Absolute reliability of adipose tissue volume measurement by computed tomography: application of low-dose scan and minimal detectable change--a phantom study.

Takahiro Onuma1, Tamotsu Kamishima, Tsukasa Sasaki, Motomichi Sakata.   

Abstract

Metabolic syndrome increases the risk of developing diabetes and cardiovascular disease, particularly heart failure. Abdominal obesity is commonly assessed by measurement of the waist circumference, which exhibits a positive correlation with the visceral fat area measured on computed tomography (CT). CT is an excellent technique for measurement of cross-sectional areas of adipose tissue, but the exposure to ionizing radiation limits broad and repeated application in healthy subjects. Our purpose in this study was to determine the reliability of low-dose CT for abdominal fat quantification as compared with a standard CT protocol. A phantom was scanned by use of changes in the volume of vegetable oil, simulating visceral and subcutaneous adipose tissue, and by changes in the tube current-time products (25-300 mAs). We measured the volume of vegetable oil for each mAs value, and we calculated the minimal detectable change (MDC) in the volume by making repeated measurements. The measured volume of vegetable oil at 50 mAs and higher was not significantly different (p > 0.05), but that at 25 mAs was significantly different (p < 0.001), from that at 300 mAs. The MDC was less than 0.4 ml regardless of the mAs value at all mAs values assessed. We suggest that the adipose tissue volume is determined accurately by CT at 50 mAs (75 % reduction of radiation exposure compared with the standard dose).

Entities:  

Mesh:

Year:  2015        PMID: 26055571     DOI: 10.1007/s12194-015-0322-5

Source DB:  PubMed          Journal:  Radiol Phys Technol        ISSN: 1865-0333


  21 in total

Review 1.  Understanding the relevance of measured change through studies of responsiveness.

Authors:  D E Beaton
Journal:  Spine (Phila Pa 1976)       Date:  2000-12-15       Impact factor: 3.468

2.  Comparison of low-dose CT and MR for measurement of intra-abdominal adipose tissue: a phantom and human study.

Authors:  Dae Young Yoon; Jeung Hee Moon; Heon Kyu Kim; Chul Soon Choi; Suk Ki Chang; Eun Joo Yun; Young Lan Seo
Journal:  Acad Radiol       Date:  2008-01       Impact factor: 3.173

Review 3.  Metabolic syndrome--a new world-wide definition. A Consensus Statement from the International Diabetes Federation.

Authors:  K G M M Alberti; P Zimmet; J Shaw
Journal:  Diabet Med       Date:  2006-05       Impact factor: 4.359

4.  Statistical methods for assessing agreement between two methods of clinical measurement.

Authors:  J M Bland; D G Altman
Journal:  Lancet       Date:  1986-02-08       Impact factor: 79.321

5.  Measurement of abdominal and visceral fat with computed tomography and dual-energy x-ray absorptiometry.

Authors:  M D Jensen; J A Kanaley; J E Reed; P F Sheedy
Journal:  Am J Clin Nutr       Date:  1995-02       Impact factor: 7.045

6.  Effects of abdominal fat distribution parameters on severity of acute pancreatitis.

Authors:  D P O'Leary; D O'Neill; P McLaughlin; S O'Neill; E Myers; M M Maher; H P Redmond
Journal:  World J Surg       Date:  2012-07       Impact factor: 3.352

7.  A novel method of measuring intra-abdominal fat volume using helical computed tomography.

Authors:  J Kobayashi; N Tadokoro; M Watanabe; M Shinomiya
Journal:  Int J Obes Relat Metab Disord       Date:  2002-03

8.  Definition, diagnosis and classification of diabetes mellitus and its complications. Part 1: diagnosis and classification of diabetes mellitus provisional report of a WHO consultation.

Authors:  K G Alberti; P Z Zimmet
Journal:  Diabet Med       Date:  1998-07       Impact factor: 4.359

9.  Accurate tissue area measurements with considerably reduced radiation dose achieved by patient-specific CT scan parameters.

Authors:  J Brandberg; L Lonn; E Bergelin; L Sjostrom; E Forssell-Aronsson; G Starck
Journal:  Br J Radiol       Date:  2008-06-30       Impact factor: 3.039

10.  How much change is true change? The minimum detectable change of the Berg Balance Scale in elderly people.

Authors:  Declan Donoghue; Emma K Stokes
Journal:  J Rehabil Med       Date:  2009-04       Impact factor: 2.912

View more
  3 in total

Review 1.  Diagnostic imaging in the management of patients with metabolic syndrome.

Authors:  Seo Rin Kim; Lilach O Lerman
Journal:  Transl Res       Date:  2017-11-22       Impact factor: 7.012

2.  Renal Adiposity Does not Preclude Quantitative Assessment of Renal Function Using Dual-Energy Multidetector CT in Mildly Obese Human Subjects.

Authors:  Christopher M Ferguson; Alfonso Eirin; Gregory J Michalak; Ahmad F Hedayat; Abdelrhman M Abumoawad; Ahmed Saad; Xiangyang Zhu; Stephen C Textor; Cynthia H McCollough; Lilach O Lerman
Journal:  Acad Radiol       Date:  2019-01-14       Impact factor: 3.173

3.  Reproducibility and Repeatability of Computer Tomography-based Measurement of Abdominal Subcutaneous and Visceral Adipose Tissues.

Authors:  Yuan-Hao Lee; Hsing-Fen Hsiao; Hou-Ting Yang; Shih-Yi Huang; Wing P Chan
Journal:  Sci Rep       Date:  2017-01-10       Impact factor: 4.379

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.