Measurement of collimator hole angulation and camera head tilt for slant and parallel hole collimators used in SPECT.

E Busemann-Sokole1.   

Abstract

Collimator hole angulation was measured at 16 locations in three collimators used in SPECT. The parallel hole collimator measured -0.1 +/- -0.1 s.d. degrees in the X and -0.2 +/- 0.1 s.d. degrees in the Y direction. One 30 degree nominal slant hole collimator revealed a 0.1 +/- 0.3 s.d. degrees angulation in X and 26.5 +/- 1.1 s.d. degrees angulation in Y, which was unacceptable. The replacement collimator measured 0.0 +/- 0.1 s.d. degrees in X and 29.6 +/- 0.2 s.d. degrees in Y. Determination of collimator hole angulation is recommended as an acceptance test for SPECT systems. The net angle of tilt with respect to the orthogonal resulting from a particular collimator and camera head tilt was determined from summed projection images over 360 degrees of a point source placed off the axis of rotation. These measurements were sensitive within a 0.5 degree tilt angle. The method is suggested as a routine quality control procedure to optimize camera head tilt to a particular collimator. It may also reveal unexpected mechanical misalignments in the camera gantry.

Entities:  

Mesh:

Year:  1987        PMID: 3498806

Source DB:  PubMed          Journal:  J Nucl Med        ISSN: 0161-5505            Impact factor:   10.057


  5 in total

1.  EANM/ESC procedural guidelines for myocardial perfusion imaging in nuclear cardiology.

Authors:  B Hesse; K Tägil; A Cuocolo; C Anagnostopoulos; M Bardiés; J Bax; F Bengel; E Busemann Sokole; G Davies; M Dondi; L Edenbrandt; P Franken; A Kjaer; J Knuuti; M Lassmann; M Ljungberg; C Marcassa; P Y Marie; F McKiddie; M O'Connor; E Prvulovich; R Underwood; B van Eck-Smit
Journal:  Eur J Nucl Med Mol Imaging       Date:  2005-07       Impact factor: 9.236

2.  Acceptance testing for nuclear medicine instrumentation.

Authors:  Ellinor Busemann Sokole; Anna Płachcínska; Alan Britten
Journal:  Eur J Nucl Med Mol Imaging       Date:  2010-03       Impact factor: 9.236

3.  Routine quality control recommendations for nuclear medicine instrumentation.

Authors:  Ellinor Busemann Sokole; Anna Płachcínska; Alan Britten; Maria Lyra Georgosopoulou; Wendy Tindale; Rigobert Klett
Journal:  Eur J Nucl Med Mol Imaging       Date:  2010-03       Impact factor: 9.236

Review 4.  Multidetector single-photon emission tomography: are two (or three or four) heads really better than one?

Authors:  J M Links
Journal:  Eur J Nucl Med       Date:  1993-05

5.  Geometric Calibration and Image Reconstruction for a Segmented Slant-Hole Stationary Cardiac SPECT System.

Authors:  Yanfei Mao; Zhicong Yu; Gengsheng L Zeng
Journal:  J Nucl Med Technol       Date:  2015-05-08
  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.