image quality and diagnostic accuracy for the detection of pulmonary embolism.

Authors:  Doris Leithner; Tatjana Gruber-Rouh; Martin Beeres; Julian L Wichmann; Scherwin Mahmoudi; Simon S Martin; Lukas Lenga; Moritz H Albrecht; Christian Booz; Thomas J Vogl; Jan-Erik Scholtz
Journal:  Br J Radiol       Date:  2018-06-05       Impact factor: 3.039

2.  Reaching for better image quality and lower radiation dose in head and neck CT: advanced modeled and sinogram-affirmed iterative reconstruction in combination with tube voltage adaptation.

Authors:  Andrea I Schmid; Michael Uder; Michael M Lell
Journal:  Dentomaxillofac Radiol       Date:  2016-09-08       Impact factor: 2.419

3.  Diagnostic yield of 90-kVp low-tube-voltage carotid and intracerebral CT-angiography: effects on radiation dose, image quality and diagnostic performance for the detection of carotid stenosis.

Authors:  Doris Leithner; Julian L Wichmann; Scherwin Mahmoudi; Simon S Martin; Moritz H Albrecht; Thomas J Vogl; Jan-Erik Scholtz
Journal:  Br J Radiol       Date:  2018-03-08       Impact factor: 3.039

4.  Advances in cardiac CT contrast injection and acquisition protocols.

Authors:  Jan-Erik Scholtz; Brian Ghoshhajra
Journal:  Cardiovasc Diagn Ther       Date:  2017-10

5.  Comparison of low- and ultralow-dose computed tomography protocols for quantitative lung and airway assessment.

Authors:  Emily Hammond; Chelsea Sloan; John D Newell; Jered P Sieren; Melissa Saylor; Craig Vidal; Shayna Hogue; Frank De Stefano; Alexa Sieren; Eric A Hoffman; Jessica C Sieren
Journal:  Med Phys       Date:  2017-08-02       Impact factor: 4.071

6.  Single-source chest-abdomen-pelvis cancer staging on a third generation dual-source CT system: comparison of automated tube potential selection to second generation dual-source CT.

Authors:  Clara Park; Tatjana Gruber-Rouh; Doris Leithner; Amelie Zierden; Mortiz H Albrecht; Julian L Wichmann; Boris Bodelle; Mohamed Elsabaie; Jan-Erik Scholtz; Moritz Kaup; Thomas J Vogl; Martin Beeres
Journal:  Cancer Imaging       Date:  2016-10-10       Impact factor: 3.909

7.  Two-Phase Contrast Injection Protocol for Pediatric Cardiac Computed Tomography in Children with Congenital Heart Disease.

Authors:  Naoki Fukuyama; Akira Kurata; Naoto Kawaguchi; Ryo Tashiro; Takashi Higaki; Takahiro Yokoi; Yuki Tanabe; Hikaru Nishiyama; Toshihide Itoh; Teruhito Kido; Masao Miyagawa; Teruhito Mochizuki
Journal:  Pediatr Cardiol       Date:  2017-12-06       Impact factor: 1.655

8.  Third-generation dual-source CT of the neck using automated tube voltage adaptation in combination with advanced modeled iterative reconstruction: evaluation of image quality and radiation dose.

Authors:  Jan-Erik Scholtz; Julian L Wichmann; Kristina Hüsers; Moritz H Albrecht; Martin Beeres; Ralf W Bauer; Thomas J Vogl; Boris Bodelle
Journal:  Eur Radiol       Date:  2015-11-11       Impact factor: 5.315

9.  Body composition determinants of radiation dose during abdominopelvic CT.

Authors:  Patrick D McLaughlin; Liam Chawke; Maria Twomey; Kevin P Murphy; Siobhán B O'Neill; Sebastian R McWilliams; Karl James; Richard G Kavanagh; Charles Sullivan; Faimee E Chan; Niamh Moore; Owen J O'Connor; Joseph A Eustace; Michael M Maher
Journal:  Insights Imaging       Date:  2017-10-23

10.  Automatic tube potential selection with tube current modulation in coronary CT angiography: Can it achieve consistent image quality among various individuals?

Authors:  Xiao-Ping Wang; Xiao-Mei Zhu; Yin-Su Zhu; Wang-Yan Liu; Xiao-Han Yang; Wei-Wei Huang; Yi Xu; Li-Jun Tang
Journal:  Exp Ther Med       Date:  2018-05-11       Impact factor: 2.447

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.