Jeam Haroldo Oliveira Barbosa1, Antonio Carlos Santos2, Carlos Ernesto Garrido Salmon3. 1. Master, Fellow PhD degree, Faculdade de Filosofia, Ciências e Letras de Ribeirão Preto da Universidade de São Paulo (FFCLRP-USP), Ribeirão Preto, SP, Brazil. 2. PhD, Professor, Faculdade de Medicina de Ribeirão Preto da Universidade de São Paulo (FMRP-USP), Ribeirão Preto, SP, Brazil. 3. PhD, Professor, Faculdade de Filosofia, Ciências e Letras de Ribeirão Preto da Universidade de São Paulo (FFCLRP-USP), Ribeirão Preto, SP, Brazil.
Abstract
OBJECTIVE: To present a detailed explanation on the processing of magnetic susceptibility weighted imaging (SWI), demonstrating the effects of echo time and sensitive mask on the differentiation between calcification and hemosiderin. MATERIALS AND METHODS: Computed tomography and magnetic resonance (magnitude and phase) images of six patients (age range 41- 54 years; four men) were retrospectively selected. The SWI images processing was performed using the Matlab's own routine. RESULTS: Four out of the six patients showed calcifications at computed tomography images and their SWI images demonstrated hyperintense signal at the calcification regions. The other patients did not show any calcifications at computed tomography, and SWI revealed the presence of hemosiderin deposits with hypointense signal. CONCLUSION: The selection of echo time and of the mask may change all the information on SWI images, and compromise the diagnostic reliability. Amongst the possible masks, the authors highlight that the sigmoid mask allows for contrasting calcifications and hemosiderin on a single SWI image.
OBJECTIVE: To present a detailed explanation on the processing of magnetic susceptibility weighted imaging (SWI), demonstrating the effects of echo time and sensitive mask on the differentiation between calcification and hemosiderin. MATERIALS AND METHODS: Computed tomography and magnetic resonance (magnitude and phase) images of six patients (age range 41- 54 years; four men) were retrospectively selected. The SWI images processing was performed using the Matlab's own routine. RESULTS: Four out of the six patients showed calcifications at computed tomography images and their SWI images demonstrated hyperintense signal at the calcification regions. The other patients did not show any calcifications at computed tomography, and SWI revealed the presence of hemosiderin deposits with hypointense signal. CONCLUSION: The selection of echo time and of the mask may change all the information on SWI images, and compromise the diagnostic reliability. Amongst the possible masks, the authors highlight that the sigmoid mask allows for contrasting calcifications and hemosiderin on a single SWI image.
Entities:
Keywords:
Calcification; Hemosiderin; Magnetic resonance imaging; Magnetic susceptibility; SWI
In the clinical practice, the most frequently utilized magnetic resonance imaging
modality is fundamentally weighted by relaxation time of tissues. At these images, the
contrast is based on the significant differences between transverse relaxation (T2)
and/or longitudinal relaxation (T1) in the study region. However, relaxometry-weighted
images cannot differentiate between hemosiderin and calcifications since both reduce the
relaxation times due to inhomogeneity of the magnetic field generated by paramagnetic
and diamagnetic atoms, respectively. On the other hand, susceptibility-weighted imaging
(SWI) enhances the contrast of calcifications and hemosiderin deposits(. Thus, SWI has supplemented the clinical diagnosis of neurological
disorders (cranioencephalic trauma and harmful clots), hemorrhagic disorders (vascular
malformation, cerebral infarction and neoplasias) and neuroinfectious conditions
(neurotoxoplasmosis and neurocysticercosis)(.The SWI is generated by means of multiplication of magnitude image by a mask with values
between 0 and 1 obtained from the phase image. Both images are generally acquired with a
conventional clinical gradient echo (GE) sequence with appropriate echo time
(TE)(. The main information on
the susceptibility distribution (χ) on the SWI is on the phase image where the
phase (φ) in a determined position (r) and with a certain TE is
approximately:where: γ is the gyromagnetic ratio of the hydrogen nucleus. If the phase of the
nuclei of water molecules contained in the greatest part of the healthy brain tissue is
considered as a reference phase, the equation 1 can be expressed in terms of phase
difference:where: ∆χ(r) represents the difference of
