Pulmonary vascular permeability index and global end-diastolic volume: are the data consistent in patients with femoral venous access for transpulmonary thermodilution: a prospective observational study.

Helena Berbara1, Sebastian Mair2, Analena Beitz3, Benedikt Henschel4, Roland M Schmid5, Wolfgang Huber6.   

Abstract

BACKGROUND: Transpulmonary thermodilution (TPTD) derived parameters are used to direct fluid management in ICU-patients. Extravascular lung water EVLW and its ratio to pulmonary blood volume (pulmonary vascular permeability index PVPI) have been associated with mortality. In single indicator TPTD pulmonary blood volume (PBV) is estimated to be 25% of global end-diastolic volume (GEDV). A recent study demonstrated marked overestimation of GEDV indexed to body-surface area (BSA; GEDVI) when using a femoral central venous catheter (CVC) for indicator injection due to the additional volume measured in the vena cava inferior. Therefore, a correction formula derived from femoral TPTD and biometric data has been suggested. Consequence, one of the commercially available TPTD-devices (PiCCO; Pulsion Medical Systems, Germany) requires information about CVC site. Correction of GEDVI for femoral CVC can be assumed. However, there is no data if correction also pertains to unindexed GEDV, which is used for calculation of PBV and PVPI. Therefore, we investigated, if also GEDV, PBV and PVPI are corrected by the new PiCCO-algorithm.
METHODS: In this prospective study 110 triplicate TPTDs were performed within 30 hours in 11 adult ICU-patients with PiCCO-monitoring and femoral CVC. We analyzed if the femoral TPTD correction formula for GEDVI was also applied to correct GEDV. Furthermore, we compared PVPIdisplayed to PVPIcalculated which was calculated as EVLWdisplayed/(0.25*GEDVdisplayed).
RESULTS: Multiplication of GEDVIdisplayed by BSA resulted in GEDVcalculated which was not significantly different to GEDVdisplayed (1459 ± 365 mL vs. 1459 ± 366 mL) suggesting that correction for femoral indicator injection also pertains to GEDVdisplayed. However, PVPIdisplayed was significantly lower than PVPIcalculated (1.64 ± 0.57 vs. 2.27 ± 0.72; p < 0.001). In addition to a bias of -0.64 ± 0.22 there was a percentage error of 22%. Application of the correction formula suggested for GEDVI to PVPIdisplayed reduced the bias of PVPIdisplayed compared to EVLW/PBV from -0.64 ± 0.22 to -0.10 ± 0.05 and the percentage error from 22% to 4%.
CONCLUSIONS: Correction for femoral CVC in the PiCCO-device pertains to both GEDVIdisplayed and GEDVdisplayed, but not to PVPIdisplayed. To provide consistent information, PVPI should be calculated based on GEDVcorrected in case of femoral CVC.

Entities:  

Mesh:

Year:  2014        PMID: 25928560      PMCID: PMC4448283          DOI: 10.1186/1471-2253-14-81

Source DB:  PubMed          Journal:  BMC Anesthesiol        ISSN: 1471-2253            Impact factor:   2.217


Background

Appropriate fluid supply is a cornerstone of critical care. In addition to cardiac output CO, transpulmonary thermodilution (TPTD) provides extravascular lung water (EVLW) reflecting pulmonary edema and the preload parameter global end-diastolic volume (GEDV) [1-6]. Increased EVLW is associated to mortality [3,7-11]. GEDV and its changes are associated to CO [6] and algorithms based on GEDV have been shown to improve outcome [12]. Moreover, the relation of EVLW and pulmonary blood volume (PBV) might be useful to differentiate the etiology of pulmonary edema [13,14]. PBV can be determined using double-indicator TPTD technique [13]. Single indicator TPTD is much more common in clinical routine and estimates PBV as about 25% of GEDV [14]. EVLW/PBV-ratio has been termed pulmonary vascular permeability index “PVPI”. Several studies demonstrated that higher values of PVPI are associated to pulmonary edema with increased permeability of the alveolo-capillary barrier [14]. Increased PVPI has been found in patients with ARDS resulting from pulmonary (e.g. pneumonia) or secondary (e.g. sepsis) pulmonary impairment. By contrast hydrostatic pulmonary edema due to congestive heart failure usually does not result in marked increases of PVPI since EVLW and PBV are increased to a similar extent [13,14]. Although TPTD has become part of clinical routine, several pitfalls have to be kept in mind. Two recent studies demonstrated marked overestimation of GEDV in case of performing TPTD indicator injection using a femoral venous access due to the additional volume of vena cava inferior (VCI) participating in TPTD [15,16]; Figure 1).
Figure 1

