Photon counting spectral breast CT: effect of adaptive filtration on CT numbers, noise, and contrast to noise ratio.

Justin D Silkwood1, Kenneth L Matthews, Polad M Shikhaliev.   

Abstract

PURPOSE: Photon counting spectral (PCS) computed tomography (CT) shows promise for breast imaging. An issue with current photon-counting detectors is low count rate capabilities, artifacts resulting from nonuniform count rate across the field of view, and suboptimal spectral information. These issues are addressed in part by using tissue-equivalent adaptive filtration of the x-ray beam. The purpose of the study was to investigate the effect of adaptive filtration on different aspects of PCS breast CT.
METHODS: The theoretical formulation for the filter shape was derived for different filter materials and evaluated by simulation and an experimental prototype of the filter was fabricated from a tissue-like material (acrylic). The PCS CT images of a glandular breast phantom with adipose and iodine contrast elements were simulated at 40, 60, 90, and 120 kVp tube voltages, with and without adaptive filter. The CT numbers, CT noise, and contrast-to-noise ratio (CNR) were compared for spectral CT images acquired with and without adaptive filters. Similar comparison was made for material-decomposed PCS CT images.
RESULTS: The adaptive filter improved the uniformity of CT numbers, CT noise, and CNR in both ordinary and material decomposed PCS CT images. At the same tube output the average CT noise with adaptive filter, although uniform, was higher than the average noise without adaptive filter due to x-ray absorption by the filter. Increasing tube output, so that average skin exposure with the adaptive filter was same as without filter, made the noise with adaptive filter comparable to or lower than that without adaptive filter. Similar effects were observed when energy weighting was applied, and when material decompositions were performed using energy selective CT data.
CONCLUSIONS: An adaptive filter decreases count rate requirements to the photon counting detectors which enables PCS breast CT based on commercially available detector technologies. Adaptive filter also improves image quality in PCS breast CT by decreasing beam hardening artifacts and by eliminating spatial nonuniformities of CT numbers, noise, and CNR.

Entities:  

Mesh:

Substances:

Year:  2013        PMID: 23635275     DOI: 10.1118/1.4800504

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  8 in total

1.  Characterization of scatter magnitude and distribution in dedicated breast computed tomography with bowtie filters.

Authors:  Kimberly Kontson; Robert J Jennings
Journal:  J Med Imaging (Bellingham)       Date:  2014-12-18

Review 2.  Photon-counting Detector CT: System Design and Clinical Applications of an Emerging Technology.

Authors:  Shuai Leng; Michael Bruesewitz; Shengzhen Tao; Kishore Rajendran; Ahmed F Halaweish; Norbert G Campeau; Joel G Fletcher; Cynthia H McCollough
Journal:  Radiographics       Date:  2019 May-Jun       Impact factor: 5.333

3.  Cascaded systems analysis of photon counting detectors.

Authors:  J Xu; W Zbijewski; G Gang; J W Stayman; K Taguchi; M Lundqvist; E Fredenberg; J A Carrino; J H Siewerdsen
Journal:  Med Phys       Date:  2014-10       Impact factor: 4.071

4.  Volumetric CT with sparse detector arrays (and application to Si-strip photon counters).

Authors:  A Sisniega; W Zbijewski; J W Stayman; J Xu; K Taguchi; E Fredenberg; Mats Lundqvist; J H Siewerdsen
Journal:  Phys Med Biol       Date:  2015-11-27       Impact factor: 3.609

Review 5.  Tutorial on X-ray photon counting detector characterization.

Authors:  Liqiang Ren; Bin Zheng; Hong Liu
Journal:  J Xray Sci Technol       Date:  2018       Impact factor: 1.535

6.  Evaluation of conventional imaging performance in a research whole-body CT system with a photon-counting detector array.

Authors:  Zhicong Yu; Shuai Leng; Steven M Jorgensen; Zhoubo Li; Ralf Gutjahr; Baiyu Chen; Ahmed F Halaweish; Steffen Kappler; Lifeng Yu; Erik L Ritman; Cynthia H McCollough
Journal:  Phys Med Biol       Date:  2016-02-02       Impact factor: 3.609

7.  Effects of kV, filtration, dose, and object size on soft tissue and iodine contrast in dedicated breast CT.

Authors:  Andrew M Hernandez; Craig K Abbey; Peymon Ghazi; George Burkett; John M Boone
Journal:  Med Phys       Date:  2020-04-27       Impact factor: 4.071

Review 8.  Dedicated breast CT: state of the art-Part I. Historical evolution and technical aspects.

Authors:  Yueqiang Zhu; Avice M O'Connell; Yue Ma; Aidi Liu; Haijie Li; Yuwei Zhang; Xiaohua Zhang; Zhaoxiang Ye
Journal:  Eur Radiol       Date:  2021-08-03       Impact factor: 7.034

  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.