χtissue - χwater. Thus, for relatively short echo
times, such phase variation assumes values between 0 and +π for paramagnetic
tissue (∆χ > 0) and -π to 0 for diamagnetic tissue
(∆χ < 0). The change in the signal in both cases may be
explained by the magnetic properties of the iron and calcium atoms. The authors
highlight that the above descriptions consider the "left-hand" reference system, which
defines the 0 intensity on the phase image as π, and the intensity 4095 as
-π. On the other hand, the "right-hand" system defines such an interval as the
contrary (-π to π). Each manufacturer adopts a reference system. In order
to find out the reference system one has only to observe the phase values of the blood
vessels and basal ganglia, which are paramagnetic.Although SWI is already a recognized technique, computed tomography (CT) is still
considered to be a conclusive technique in the differential diagnosis of cerebral
calcification(. Besides the
already known economic factors related to CT and MRI, the authors consider that a more
extensive utilization of SWI as a conclusive technique for the mentioned diagnosis is
affected by two extremely important choices in the correct use of this technique,
namely, the TE value in the images acquisition and the type of mask in the
postprocessing phase. Such two choices influence the magnetic susceptibility contrast on
the SWI. It is difficult to interpret the effects of such parameters because of the
phase limitation to values within the interval -π to +π. As the value of
the tissue phase exceeds the π scale for long TE, the image phase assumes the
added value of -2π. Such effect is called wrapping(. The incorrect selection of TE may lead to erroneous
definition of the magnetic susceptibility contrast on the SWI.The present study describes in detail the SWI processing, highlighting the effect of the
selection of the TE and of the sensitive mask to differentiate between calcification and
hemosiderin, simultaneously in six patients with different diagnostic hypotheses.
Additionally, the authors evaluated the effectiveness of this procedure, considering CT
as the gold standard.
MATERIALS AND METHODS
Subjects
Six patients (four men and two women; age range between 41 and 54 years) were
selected to illustrate the application of SWI as well the technique validity. The
retrospective selection was made after the images acquisition, according to the
following criteria: a) diagnostic suspicion of neurotoxoplasmosis, ventricular cysts
and hemosiderin deposition; b) neurological CT and MRI studies at an interval of at
least 3 months (except for one patient with suspected neurotoxoplasmosis, with an
interval of 10 months between examinations); c) absence of artifacts at both imaging
modalities; d) reports issued and reviewed by at least two neuroradiologists, and
confirmation of the diagnostic suspicion.
Imaging parameters
MRI scans were performed in the axial plane with a 3DGE (PRESTO) sequence, in a 3.0T
magnetic field (Achieva; Philips), with 8-channel SENSE head coil. Phase and
magnitude data were stored. The sequence parameters were the following: TE, 23.1 ms;
repetition time (TR), 16.3 ms; flip angle, 10°; spatial resolution, 0.57 ×
0.57 × 0.85 mm3; field of view (FOV), 218 × 218 × 127
mm3; SENSE, 2.The CT images were acquired in a Brilliance Big Bore (Philips) apparatus in the axial
plane, with the following parameters: 257 mAs/120 kV; spatial resolution, 0.39
× 0.39 × 5.0 mm3; FOV, 199 × 199 × 150
mm3.
SWI processing
Generally, SWIs are formed by multiplying the phase mask to the magnitude image
(Figure 1). The generation of the phase mask
is generated in two steps: Low frequency elimination using a high-pass filter and
mask generation considering the desired phase interval. A detailed explanation about
this process is below.
Figure 1
Diagram describing SWI processing with sigmoid mask. Patient with calcification
deposit.
Diagram describing SWI processing with sigmoid mask. Patient with calcification
deposit.The filter is applied to the phase image to eliminate or attenuate the low spatial
frequency components formed by the inhomogeneity of the background magnetic field
(Figure 1)(. Such components are responsible for smoother spatial
variations on the images, thus they do not represent local variables, rather global,
such as total contrast and average pixel intensity. The background magnetic field is
generated by outside sources, such as, inhomogeneity of the static magnetic field
(B0) and non-uniformity of the radiofrequency field.
