Constrained Source Space MR Spectroscopy: Multiple Voxels, No Gradient Readout.

K Landheer1, A Sahgal2, S Das3, S J Graham4.   

Abstract

BACKGROUND AND
PURPOSE: Our goal was to develop a novel technique for measuring a small number of localized spectra simultaneously and in a time-efficient manner.
MATERIALS AND METHODS: Using appropriate radiofrequency pulses, the magnetization from multiple voxels is excited simultaneously and then separated (reconstructed) by using the individual coil-sensitivity profiles from a multichannel receiver coil. Because no gradients are used for k-space encoding, constrained source space MR spectroscopy provides a time advantage over conventional spectroscopic imaging and an improved signal-to-noise ratio per square root of unit time over single-voxel spectroscopy applied at each successive location. In the present work, we considered prototype application of constrained source space MR spectroscopy for 2 voxels.
RESULTS: Experimental data from healthy volunteers and simulation results showed that constrained source space MR spectroscopy is effective at extracting 2 independent spectra even in the challenging scenario of the voxels being closely spaced. Also, from 6 patients with various types of brain cancer we obtained 2-voxel constrained source space MR spectroscopy data, which showed spectra of clinical quality in half the time required to perform successive single-voxel MR spectroscopy.
CONCLUSIONS: Constrained source space MR spectroscopy provides clinical quality spectra and could be used to probe multiple voxels simultaneously in combination with Hadamard encoding for further scan-time reductions.
© 2015 by American Journal of Neuroradiology.

Entities:  

Mesh:

Year:  2015        PMID: 26089315      PMCID: PMC7964693          DOI: 10.3174/ajnr.A4319

Source DB:  PubMed          Journal:  AJNR Am J Neuroradiol        ISSN: 0195-6108            Impact factor:   3.825


  20 in total

1.  SENSE: sensitivity encoding for fast MRI.

Authors:  K P Pruessmann; M Weiger; M B Scheidegger; P Boesiger
Journal:  Magn Reson Med       Date:  1999-11       Impact factor: 4.668

2.  An electromagnetic reverse method of coil sensitivity mapping for parallel MRI - theoretical framework.

Authors:  Jin Jin; Feng Liu; Ewald Weber; Yu Li; Stuart Crozier
Journal:  J Magn Reson       Date:  2010-08-17       Impact factor: 2.229

3.  An automated quantitation of short echo time MRS spectra in an open source software environment: AQSES.

Authors:  Jean-Baptiste Poullet; Diana M Sima; Arjan W Simonetti; Bart De Neuter; Leentje Vanhamme; Philippe Lemmerling; Sabine Van Huffel
Journal:  NMR Biomed       Date:  2007-08       Impact factor: 4.044

4.  Parameter relations for the Shinnar-Le Roux selective excitation pulse design algorithm [NMR imaging].

Authors:  J Pauly; P Le Roux; D Nishimura; A Macovski
Journal:  IEEE Trans Med Imaging       Date:  1991       Impact factor: 10.048

Review 5.  Spatial localization in NMR spectroscopy in vivo.

Authors:  P A Bottomley
Journal:  Ann N Y Acad Sci       Date:  1987       Impact factor: 5.691

6.  Localized 2D J-resolved 1H MR spectroscopy: strong coupling effects in vitro and in vivo.

Authors:  L N Ryner; J A Sorenson; M A Thomas
Journal:  Magn Reson Imaging       Date:  1995       Impact factor: 2.546

7.  Single-shot magnetic resonance spectroscopic imaging with partial parallel imaging.

Authors:  Stefan Posse; Ricardo Otazo; Shang-Yueh Tsai; Akio Ernesto Yoshimoto; Fa-Hsuan Lin
Journal:  Magn Reson Med       Date:  2009-03       Impact factor: 4.668

8.  Field, coil, and echo-time influence on sensitivity and reproducibility of brain proton MR spectroscopy.

Authors:  M Inglese; M Spindler; J S Babb; P Sunenshine; M Law; O Gonen
Journal:  AJNR Am J Neuroradiol       Date:  2006-03       Impact factor: 3.825

9.  Parallel spectroscopic imaging reconstruction with arbitrary trajectories using k-space sparse matrices.

Authors:  Meng Gu; Chunlei Liu; Daniel M Spielman
Journal:  Magn Reson Med       Date:  2009-02       Impact factor: 4.668

10.  Accelerated regularized estimation of MR coil sensitivities using augmented Lagrangian methods.

Authors:  Michael J Allison; Sathish Ramani; Jeffrey A Fessler
Journal:  IEEE Trans Med Imaging       Date:  2012-11-22       Impact factor: 10.048