Comparison of transpulmonary indicator dilution (TPTD) using femoral central venous catheter (CVC) instead of jugular or subclavian CVC. Obviously, the distribution volume for the indicator is augmented by the volume of vena cava inferior in case of femoral indicator injection. This results in diminished indicator concentration and overestimation of global enddiastolic volume GEDV. In single indicator dilution technique GEDV is calculated as difference of the total dilution volume (intrathoracic thermovolume ITTV) minus the sum of pulmonary blood volume PBV and extravascular lung water EVLW: GEDV = ITTV- (EVLW + PBV). The sum (PBV + EVLW) is thermed pulmonary thermovolume PTV, which is directly derived from the thermodilution curve. Therefore, it can be assumed not to be altered by the additional dilution volume of VCI participating in TPTD.

Comparison of transpulmonary indicator dilution (TPTD) using femoral central venous catheter (CVC) instead of jugular or subclavian CVC. Obviously, the distribution volume for the indicator is augmented by the volume of vena cava inferior in case of femoral indicator injection. This results in diminished indicator concentration and overestimation of global enddiastolic volume GEDV. In single indicator dilution technique GEDV is calculated as difference of the total dilution volume (intrathoracic thermovolume ITTV) minus the sum of pulmonary blood volume PBV and extravascular lung water EVLW: GEDV = ITTV- (EVLW + PBV). The sum (PBV + EVLW) is thermed pulmonary thermovolume PTV, which is directly derived from the thermodilution curve. Therefore, it can be assumed not to be altered by the additional dilution volume of VCI participating in TPTD. The authors suggested a correction formula for GEDV indexed to predicted body surface area (GEDVI) based on data derived from femoral TPTD and biometric information. As a consequence of these data one the manufacturer of the PiCCO device (Pulsion Medical Systems, Feldkirchen, Germany) implemented a new software requiring the information about CVC site (femoral or jugular/subclavian) and correcting GEDVI in case of femoral injection. Although jugular and subclavian vein access might be the preferred sites for CVC insertion, several recent studies on catheter related bloodstream infections have demonstrated that femoral venous access was used in about 20 to 35% of all catheter insertions [17,18]. Therefore, results of TPTD might be substantially altered in about one third of cases without appropriate correction. To the best of our knowledge there are no data available if the new PiCCO algorithm also corrects unindexed GEDV and PVPI in case of femoral CVC. Furthermore, data on PVPI were not investigated in both previous papers on TPTD using a femoral CVC [15,16]. Therefore, it was the aim of our study to investigate if in case of femoral indicator injection GEDV in the new PiCCO algorithm is given based on the correction of GEDVI and PVPI is based on corrected GEDV and corrected pulmonary blood volume.

Methods

The study was approved by the institutional review board (Ethikkommission der Fakultät für Medizin der Technischen Universität München, Munich, Germany). All patients or their legal representatives gave written informed consent. We prospectively performed 110 TPTDs in 11 patients with femoral indicator injection. All patients were recruited between October 2013 and January 2014. Ten triplicate TPTDs per patient were performed with 15 mL cold saline within a total of 30 h. TPTD was performed as described previously using the PiCCO-2 device (Pulsion Medical Systems, Feldkirchen, Germany) [6,16,19] and a 5-lumen CVC (Multicath 5; Vygon; Aachen, Germany). All PiCCO-2 devices were equipped with the V3.1 algorithm requiring information about the venous catheter site which was set to “femoral” CVC in all patients. Immediately after triplicate TPTD absolute and indexed mean values were documented for GEDV, GEDVI, EVLW, EVLWI and PVPI as displayed by the device. Parameters displayed by the device were subscripted with “displayed” (see Table 1).
Table 1

