Literature DB >> 26052176

Task-driven imaging in cone-beam computed tomography.

G J Gang1, J W Stayman1, S Ouadah1, T Ehtiati2, J H Siewerdsen1.   

Abstract

PURPOSE: Conventional workflow in interventional imaging often ignores a wealth of prior information of the pan class="Species">patient anatomy and the imaging task. This work introduces a task-driven imaging framework that utilizes such information to prospectively design acquisition and reconstruction techniques for cone-beam CT (CBCT) in a manner that maximizes task-based performance in subsequent imaging procedures.
METHODS: The framework is employed in jointly optimizing tube current modulation, orbital tilt, and reconstruction parameters in filtered backprojection reconstruction for interventional imaging. Theoretical predictors of noise and resolution relates acquisition and reconstruction parameters to task-based detectability. Given a patient-specific prior image and specification of the imaging task, an optimization algorithm prospectively identifies the combination of imaging parameters that maximizes task-based detectability. Initial investigations were performed for a variety of imaging tasks in an elliptical phantom and an anthropomorphic head phantom.
RESULTS: Optimization of tube current modulation and view-dependent reconstruction kernel was shown to have greatest benefits for a directional task (e.g., identification of device or tissue orientation). The task-driven approach yielded techniques in which the dose and sharp kernels were concentrated in views contributing the most to the signal power associated with the imaging task. For example, detectability of a line pair detection task was improved by at least three fold compared to conventional approaches. For radially symmetric tasks, the task-driven strategy yielded results similar to a minimum variance strategy in the absence of kernel modulation. Optimization of the orbital tilt successfully avoided highly attenuating structures that can confound the imaging task by introducing noise correlations masquerading at spatial frequencies of interest.
CONCLUSIONS: This work demonstrated the potential of a task-driven imaging framework to improve image quality and reduce dose beyond that achievable with conventional imaging approaches.

Entities:  

Year:  2015        PMID: 26052176      PMCID: PMC4457372          DOI: 10.1117/12.2082169

Source DB:  PubMed          Journal:  Proc SPIE Int Soc Opt Eng        ISSN: 0277-786X


  16 in total

1.  Dose reduction in CT by anatomically adapted tube current modulation. I. Simulation studies.

Authors:  M Gies; W A Kalender; H Wolf; C Suess
Journal:  Med Phys       Date:  1999-11       Impact factor: 4.071

2.  Flat-panel cone-beam computed tomography for image-guided radiation therapy.

Authors:  David A Jaffray; Jeffrey H Siewerdsen; John W Wong; Alvaro A Martinez
Journal:  Int J Radiat Oncol Biol Phys       Date:  2002-08-01       Impact factor: 7.038

3.  Volume CT with a flat-panel detector on a mobile, isocentric C-arm: pre-clinical investigation in guidance of minimally invasive surgery.

Authors:  J H Siewerdsen; D J Moseley; S Burch; S K Bisland; A Bogaards; B C Wilson; D A Jaffray
Journal:  Med Phys       Date:  2005-01       Impact factor: 4.071

4.  Dose and image quality for a cone-beam C-arm CT system.

Authors:  Rebecca Fahrig; Robert Dixon; Thomas Payne; Richard L Morin; Arundhuti Ganguly; Norbert Strobel
Journal:  Med Phys       Date:  2006-12       Impact factor: 4.071

5.  Mean and variance of implicitly defined biased estimators (such as penalized maximum likelihood): applications to tomography.

Authors:  J A Fessler
Journal:  IEEE Trans Image Process       Date:  1996       Impact factor: 10.856

6.  1996 RSNA Eugene P. Pendergrass New Horizons Lecture. Image-guided procedures and the operating room of the future.

Authors:  F A Jolesz
Journal:  Radiology       Date:  1997-09       Impact factor: 11.105

7.  Task-based detectability in CT image reconstruction by filtered backprojection and penalized likelihood estimation.

Authors:  Grace J Gang; J Webster Stayman; Wojciech Zbijewski; Jeffrey H Siewerdsen
Journal:  Med Phys       Date:  2014-08       Impact factor: 4.071

8.  Statistically defined backgrounds: performance of a modified nonprewhitening observer model.

Authors:  A E Burgess
Journal:  J Opt Soc Am A Opt Image Sci Vis       Date:  1994-04       Impact factor: 2.129

9.  Tomotherapy: a new concept for the delivery of dynamic conformal radiotherapy.

Authors:  T R Mackie; T Holmes; S Swerdloff; P Reckwerdt; J O Deasy; J Yang; B Paliwal; T Kinsella
Journal:  Med Phys       Date:  1993 Nov-Dec       Impact factor: 4.071

10.  Analysis of Fourier-domain task-based detectability index in tomosynthesis and cone-beam CT in relation to human observer performance.

Authors:  Grace J Gang; Junghoon Lee; J Webster Stayman; Daniel J Tward; W Zbijewski; Jerry L Prince; Jeffrey H Siewerdsen
Journal:  Med Phys       Date:  2011-04       Impact factor: 4.071

View more