Attenuation correction for flexible magnetic resonance coils in combined magnetic resonance/positron emission tomography imaging.

Mootaz Eldib1, Jason Bini, Claudia Calcagno, Philip M Robson, Venkatesh Mani, Zahi A Fayad.   

Abstract

INTRODUCTION: Attenuation correction for magnetic resonance (MR) coils is a new challenge that came about with the development of combined MR and positron emission tomography (PET) imaging. This task is difficult because such coils are not directly visible on either PET or MR acquisitions with current combined scanners and are therefore not easily localized in the field of view. This issue becomes more evident when trying to localize flexible MR coils (eg, cardiac or body matrix coil) that change position and shape from patient to patient and from one imaging session to another. In this study, we proposed a novel method to localize and correct for the attenuation and scatter of a flexible MR cardiac coil, using MR fiducial markers placed on the surface of the coil to allow for accurate registration of a template computed tomography (CT)-based attenuation map.
MATERIALS AND METHODS: To quantify the attenuation properties of the cardiac coil, a uniform cylindrical water phantom injected with 18F-fluorodeoxyglucose (18F-FDG) was imaged on a sequential MR/PET system with and without the flexible cardiac coil. After establishing the need to correct for the attenuation of the coil, we tested the feasibility of several methods to register a precomputed attenuation map to correct for the attenuation. To accomplish this, MR and CT visible markers were placed on the surface of the cardiac flexible coil. Using only the markers as a driver for registration, the CT image was registered to the reference image through a combination of rigid and deformable registration. The accuracy of several methods was compared for the deformable registration, including B-spline, thin-plate spline, elastic body spline, and volume spline. Finally, we validated our novel approach both in phantom and patient studies.
RESULTS: The findings from the phantom experiments indicated that the presence of the coil resulted in a 10% reduction in measured 18F-FDG activity when compared with the phantom-only scan. Local underestimation reached 22% in regions of interest close to the coil. Various registration methods were tested, and the volume spline was deemed to be the most accurate, as measured by the Dice similarity metric. The results of our phantom experiments showed that the bias in the 18F-FDG quantification introduced by the presence of the coil could be reduced by using our registration method. An overestimation of only 1.9% of the overall activity for the phantom scan with the coil attenuation map was measured when compared with the baseline phantom scan without coil. A local overestimation of less than 3% was observed in the ROI analysis when using the proposed method to correct for the attenuation of the flexible cardiac coil. Quantitative results from the patient study agreed well with the phantom findings.
CONCLUSIONS: We presented and validated an accurate method to localize and register a CT-based attenuation map to correct for the attenuation and scatter of flexible MR coils. This method may be translated to clinical use to produce quantitatively accurate measurements with the use of flexible MR coils during MR/PET imaging.

Entities:  

Mesh:

Year:  2014        PMID: 24056110      PMCID: PMC4011564          DOI: 10.1097/RLI.0b013e3182a530f8

Source DB:  PubMed          Journal:  Invest Radiol        ISSN: 0020-9996            Impact factor:   6.016


  25 in total

1.  Nonrigid registration using free-form deformations: application to breast MR images.

Authors:  D Rueckert; L I Sonoda; C Hayes; D L Hill; M O Leach; D J Hawkes
Journal:  IEEE Trans Med Imaging       Date:  1999-08       Impact factor: 10.048

2.  Method for transforming CT images for attenuation correction in PET/CT imaging.

Authors:  Jonathan P J Carney; David W Townsend; Vitaliy Rappoport; Bernard Bendriem
Journal:  Med Phys       Date:  2006-04       Impact factor: 4.071

3.  Image artifacts from MR-based attenuation correction in clinical, whole-body PET/MRI.

Authors:  Sune H Keller; Søren Holm; Adam E Hansen; Bernhard Sattler; Flemming Andersen; Thomas L Klausen; Liselotte Højgaard; Andreas Kjær; Thomas Beyer
Journal:  MAGMA       Date:  2012-09-21       Impact factor: 2.310

4.  Hybrid MRI/PET of the heart: a new complementary imaging technique for simultaneous acquisition of MRI and PET data.

Authors:  Thomas Schlosser; Felix Nensa; Amir A Mahabadi; Thorsten D Poeppel
Journal:  Heart       Date:  2012-11-13       Impact factor: 5.994

5.  Simultaneous PET and MR imaging.

Authors:  Y Shao; S R Cherry; K Farahani; K Meadors; S Siegel; R W Silverman; P K Marsden
Journal:  Phys Med Biol       Date:  1997-10       Impact factor: 3.609

6.  Design and performance evaluation of a whole-body Ingenuity TF PET-MRI system.

Authors:  H Zaidi; N Ojha; M Morich; J Griesmer; Z Hu; P Maniawski; O Ratib; D Izquierdo-Garcia; Z A Fayad; L Shao
Journal:  Phys Med Biol       Date:  2011-04-20       Impact factor: 3.609

7.  Toward simultaneous PET/MR breast imaging: systematic evaluation and integration of a radiofrequency breast coil.

Authors:  Bassim Aklan; Daniel H Paulus; Evelyn Wenkel; Harald Braun; Bharath K Navalpakkam; Susanne Ziegler; Christian Geppert; Eric E Sigmund; Amy Melsaether; Harald H Quick
Journal:  Med Phys       Date:  2013-02       Impact factor: 4.071

8.  Tissue classification as a potential approach for attenuation correction in whole-body PET/MRI: evaluation with PET/CT data.