Terminology of thermodilution-derived parameters

SubscriptMeaningParameters analyzed
“displayed”Values as displayed by the PiCCO monitorGEDVIdisplayed
GEDVdisplayed
EVLWdisplayed
PVPIdisplayed
“calculated”Parameters calculated based on parameters displayed on the monitor.GEDVcalculated
These parameters should be consistent with corresponding displayed values.PVPIcalculated
“corrected”Parameters derived from correction of the displayed values. Corrected parameters were used for comparison to calculated values (only used in case of inconsistency of displayed and calculated values).PVPIcorrected
GEDVuncorrected
Terminology of thermodilution-derived parameters The main purpose of the study was to investigate the consistency of displayed values by comparing displayed data (e.g. GEDVdisplayed, PVPIdisplayed) to the corresponding values derived from calculation specified with the subscript “calculated” (GEDVcaculated, PVPIcalculated). If displayed and calculated values were inconsistent, we investigated whether the inconsistency could be explained by (non-) application of the recently suggested correction formula for the displayed values parameters [16]. In this case “ex post” correction by our formula should result in a consistency of the “corrected” values (with the subscript “corrected”) and the calculated values. To analyze if the femoral TPTD correction formula for GEDVIdisplayed was only applied for GEDVIdisplayed or if also GEDVdisplayed is corrected in case of femoral indicator injection, we calculated GEDVcalculated using the formula: Finally, GEDVdisplayed was compared to GEDVcalculated. In a second step we analyzed, if PVPIdisplayed is calculated based on a GEDVdisplayed corrected for femoral indicator injection. Therefore, we compared PVPIdisplayed to PVPIcalculated using the formula PVPIcalculated = EVLWdisplayed/(0.25 * GEDVdisplayed). Predicted bodyweight BWpredicted was calculated using the formula: Predicted body surface area BSApredicted was calculated using BWpredicted instead of actual bodyweight in the Dubois formula: The correction formula suggested for correction of uncorrected femoral indicator injection derived GEDVuncorrected is GEDVIcorrected [mL / m2] = 0.539 * GEDVIuncorrected - 15.17 + 24.49 * CIuncorrected +2.311* BWideal[16]. Ideal bodyweight BWideal was calculated using the formula: Correlation of GEDVcalculated and GEDVIdisplayed was analyzed using Spearman correlation. Comparisons of GEDVdisplayed vs. GEDVcalculated, PVPIdisplayed vs. PVPIcalculated and PVPIcorrected vs. PVPIcalculated were performed using Wilcoxon-test for paired samples, respectively. Bland-Altman analyses and calculation of percentage error (PE) were used for these comparisons as described previously [20,21]. Bland-Altman analyses were corrected for repeated measurements allowing variability of true values within each subject [22]. All analyses were performed using IBM SPSS Statistics 21 (SPSS inc., Chicago, IL, USA).

Results

Patients characteristics

Five female and six male patients were included. Patients characteristics are shown in Table 2.
Table 2

Patients characteristics (mean ± standard deviation; numbers and percentages)

Gender5/11 (45%) female; 6/11 (55%)male
Age [years]59.9 ± 11.9
Height [cm]171.1 ± 11.7
APACHE-II score16 ± 6
SOFA score6.5 ± 2.5
Actual bodyweight [kg]72.8 ± 19.1
Predicted bodyweight [kg]65.0 ± 12.0
Ideal bodyweight [kg]62.5 ± 11.4
Predicted body surface area [m2]1.76 ± 0.22
Aetiology 
   Sepsis7/11 (64%)
   ARDS1/11 (9%)
   Cirrhosis1/11 (9%)
   Cardiogenic shock2/11 (18%)
Measurements under mechanical ventilation60/110 (54.5%)
Measurements under vasopressors34/110 (30.9%)
Patients characteristics (mean ± standard deviation; numbers and percentages) GEDVcalculated was obtained by multiplying GEDVIdisplayed by BSApredicted. As demonstrated in Figure 2 GEDVcalculated and GEDVdisplayed significantly correlated (r2 = 1.0; p < 0.001) and were not significantly different (1459 ± 365 mL vs. 1459 ± 366 mL).
Figure 2

Scatter plot depicting the correlation of GEDV as displayed by the PiCCO-2 device (GEDV ) vs. GEDV calculated as product GEDVI * predicted body surface area.