Several types of high-pass filters can be used, here we applied a simple version as
suggested by Haacke et al.(. In
this case, the filtered phase image is calculated from the original complex image in
the 2D k-space, truncating it in the central n x n points, adding zeros in the
elements outside the truncated region and then making the subtraction of the
resultant truncated phase image and the original phase image, in the real space. The
window size (n × n) will depend on the relation between signal-noise ratio
loss on the phase image and the background field elimination, a convenient value is
64 × 64 considering brain images with a FOV close to 256 mm.The phase mask generation is the most important step in SWI processing. The
paramagnetic or diamagnetic contrast weighting is defined in this step. The phase
image may assume only values of -π to +π by the Fourier transform
properties and limitations of the system of MRI signal acquisition. Thus,
discontinuity values of up to 2π may be observed on the phase image, in case
the value exceeds such limits. Such effect is more common for long echo times, as
shown on Figure 2. Such figure presents phase
images with TE = 23 ms (Figure 2A) and TE = 40
ms (Figure 2B), the latter obtained by means of
simulation of phase evolution from the first one. According to the equation 2 for
values of susceptibility difference between hemosiderin and white matter (∆χ =
0.25 ppm)(, one can observe in
the chart on Figure 2, that the phase
difference for TE = 23 ms is approximately -π/2, and for TE = 40 ms this value
should be +π. Wrapping is also observed for the susceptibility difference
between calcifications and white matter (∆χ = -0.15 ppm)(, shown on the chart: for TE = 23 ms,
the phase is π/2 and for TE = 40 ms the phase is -2π/3. Thus, the phase
evolved with the TE and according to the tissue susceptibility difference.
Figure 2
Wrapping effect on phase image of a patient with hemosiderin deposit.
A and B phase images with TE = 23 ms and 40 ms,
respectively. The chart includes phase values simulated with the “left hand”
system for regions with hemosiderin and calcification deposits, considering a
3T field. The highlighted spots on the chart indicate phase values for TE = 23
ms and TE = 40 ms.
Wrapping effect on phase image of a patient with hemosiderin deposit.
A and B phase images with TE = 23 ms and 40 ms,
respectively. The chart includes phase values simulated with the “left hand”
system for regions with hemosiderin and calcification deposits, considering a
3T field. The highlighted spots on the chart indicate phase values for TE = 23
ms and TE = 40 ms.The wrapping suggests undesirable TE values for the differentiation of hemosiderin
deposition and calcification lesions, depending on the magnetic field strengths. For
3T, as exemplified on Figure 2, with echo times
approaching 13 and 27 ms, the phase difference in the central region of both types of
lesions tend to a single value, making a correct distinction more difficult.
Additionally, depending on the interval where the selected TE is situated,
intensities will alternate between both types of lesions.The phase mask is based on the linear relation between the phase evolution and the
magnetic susceptibility (equation 1), valid for all the points of the image acquired
in a determined magnetic field (B0) and with a single TE.
Such mask is applied to enhance the differences on the phase images on the basis of
magnitude images with a more anatomical aspect of easier interpretation.Different masks may be defined, always with values between 0 and 1 (Figure 3). A common example that is easy to be
illustrated is the negative mask, assuming a unitary value for phases > 0 and from
0 to 1 for phase values in the interval [-π, 0]. Another example
is the positive mask, assuming 1 for phase values < 0 and 0 to 1 for the interval
[0, π]. The previous masks may be individually utilized to
enhance a specific type of lesion. On the other hand, the sigmoid mask( is more general and allows for
differentiation between paramagnetic and diamagnetic tissues on a single
susceptibility-weighted image. Such mask follows the behavior of the sigmoid function
described on equation 3:
Figure 3
Phase masks defined with values between 0 and 1 according to the phase
values.
Phase masks defined with values between 0 and 1 according to the phase
values.The adjustment parameters "a" and "b" are freely selected as to achieve a good
contrast-noise ratio(. In this
paper, a sigmoid mask (parameters a = 0.5 and b = -0.15) was used to differentiate
paramagnetic and diamagnetic tissues.Subsequently, the phase mask is multiplied n times over the
magnitude image to generate the susceptibility-weighted image. The number
n of multiplications should be selected to optimize the
contrast-noise ratio at SWI, as demonstrated by Reichenbach et al., such number
should be close to 4(. Figure 4 exemplifies the change of the
contrast-noise for the use of sigmoid mask multiplied 4, 6 and 8 times over the
magnitude image of a patient with hemosiderin deposition.