Authors:  Axel Martinez-Möller; Michael Souvatzoglou; Gaspar Delso; Ralph A Bundschuh; Christophe Chefd'hotel; Sibylle I Ziegler; Nassir Navab; Markus Schwaiger; Stephan G Nekolla
Journal:  J Nucl Med       Date:  2009-03-16       Impact factor: 10.057

9.  Magnetic resonance-based attenuation correction for PET/MR hybrid imaging using continuous valued attenuation maps.

Authors:  Bharath K Navalpakkam; Harald Braun; Torsten Kuwert; Harald H Quick
Journal:  Invest Radiol       Date:  2013-05       Impact factor: 6.016

10.  PET/MR for therapy response evaluation in malignant lymphoma: initial experience.

Authors:  Ivan Platzek; Bettina Beuthien-Baumann; Jens Langner; Manuel Popp; Georg Schramm; Rainer Ordemann; Michael Laniado; Jörg Kotzerke; Jörg van den Hoff
Journal:  MAGMA       Date:  2012-09-16       Impact factor: 2.310

View more
  13 in total

1.  Feasibility of (18)F-Fluorodeoxyglucose radiotracer dose reduction in simultaneous carotid PET/MR imaging.

Authors:  Mootaz Eldib; Jason Bini; Olivier Lairez; David D Faul; Niels Oesingmann; Zahi A Fayad; Venkatesh Mani
Journal:  Am J Nucl Med Mol Imaging       Date:  2015-06-15

2.  Quantitative carotid PET/MR imaging: clinical evaluation of MR-Attenuation correction versus CT-Attenuation correction in (18)F-FDG PET/MR emission data and comparison to PET/CT.

Authors:  Jason Bini; Philip M Robson; Claudia Calcagno; Mootaz Eldib; Zahi A Fayad
Journal:  Am J Nucl Med Mol Imaging       Date:  2015-02-15

3.  Markerless attenuation correction for carotid MRI surface receiver coils in combined PET/MR imaging.

Authors:  Mootaz Eldib; Jason Bini; Philip M Robson; Claudia Calcagno; David D Faul; Charalampos Tsoumpas; Zahi A Fayad
Journal:  Phys Med Biol       Date:  2015-05-28       Impact factor: 3.609

Review 4.  Attenuation Correction for Magnetic Resonance Coils in Combined PET/MR Imaging: A Review.

Authors:  Mootaz Eldib; Jason Bini; David D Faul; Niels Oesingmann; Charalampos Tsoumpas; Zahi A Fayad
Journal:  PET Clin       Date:  2015-11-27

Review 5.  Emerging role of integrated PET-MRI in osteoarthritis.

Authors:  Amarnath Jena; Sangeeta Taneja; Prerana Rana; Nidhi Goyal; Abhishek Vaish; Rajesh Botchu; Raju Vaishya
Journal:  Skeletal Radiol       Date:  2021-06-29       Impact factor: 2.199

Review 6.  Applications of PET-MRI in musculoskeletal disease.

Authors:  Feliks Kogan; Stephen M Broski; Daehyun Yoon; Garry E Gold
Journal:  J Magn Reson Imaging       Date:  2018-07       Impact factor: 4.813

7.  Evaluation of a Dedicated Radiofrequency Carotid PET/MRI Coil.

Authors:  Mueez Aizaz; Jochem A J van der Pol; Roel Wierts; Hans Zwart; Abe J van der Werf; Joachim E Wildberger; Jan A Bucerius; Rik P M Moonen; Marianne Eline Kooi
Journal:  J Clin Med       Date:  2022-05-04       Impact factor: 4.241

8.  MLAA-based attenuation correction of flexible hardware components in hybrid PET/MR imaging.

Authors:  Thorsten Heußer; Christopher M Rank; Yannick Berker; Martin T Freitag; Marc Kachelrieß
Journal:  EJNMMI Phys       Date:  2017-03-01

9.  PET Image Quality Improvement for Simultaneous PET/MRI with a Lightweight MRI Surface Coil.

Authors:  Timothy W Deller; Nicholas K Mathew; Samuel A Hurley; Chad M Bobb; Alan B McMillan
Journal:  Radiology       Date:  2020-11-03       Impact factor: 11.105

10.  Scan-rescan measurement repeatability of 18F-FDG PET/MR imaging of vascular inflammation.

Authors:  Philip M Robson; Audrey Kaufman; Alison Pruzan; Marc R Dweck; Maria-Giovanna Trivieri; Ronan Abgral; Nicolas A Karakatsanis; Patrick M Brunner; Emma Guttman; Zahi A Fayad; Venkatesh Mani
Journal:  J Nucl Cardiol       Date:  2021-05-27       Impact factor: 3.872

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.