Scatter plot depicting the correlation of GEDV as displayed by the PiCCO-2 device (GEDV ) vs. GEDV calculated as product GEDVI * predicted body surface area. This confirms that correction for femoral venous catheter site pertains to both GEDVIdisplayed and GEDVdisplayed. Consequently, calculation of PVPIcalculated based on EVLWdisplayed and GEDVdisplayed should result in a PVPIcalculated identical with PVPIdisplayed. However, PVPIdisplayed was significantly lower than PVPIcalculated (Figure 3; 1.64 ± 0.57 vs. 2.27 ± 0.72; p < 0.001).In addition to a bias of -0.64 ± 0.22 Bland-Altman analysis demonstrated a percentage error of 22% and upper and lower limits of agreement of -0.199 and -1.075 (Figure 4).
Figure 3

Boxplots plots comparing pulmonary vascular permeability index PVPI as displayed by the PiCCO-2 device (PVPI) vs. PVPI calculated as ratio of EVLW/PBV (extravascular lung water/pulmonary blood volume) vs. PVPI corrected using the formula suggested for correction of femoral indicator injection derived GEDV [16]: GEDVI[mL / m] = 0.539 * GEDVI- 15.17 + 24.49 * CI2.311* BW. Assuming that PVPIdisplayed was calculated based on GEDV and PBV not corrected for femoral injection, PVPIcorrected was calculated by multiplying PVPIdisplayed with the ratio 0.25*GEDVuncorrected/0.25*GEDVcorrected.

Figure 4

Bland Altman plot comparing pulmonary vascular permeability index PVPI as displayed by the PiCCO-2 device (PVPI ) vs. PVPI calculated as ratio of EVLW/PBV (extravascular lung water/pulmonary blood volume).

Boxplots plots comparing pulmonary vascular permeability index PVPI as displayed by the PiCCO-2 device (PVPI) vs. PVPI calculated as ratio of EVLW/PBV (extravascular lung water/pulmonary blood volume) vs. PVPI corrected using the formula suggested for correction of femoral indicator injection derived GEDV [16]: GEDVI[mL / m] = 0.539 * GEDVI- 15.17 + 24.49 * CI2.311* BW. Assuming that PVPIdisplayed was calculated based on GEDV and PBV not corrected for femoral injection, PVPIcorrected was calculated by multiplying PVPIdisplayed with the ratio 0.25*GEDVuncorrected/0.25*GEDVcorrected. Bland Altman plot comparing pulmonary vascular permeability index PVPI as displayed by the PiCCO-2 device (PVPI ) vs. PVPI calculated as ratio of EVLW/PBV (extravascular lung water/pulmonary blood volume). This could be explained if instead of the corrected GEDVdisplayed an uncorrected GEDVuncorrected was used for calculation of PBV and PVPIdisplayed. Based on this hypothesis and also assuming that the formula for femoral indicator injection suggested by our group is used by the latest PiCCO algorithm we determined PVPIcorrected by multiplying PVPIcalculated by the ratio GEDVuncorrected/GEDVdisplayed. As shown in Figure 5 the bias of PVPIdisplayed.d to EVLW/PBV could be reduced from -0.64 ± 0.22 to -0.10 ± 0.05. Furthermore, percentage error decreased from 22% to 4% with upper and lower limits of agreement of -0.009 and -0.190 (Figure 5).
Figure 5

Bland Altman plot comparing pulmonary vascular permeability index PVPI corrected by the correction formula for femoral indicator injection (PVPI ; see Figure3) vs. PVPI calculated as ratio of EVLW/PBV (extravascular lung water/pulmonary blood volume).

Bland Altman plot comparing pulmonary vascular permeability index PVPI corrected by the correction formula for femoral indicator injection (PVPI ; see Figure3) vs. PVPI calculated as ratio of EVLW/PBV (extravascular lung water/pulmonary blood volume).