Figure 4
Susceptibility-weighted images of a patient with hemosiderin deposit (indicated
with the dashed line), with different sigmoid mask multiplications
A:
n = 4; B:
n = 6; C:
n = 8. The chart includes signal intensity values for SWI
along the profile indicated by the dotted line on the magnified image.
Susceptibility-weighted images of a patient with hemosiderin deposit (indicated
with the dashed line), with different sigmoid mask multiplications
A:
n = 4; B:
n = 6; C:
n = 8. The chart includes signal intensity values for SWI
along the profile indicated by the dotted line on the magnified image.In the present study, the SWI processing was performed in a home-made Matlab routine.
In summary, a high-pass (HP) filter with window size 64 × 64( was selected and a sigmoid mask
× 4 was applied(.
Effect of the spatial distribution of the phase in regions adjacent to the
lesion
In addition to the punctual dependence between the phase difference and the tissue
susceptibility difference, as approached by equation 1, the
∆φ(r,TE) is also affected in the adjacent areas, a fact
that is frequently observed at the sections inferiorly and superiorly to the
hemosiderin deposit or calcification (Figure
5). A more general way to write the equation 2 is to express
∆φ(r,TE) as a function linearly dependent of the
perturbation in the main magnetic field ∆B(r) along the
B0 vector; the other components are not considered:
Figure 5
SWI filtered phase images (64 × 64) of a patient with calcification and
other with CT-proved hemosiderin deposit. These images demonstrate the
non-local of the phase evolution the on the section inferiorly and superiorly
to the hemosiderin deposit and with opposite behavior for calcification and
hemosiderin.
SWI filtered phase images (64 × 64) of a patient with calcification and
other with CT-proved hemosiderin deposit. These images demonstrate the
non-local of the phase evolution the on the section inferiorly and superiorly
to the hemosiderin deposit and with opposite behavior for calcification and
hemosiderin.As calcification and hemosiderin deposits with spherical geometry centered at the
origin are considered, the perturbation of the magnetic field at an arbitrary point
resulting from such deposits might be written as follows(:where: a indicates the radius of the spherical lesion, and Θ is the
angle between the position vector (r) and the main magnetic field
axis (z axis).Using the equations 4, 5a and 5b, the authors simulated the effect of the phase
evolution on the sections inferiorly and superiorly to the calcification and
hemosiderin deposition with a 10 mm radius, and considering a 3T magnetic field
intensity and TE = 23 ms. In such simulation, ∆χ = 0.25 ppm for hemosiderin,
and ∆χ = -0.20 ppm for calcification were assumed.
RESULTS
Only four (1, 2, 3 and 4) out of the six patients presented calcifications at the CT
images. For such patients, the susceptibility-weighted images presented hyperintense
signal on the calcification regions (Figure 6,
continuous arrows). The other two patients (5 and 6) did not present any detectable
calcifications at the CT images and presented hemosiderin deposits with hypointensity at
SWI(Figure 6, dotted arrows).
Figure 6
Magnitude, mask (multiplied 4 times), SWI and CT images on each one of the columns
for all the individuals involved in the present study. The SWI processed with
sigmoid mask and the respective CT image demonstrate the regions of calcification
(continuous arrow) or hemosiderin (dotted arrow), depending on the case.
Magnitude, mask (multiplied 4 times), SWI and CT images on each one of the columns
for all the individuals involved in the present study. The SWI processed with
sigmoid mask and the respective CT image demonstrate the regions of calcification
(continuous arrow) or hemosiderin (dotted arrow), depending on the case.The HP filter failed to filter some wrapping artifacts on the SWI of the patients 2 and
4. Generally, such artifacts occurred in regions of great magnetic susceptibility
difference, as the region of the nasal sinuses and ears.Figure 7 shows the phase evolution effect
inferiorly and superiorly to the radius of the simulated calcification and hemosiderin
deposition. Regions outside the lesion show phase values induced by the adjacent
lesion.
Figure 7
Image of calcification and hemosiderin deposit with spherical geometry to simulate
the phase evolution effect on the sections inferiorly and superiorly to the sphere
with a 10 mm-radius. 3T and TE = 23 ms were considered.
