Geometric Calibration and Image Reconstruction for a Segmented Slant-Hole Stationary Cardiac SPECT System.

Yanfei Mao1, Zhicong Yu2, Gengsheng L Zeng3.   

Abstract

UNLABELLED: A dedicated stationary cardiac single-photon emission computed tomography (SPECT) system with a novel segmented slant-hole collimator has been developed. The goal of this paper is to calibrate this new imaging geometry with a point source.
METHODS: Unlike the commercially available dedicated cardiac SPECT systems, which are specialized and can be used only to image the heart, our proposed cardiac system is based on a conventional SPECT system but with a segmented slant-hole collimator replacing the collimator. For a dual-head SPECT system, 2 segmented collimators, each with 7 sections, are arranged in an L-shaped configuration such that they can produce a complete cardiac SPECT image with only one gantry position. A calibration method was developed to estimate the geometric parameters of each collimator section as well as the detector rotation radius, under the assumption that the point source location is calculated using the central-section data. With a point source located off the rotation axis, geometric parameters for each collimator section can be estimated independently. The parameters estimated individually are further improved by a joint objective function that uses all collimator sections simultaneously and incorporates the collimator symmetry information.
RESULTS: Estimation results and images reconstructed from estimated parameters are presented for both simulated and real data acquired from a prototype collimator. The calibration accuracy was validated by computer simulations with an error of about 0.1° for the slant angles and about 1 mm for the rotation radius. Reconstructions of a heart-insert phantom did not show any image artifacts of inaccurate geometric parameters.
CONCLUSION: Compared with the detector's intrinsic resolution, the estimation error is small and can be ignored. Therefore, the accuracy of the calibration is sufficient for cardiac SPECT imaging.
© 2015 by the Society of Nuclear Medicine and Molecular Imaging, Inc.

Entities:  

Keywords:  cardiac single photon emission computed tomography (SPECT); collimator; geometric calibration; slant-hole

Mesh:

Year:  2015        PMID: 25956691      PMCID: PMC5297456          DOI: 10.2967/jnmt.114.153668

Source DB:  PubMed          Journal:  J Nucl Med Technol        ISSN: 0091-4916


  9 in total

1.  Characterization of pinhole SPECT acquisition geometry.

Authors:  Dirk Bequé; Johan Nuyts; Guy Bormans; Paul Suetens; Patrick Dupont
Journal:  IEEE Trans Med Imaging       Date:  2003-05       Impact factor: 10.048

2.  Optimization of geometrical calibration in pinhole SPECT.

Authors:  Dirk Bequé; Johan Nuyts; Paul Suetens; Guy Bormans
Journal:  IEEE Trans Med Imaging       Date:  2005-02       Impact factor: 10.048

3.  GATE: a simulation toolkit for PET and SPECT.

Authors:  S Jan; G Santin; D Strul; S Staelens; K Assié; D Autret; S Avner; R Barbier; M Bardiès; P M Bloomfield; D Brasse; V Breton; P Bruyndonckx; I Buvat; A F Chatziioannou; Y Choi; Y H Chung; C Comtat; D Donnarieix; L Ferrer; S J Glick; C J Groiselle; D Guez; P F Honore; S Kerhoas-Cavata; A S Kirov; V Kohli; M Koole; M Krieguer; D J van der Laan; F Lamare; G Largeron; C Lartizien; D Lazaro; M C Maas; L Maigne; F Mayet; F Melot; C Merheb; E Pennacchio; J Perez; U Pietrzyk; F R Rannou; M Rey; D R Schaart; C R Schmidtlein; L Simon; T Y Song; J M Vieira; D Visvikis; R Van de Walle; E Wieërs; C Morel
Journal:  Phys Med Biol       Date:  2004-10-07       Impact factor: 3.609

4.  A novel approach to multipinhole SPECT for myocardial perfusion imaging.

Authors:  Tobias Funk; Dennis L Kirch; John E Koss; Elias Botvinick; Bruce H Hasegawa
Journal:  J Nucl Med       Date:  2006-04       Impact factor: 10.057

5.  Performance evaluation of D-SPECT: a novel SPECT system for nuclear cardiology.

Authors:  Kjell Erlandsson; Krzysztof Kacperski; Dean van Gramberg; Brian F Hutton
Journal:  Phys Med Biol       Date:  2009-04-08       Impact factor: 3.609

6.  Comparison of simultaneous dual-isotope multipinhole SPECT with rotational SPECT in a group of patients with coronary artery disease.

Authors:  Peter P Steele; Dennis L Kirch; John E Koss
Journal:  J Nucl Med       Date:  2008-06-13       Impact factor: 10.057

7.  Estimation of geometrical parameters and collimator evaluation for cone beam tomography.

Authors:  G T Gullberg; B M Tsui; C R Crawford; J G Ballard; J T Hagius
Journal:  Med Phys       Date:  1990 Mar-Apr       Impact factor: 4.071

8.  Measurement of collimator hole angulation and camera head tilt for slant and parallel hole collimators used in SPECT.

Authors:  E Busemann-Sokole
Journal:  J Nucl Med       Date:  1987-10       Impact factor: 10.057

9.  A fast cardiac gamma camera with dynamic SPECT capabilities: design, system validation and future potential.

Authors:  Moshe Bocher; Ira M Blevis; Leonid Tsukerman; Yigal Shrem; Gil Kovalski; Lana Volokh
Journal:  Eur J Nucl Med Mol Imaging       Date:  2010-06-29       Impact factor: 9.236

  9 in total
  1 in total

1.  Multimodality calibration for simultaneous fluoroscopic and nuclear imaging.

Authors:  Casper Beijst; Mattijs Elschot; Sandra van der Velden; Hugo W A M de Jong
Journal:  EJNMMI Phys       Date:  2016-08-30
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.