Discussion

Despite improving outcome by a number of specific and unspecific approaches such as low tidal volume [23], extracorporeal lung assist [24] and prone positioning [25] ARDS still carries a high mortality of up to 65% [23-26]. Well balanced fluid management providing appropriate resuscitation [27] as well as avoiding fluid overload-particularly of the lungs-is another obvious approach to improve outcome in ARDS. EVLW has been validated as a significant predictor of outcome in ARDS [3,7-11]. However, with regard to fluid management it is important, if increased EVLW derives from inflammatory capillary leakage of the lungs or from general fluid overload and congestive heart failure. Relating EVLW to pulmonary blood volume, PVPI has been demonstrated to discriminate inflammatory from congestive pulmonary failure [13,14]. Therefore, combined use of EVLW and PVPI has been suggested to specify therapy in acute pulmonary failure. However, there are a number of studies suggesting pitfalls in using these parameters derived from TPTD: Due to its current indexation solely to bodyweight EVLWI might be inappropriately lowered in obese patients. Therefore, weight correction formulas [19] and indexation of EVLWI to height [19,28] have been shown to improve usefulness of EVLWI. Furthermore, femoral central vein access for TPTD indicator injection inappropriately increases GEDV(I) [15,16]. Therefore, a correction formula for GEDVI in case of femoral CVC has been suggested [16]. Requirement of information about the CVC site in the last PiCCO algorithm suggests that some kind of correction for femoral CVC is provided. Due to the assumption that PBV is 25% of GEDV in single indicator TPTD [6,12], overestimation of GEDV in femoral indicator TPTD indicator injection also pertains to PVPI. However, no information is available, – if correction provided by the new PiCCO-algorithm applies only to GEDVI or also to GEDV – if this correction is similar to the formula suggested by Saugel et al. [16] and – if PVPIdisplayed is based on PBV and GEDV derived from the correction formula suggested by our group.Therefore, we performed a prospective study in patients with femoral CVC access showing the following main results that are graphically summarized in Figure 6.
Figure 6

Algorithm for the calculation of GEDV, GEDVI, PBV and PVPI as displayed by the PiCCO-2 software V 3.1 (PVPI) and as suggested using GEDVand PBVfor the calculation of PVPI. Values not displayed are shaded.

Algorithm for the calculation of GEDV, GEDVI, PBV and PVPI as displayed by the PiCCO-2 software V 3.1 (PVPI) and as suggested using GEDVand PBVfor the calculation of PVPI. Values not displayed are shaded. GEDVdisplayed is corrected to the same degree as GEDVIdisplayed. However, PVPIdisplayed provided by the PiCCO-device is significantly lower than PVPIcalculated derived by dividing EVLWdisplayed by 0.25*GEDVdisplayed. This suggests the use of uncorrected GEDV and uncorrected PBV for the PVPI displayed by the device. This assumption is supported by PVPIcorrected values nearly identical to PVPIcalculated when correcting PVPIdisplayed using the correction formula.

Practical implications

Our data strongly suggest that the correction formula for GEDVI in case of femoral CVC indicator injection should be also applied for the calculation of PVPI. Otherwise, PVPI would remain to be artificially lowered in case of femoral CVC which also would result in different PVPI threshold for inflammatory vs. congestive pulmonary impairment.

Limitations of the study

Despite conclusive results with high statistical significance this is a single centre study with a limited number of patients and TPTD measurements. Furthermore, we have to admit that our explanations for this inconsistency are in part speculative: Although correction of PVPIdisplayed by our formula results in PVPIcorrected close to PVPIdisplayed, the hypothesis of a “neglect” of correction of PBV has to be proven by a confirmatory study. Future studies should verify the practical implications, e.g. showing improved association of PVPI with outcome when applying the correction formula also for PBV and PVPI. Furthermore, our findings could be confirmed by comparing PVPI derived from femoral and jugular indicator injection in patients with both femoral and jugular CVC. Finally, these findings pertain to the PiCCO-2 device with the V3.1 algorithm. To the best of our knowledge, there is only one other device providing PVPI which is the Edwards EV 1000 VolumeView (Edwards Lifesciences; Irvine, CA, USA). Despite some minor modifications of the thermodilution analysis this device is comparable to the PiCCO. However, the EV-1000 does not require information about the CVC site. This suggests that the device does neither correct GEDV(I) nor PVPI in case of femoral CVC site.

