| Literature DB >> 25987408 |
Elyar Ghafoori1, Muammar M Kabir1, Jian Cao2, Alexei Shvilkin3, Larisa G Tereshchenko4.
Abstract
We constructed an intracardiac vectorcardiogram from 3 configurations of intracardiac cardiovertor defibrilator (ICD) electrograms (EGMs). Six distinctive 3 lead combinations were selected out of five leads: can to right ventricular coil (RVC); RVC to superior vena cava coil (SVC); atrial lead tip (A-tip) to right ventricular (RV)-ring; can to RV-ring; RV-tip to RVC, in a patient with dual chamber ICD. Surface spatial QRS-T angle (119.8°) was similar to intracardiac spatial QRS-T angle derived from ICD EGMs combination A (101.3°), B (96.1°), C (92.8°), D (95.2), E (99.0), F (96.2) and median (101.5). Future validation of the novel method is needed.Entities:
Keywords: Implantable cardioverter defibrillator; Intracardiac electrogram; Vectorcardiography
Mesh:
Year: 2015 PMID: 25987408 PMCID: PMC4484311 DOI: 10.1016/j.jelectrocard.2015.05.008
Source DB: PubMed Journal: J Electrocardiol ISSN: 0022-0736 Impact factor: 1.438