Characteristics of an 18 MV photon beam from a Therac 20 Medical Linear Accelerator.

M S Patterson, P C Shragge.   

Abstract

The 18 MV photon beam characteristics of a Therac 20 Medical Linear Accelerator manufactured by Atomic Energy of Canada Ltd, are presented. Tissue phantom ratios (TRP's) and percent depth dose data are given; for a 10 x 10 cm field, the percent depth dose at a depth of 10 cm is 78.5 (SSD 100 cm). The relative dose factors (RDF'S) are given and are analyzed to elucidate the relative contributions from phantom scatter, collimator scatter, and backscatter from the top of the collimators into the monitor chambers. The effect of field size and depth on the penumbra is described. Crossplots of the beam at a depth of 5 cm indicate that the flattening filter could be improved; there are hot spots of 108% near the corners of 40 x 40 fields.

Entities:  

Mesh:

Year:  1981        PMID: 6798391     DOI: 10.1118/1.594833

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  4 in total

1.  Measurement and comparison of head scatter factor for 7 MV unflattened (FFF) and 6 MV flattened photon beam using indigenously designed columnar mini phantom.

Authors:  Sigamani Ashokkumar; Arunai Nambiraj; Sujit Nath Sinha; Girigesh Yadav; Kothanda Raman; Manindra Bhushan; Rajesh Thiyagarajan
Journal:  Rep Pract Oncol Radiother       Date:  2015-03-01

2.  Prediction of back-scatter radiations to a beam monitor chamber of medical linear accelerators by use of the digitized target-current-pulse analysis method.

Authors:  Yusuke Suzuki; Naoki Hayashi; Hideki Kato; Hiroshi Fukuma; Yasujiro Hirose; Makoto Kawano; Yoshio Nishii; Masaru Nakamura; Takashi Mukouyama
Journal:  Radiol Phys Technol       Date:  2012-10-25

3.  Measurement of back-scattered radiation from micro multileaf collimator into the beam monitor chamber from a dual energy linear accelerator.

Authors:  K R Muralidhar; P Narayana Murthy; N V N M Sresty; Pramod Kumar Dixit; Rajneesh Kumar; A K Raju
Journal:  J Med Phys       Date:  2007-04

4.  Comparison of Head Scatter Factor for 6MV and 10MV flattened (FB) and Unflattened (FFF) Photon Beam using indigenously Designed Columnar Mini Phantom.

Authors:  Sigamani Ashokkumar; N Arunai Nambi Raj; Sujit Nath Sinha; Girigesh Yadav; Rajesh Thiyagarajan; Kothanda Raman; Manindra Bhushan Mishra
Journal:  J Med Phys       Date:  2014-07
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.