Single-source chest-abdomen-pelvis cancer staging on a third generation dual-source CT system: comparison of automated tube potential selection to second generation dual-source CT.

Clara Park1, Tatjana Gruber-Rouh1, Doris Leithner1, Amelie Zierden1, Mortiz H Albrecht1, Julian L Wichmann1, Boris Bodelle1, Mohamed Elsabaie1, Jan-Erik Scholtz1, Moritz Kaup1, Thomas J Vogl1, Martin Beeres2.   

Abstract

BACKGROUND: Evaluation of latest generation automated attenuation-based tube potential selection (ATPS) impact on image quality and radiation dose in contrast-enhanced chest-abdomen-pelvis computed tomography examinations for gynaecologic cancer staging.
METHODS: This IRB approved single-centre, observer-blinded retrospective study with a waiver for informed consent included a total of 100 patients with contrast-enhanced chest-abdomen-pelvis CT for gynaecologic cancer staging. All patients were examined with activated ATPS for adaption of tube voltage to body habitus. 50 patients were scanned on a third-generation dual-source CT (DSCT), and another 50 patients on a second-generation DSCT. Predefined image quality setting remained stable between both groups at 120 kV and a current of 210 Reference mAs. Subjective image quality assessment was performed by two blinded readers independently. Attenuation and image noise were measured in several anatomic structures. Signal-to-noise ratio (SNR) was calculated. For the evaluation of radiation exposure, CT dose index (CTDIvol) values were compared.
RESULTS: Diagnostic image quality was obtained in all patients. The median CTDIvol (6.1 mGy, range 3.9-22 mGy) was 40 % lower when using the algorithm compared with the previous ATCM protocol (median 10.2 mGy · cm, range 5.8-22.8 mGy). A reduction in potential to 90 kV occurred in 19 cases, a reduction to 100 kV in 23 patients and a reduction to 110 kV in 3 patients of our experimental cohort. These patients received significantly lower radiation exposure compared to the former used protocol.
CONCLUSION: Latest generation automated ATPS on third-generation DSCT provides good diagnostic image quality in chest-abdomen-pelvis CT while average radiation dose is reduced by 40 % compared to former ATPS protocol on second-generation DSCT.

Entities:  

Keywords:  Automated Tube Potential Selection; Cancer; Cancer Staging; Dual-Source CT; Multidetector Computed Tomography; Neoplasms

Mesh:

Substances:

Year:  2016        PMID: 27724954      PMCID: PMC5057380          DOI: 10.1186/s40644-016-0093-1

Source DB:  PubMed          Journal:  Cancer Imaging        ISSN: 1470-7330            Impact factor:   3.909


  21 in total

1.  Dose reduction in CT by anatomically adapted tube current modulation. I. Simulation studies.

Authors:  M Gies; W A Kalender; H Wolf; C Suess
Journal:  Med Phys       Date:  1999-11       Impact factor: 4.071

Review 2.  Computed tomography--an increasing source of radiation exposure.

Authors:  David J Brenner; Eric J Hall
Journal:  N Engl J Med       Date:  2007-11-29       Impact factor: 91.245

3.  Systematic Comparison of Reduced Tube Current Protocols for High-pitch and Standard-pitch Pulmonary CT Angiography in a Large Single-center Population.

Authors:  Andreas M Bucher; Matthias J Kerl; Moritz H Albrecht; Martin Beeres; Hanns Ackermann; Julian L Wichmann; Thomas J Vogl; Ralf W Bauer; T Lehnert
Journal:  Acad Radiol       Date:  2016-02-19       Impact factor: 3.173

Review 4.  Computed tomography--old ideas and new technology.

Authors:  Dominik Fleischmann; F Edward Boas
Journal:  Eur Radiol       Date:  2011-01-20       Impact factor: 5.315

5.  Attenuation-based automatic kilovolt selection in abdominal computed tomography: effects on radiation exposure and image quality.

Authors:  Achim Eller; Matthias S May; Michael Scharf; Axel Schmid; Michael Kuefner; Michael Uder; Michael M Lell
Journal:  Invest Radiol       Date:  2012-10       Impact factor: 6.016

6.  Dose reduction in CT by anatomically adapted tube current modulation. II. Phantom measurements.

Authors:  W A Kalender; H Wolf; C Suess
Journal:  Med Phys       Date:  1999-11       Impact factor: 4.071

7.  Dose reduction in CT examination of children by an attenuation-based on-line modulation of tube current (CARE Dose).

Authors:  Holger Greess; Anton Nömayr; Heiko Wolf; Ulrich Baum; Michael Lell; Bernhard Böwing; Willi Kalender; Werner A Bautz
Journal:  Eur Radiol       Date:  2002-02-02       Impact factor: 5.315

Review 8.  Techniques and applications of automatic tube current modulation for CT.

Authors:  Mannudeep K Kalra; Michael M Maher; Thomas L Toth; Bernhard Schmidt; Bryan L Westerman; Hugh T Morgan; Sanjay Saini
Journal:  Radiology       Date:  2004-10-21       Impact factor: 11.105

9.  Topogram-based automated selection of the tube potential and current in thoraco-abdominal trauma CT - a comparison to fixed kV with mAs modulation alone.

Authors:  Claudia Frellesen; Wenzel Stock; J Matthias Kerl; Thomas Lehnert; Julian L Wichmann; Christoph Nau; Emanuel Geiger; Sebastian Wutzler; Martin Beeres; Boris Schulz; Boris Bodelle; Hanns Ackermann; Thomas J Vogl; Ralf W Bauer
Journal:  Eur Radiol       Date:  2014-05-10       Impact factor: 5.315

10.  Chest-abdomen-pelvis CT for staging in cancer patients: dose effectiveness and image quality using automated attenuation-based tube potential selection.

Authors:  Martin Beeres; Marcus Römer; Boris Bodelle; Clara Lee; Tatjana Gruber-Rouh; Emmanuel Mbalisike; Josef M Kerl; Julian L Wichmann; Boris Schulz; Thomas J Vogl; Ralf W Bauer
Journal:  Cancer Imaging       Date:  2014-09-02       Impact factor: 3.909

View more
  2 in total

1.  Image quality and radiologists' subjective acceptance using model-based iterative and deep learning reconstructions as adjuncts to ultrahigh-resolution CT in low-dose contrast-enhanced abdominopelvic CT: phantom and clinical pilot studies.

Authors:  Makiko Nishikawa; Haruhiko Machida; Yuta Shimizu; Toshiya Kariyasu; Hiroyuki Morisaka; Takuya Adachi; Takehiro Nakai; Kosuke Sakaguchi; Shun Saito; Saki Matsumoto; Masamichi Koyanagi; Kenichi Yokoyama
Journal:  Abdom Radiol (NY)       Date:  2021-12-16

2.  The Image Quality and Diagnostic Performance of CT with Low-Concentration Iodine Contrast (240 mg Iodine/mL) for the Abdominal Organs.

Authors:  Moon-Hyung Choi; Young-Joon Lee; Seung-Eun Jung
Journal:  Diagnostics (Basel)       Date:  2022-03-19
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.