Image of calcification and hemosiderin deposit with spherical geometry to simulate
the phase evolution effect on the sections inferiorly and superiorly to the sphere
with a 10 mm-radius. 3T and TE = 23 ms were considered.
DISCUSSION
SWI has been suggested to identify cerebral hemosiderin and calcification
deposition(. The proposed
analysis allowed for the differentiation of such deposits by means of SWI using a
sigmoid mask, and the results were validated by CT image.Hemosiderin deposits include Fe+3 atoms, which are paramagnetic and distort
the local magnetic Field at molecular level(. The magnetic field distortion causes a different interaction
between each nuclear spin of the tissue and the local magnetic field. Therefore, the
phase resulting from the voxel with hemosiderin deposit evolves much more than the
normal tissue in the adjacent voxel. Calcifications contain Ca+2 diamagnetic
atoms which also distort the magnetic field leading to spins phase evolution. The phase
evolution caused by the calcification occurs contrarily to Fe+3 deposits.
Therefore, for a determined TE, it is possible to differentiate hemosiderin deposit and
calcification (Figure 2).For TE = 23 ms at 3 T the calcifications demonstrated hyperintense signal at SWI with
sigmoid mask, in agreement with reports on calcifications in oligodendrogliomas,
physiological lesions and cysticercosis(, and in prostate cancer(. SWI with longer TEs may yield results opposite to the above
mentioned results. Recently, Zulfiqar et al. utilized TE = 40 ms and reported
hypointense signal for calcifications in oligodendrogliomas at SWI with negative
mask(. Additionally, they
concluded that, amongst the techniques for diagnosis of cerebral calcifications by MRI,
SWI was the most frequently indicated modality to detect calcifications in
oligodendrogliomas with results confirmed by CT image(. Other authors utilized TE = 25 ms and also reported
hypointense signal for calcifications at SWI with negative mask(, however, the static magnetic field was
less intense (1.5 T). As shown on the equation 2, the phase evolution was weaker for
smaller magnetic fields. Additionally, it is important to highlight that phase images
may assume opposite values as the "right hand" system is utilized. Thus, the authors
point out that the results reported in the literature on this matter should be
cautiously considered with a view on their application to each clinical center
reality.Also, the most MRI scanner manufacturers include the SWI sequence acquisition with
automated postprocessing in their products, so the utilization of this imaging modality
is increasing in the clinical practice. However, the authors highlight that the SWI is
postprocessed and depends on the selection of a type of mask and some parameters which
many times may be occult to radiologists. Clinical diagnoses might be erroneously
produced in the absence of knowledge about variable and/or inappropriately selected
parameters.The authors suggest a standardization of the use of SWI by a correct TE selection for
the construction of a phase mask capable of enhance the contrast of both hemosiderin and
calcification in order to avoid possible errors in clinical diagnoses. For example: For
the acquisition, a GE sequence using TE = 23 ms for 3 T magnetic field and storing the
phase and magnitude data with a reference "left-hand" system; for the SWI processing, a
HP filter with a window size of 64 × 64 and using sigmoid mask with four
multiplications.The presence of magnetic susceptibility artifacts in the region of the nasal sinuses and
ears of patients 2 and 4 (Figure 6) did not impair
the diagnosis and differentiation between hemosiderin deposition and calcifications. In
some cases, however, such artifacts prevent a complete visualization of the brain. The
use of shorter TE, filter with larger window size, and other filters might remove such
artifacts(.
CONCLUSION
SWI is a technique that produces postprocessed images weighting the magnetic
susceptibility contrast of the phase image on the magnitude image. The TE and mask
selection may change all the SWI information, affecting the diagnostic reliability.
Amongst the possible masks, the authors highlight that the use of the sigmoid mask
allows for enhancing the calcification and hemosiderin contrast on a single SWI.
Authors: Lisa C Adams; Keno Bressem; Sarah Maria Böker; Yi-Na Yvonne Bender; Dominik Nörenberg; Bernd Hamm; Marcus R Makowski Journal: Sci Rep Date: 2017-11-14 Impact factor: 4.379
Authors: Lisa C Adams; Sarah M Böker; Yvonne Y Bender; Gerd Diederichs; Eva M Fallenberg; Moritz Wagner; Bernd Hamm; Marcus R Makowski Journal: PLoS One Date: 2017-03-09 Impact factor: 3.240