Conclusion

This study suggests that the V3.1 software of the PiCCO-device corrects GEDVI and GEDV for TPTD indicator injection based on a correction formula similar to the correction recently suggested [16]. However, correction of GEDV(I) does not pertain to PVPI which is contradictory to the definition of PVPI and results in a substantial underestimation of PVPI. To make PVPI values derived from jugular and femoral indicator injection comparable, corrected GEDV and PBV should be used for calculation of PVPI in case of femoral CVC.

Abbreviations

APACHE-II: Acute physiology and chronic health evaluation-II; ARDS: Acute respiratory distress syndrome; BSA: Body surface area; CO: Cardiac output; CI: Cardiac index; CVC: Central venous catheter; EVLW: Extravascular lung water; EVLWI: Extravascular lung water index; GEDV: Global end-diastolic volume; GEDVI: Global end-diastolic volume index; ICU: Intensive care unit; ITTV: Intrathoracic thermovolume; PBV: Pulmonary blood volume; PE: Percentage error; PTV: Pulmonary thermovolume; PVPI: Pulmonary vascular permeability index; RRT: Renal replacement-therapy; ROC: Receiver operating characteristics; SOFA: Sequential organ failure assessment score; TPTD: Transpulmonary thermodilution.

Competing interests

Wolfgang Huber collaborates with Pulsion Medical Systems Feldkirchen, Germany as member of the Medical Advisory Board. The other authors declare that they have no competing interest.

Authors’ contributions

HB performed the majority of measurements, participated in analysis of the data and in drafting the manuscript and finally approved the manuscript. SM performed parts of the measurement, participated in analysis of the data and in drafting the manuscript and finally approved the manuscript. AB performed parts of the measurement, participated in analysis of the data and in drafting the manuscript and finally approved the manuscript. BH performed parts of the measurement, participated in analysis of the data and in drafting the manuscript and finally approved the manuscript. RS substantially contributed to conception and design of the study, participated in the analysis of the data, participated in drafting the manuscript and finally approved the manuscript. WH performed conception and design of the study, analyzed the data, drafted the manuscript and finally approved the manuscript.

Pre-publication history

The pre-publication history for this paper can be accessed here: http://www.biomedcentral.com/1471-2253/14/81/prepub
  28 in total

1.  A meta-analysis of studies using bias and precision statistics to compare cardiac output measurement techniques.

Authors:  L A Critchley; J A Critchley
Journal:  J Clin Monit Comput       Date:  1999-02       Impact factor: 2.502

2.  Extravascular lung water determined with single transpulmonary thermodilution correlates with the severity of sepsis-induced acute lung injury.

Authors:  Vsevolod V Kuzkov; Mikhail Y Kirov; Mikhail A Sovershaev; Vladimir N Kuklin; Evgeny V Suborov; Kristine Waerhaug; Lars J Bjertnaes
Journal:  Crit Care Med       Date:  2006-06       Impact factor: 7.598

3.  Prone positioning in severe acute respiratory distress syndrome.

Authors:  Claude Guérin; Jean Reignier; Jean-Christophe Richard; Pascal Beuret; Arnaud Gacouin; Thierry Boulain; Emmanuelle Mercier; Michel Badet; Alain Mercat; Olivier Baudin; Marc Clavel; Delphine Chatellier; Samir Jaber; Sylvène Rosselli; Jordi Mancebo; Michel Sirodot; Gilles Hilbert; Christian Bengler; Jack Richecoeur; Marc Gainnier; Frédérique Bayle; Gael Bourdin; Véronique Leray; Raphaele Girard; Loredana Baboi; Louis Ayzac
Journal:  N Engl J Med       Date:  2013-05-20       Impact factor: 91.245

4.  Prognostic value of extravascular lung water in critically ill patients.

Authors:  Samir G Sakka; Magdalena Klein; Konrad Reinhart; Andreas Meier-Hellmann
Journal:  Chest       Date:  2002-12       Impact factor: 9.410

5.  Statistical methods for assessing agreement between two methods of clinical measurement.

Authors:  J M Bland; D G Altman
Journal:  Lancet       Date:  1986-02-08       Impact factor: 79.321

6.  Ventilation with lower tidal volumes as compared with traditional tidal volumes for acute lung injury and the acute respiratory distress syndrome.

Authors:  Roy G Brower; Michael A Matthay; Alan Morris; David Schoenfeld; B Taylor Thompson; Arthur Wheeler
Journal:  N Engl J Med       Date:  2000-05-04       Impact factor: 91.245

7.  Extravascular lung water to blood volume ratios as measures of permeability in sepsis-induced ALI/ARDS.

Authors:  A B Johan Groeneveld; Joanne Verheij
Journal:  Intensive Care Med       Date:  2006-06-02       Impact factor: 17.440

8.  The incidence of infectious complications of central venous catheters at the subclavian, internal jugular, and femoral sites in an intensive care unit population.

Authors:  Kedar S Deshpande; Carlo Hatem; Harry L Ulrich; Brian P Currie; Thomas K Aldrich; Christopher W Bryan-Brown; Vladimir Kvetan
Journal:  Crit Care Med       Date:  2005-01       Impact factor: 7.598

9.  How to perform indexing of extravascular lung water: a validation study.

Authors:  Stefan Wolf; Alexander Riess; Julia F Landscheidt; Christianto B Lumenta; Ludwig Schürer; Patrick Friederich
Journal:  Crit Care Med       Date:  2013-04       Impact factor: 7.598

10.  Extravascular lung water is an independent prognostic factor in patients with acute respiratory distress syndrome.

Authors:  Mathieu Jozwiak; Serena Silva; Romain Persichini; Nadia Anguel; David Osman; Christian Richard; Jean-Louis Teboul; Xavier Monnet
Journal:  Crit Care Med       Date:  2013-02       Impact factor: 7.598

View more
  7 in total

1.  Consistency of cardiac function index and global ejection fraction with global end-diastolic volume in patients with femoral central venous access for transpulmonary thermodilution: a prospective observational study.

Authors:  Analena Beitz; Helena Berbara; Sebastian Mair; Benedikt Henschel; Tobias Lahmer; Sebastian Rasch; Roland Schmid; Wolfgang Huber
Journal:  J Clin Monit Comput       Date:  2016-04-21       Impact factor: 2.502

2.  Femoral indicator injection for transpulmonary thermodilution using the EV1000/VolumeView(®): do the same criteria apply as for the PiCCO(®)?

Authors:  Wolfgang Huber; Veit Phillip; Josef Höllthaler; Caroline Schultheiss; Bernd Saugel; Roland M Schmid
Journal:  J Zhejiang Univ Sci B       Date:  2016-07       Impact factor: 3.066

3.  Comparison of cardiac function index derived from femoral and jugular indicator injection for transpulmonary thermodilution with the PiCCO-device: A prospective observational study.

Authors:  Alexander Herner; Markus Heilmaier; Ulrich Mayr; Roland M Schmid; Wolfgang Huber
Journal:  PLoS One       Date:  2018-07-31       Impact factor: 3.240

4.  Comparison of pulmonary vascular permeability index PVPI and global ejection fraction GEF derived from jugular and femoral indicator injection using the PiCCO-2 device: A prospective observational study.

Authors:  Wolfgang Huber; Andrea Gruber; Maximilian Eckmann; Felicia Elkmann; Ines Klein; Tobias Lahmer; Ulrich Mayr; Raphael Schellnegger; Jochen Schneider; Gonzalo Batres-Baires; Lisa Fekecs; Analena Beitz; Helena Berbara; Roland Schmid; Alexander Herner
Journal:  PLoS One       Date:  2017-10-17       Impact factor: 3.240

Review 5.  Haemodynamic monitoring in acute heart failure - what you need to know.

Authors:  Karolina Jasińska-Gniadzik; Piotr Szwed; Aleksandra Gasecka; Mateusz Zawadka; Marcin Grabowski; Arkadiusz Pietrasik
Journal:  Postepy Kardiol Interwencyjnej       Date:  2022-08-20       Impact factor: 1.065

6.  Extravascular lung water in critical care: recent advances and clinical applications.

Authors:  Mathieu Jozwiak; Jean-Louis Teboul; Xavier Monnet
Journal:  Ann Intensive Care       Date:  2015-11-06       Impact factor: 6.925

Review 7.  Extravascular lung water measurements in acute respiratory distress syndrome: why, how, and when?

Authors:  Takashi Tagami; Marcus Eng Hock Ong
Journal:  Curr Opin Crit Care       Date:  2018-06       Impact factor: 3.687

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.