A comprehensive study of the mechanical performance of gantry, EPID and the MLC assembly in Elekta linacs during gantry rotation.

P Rowshanfarzad1, H L Riis2, S J Zimmermann2, M A Ebert1,3.   

Abstract

OBJECTIVE: In radiotherapy treatments, it is crucial to monitor the performance of linear accelerator (linac) components, including gantry, collimation system and electronic portal imaging device (EPID) during arc deliveries. In this study, a simple EPID-based measurement method is suggested in conjunction with an algorithm to investigate the stability of these systems at various gantry angles with the aim of evaluating machine-related errors in treatments.
METHODS: The EPID sag, gantry sag, changes in source-to-detector distance (SDD), EPID and collimator skewness, EPID tilt and the sag in leaf bank assembly owing to linac rotation were separately investigated by acquisition of 37 EPID images of a simple phantom with 5 ball bearings at various gantry angles. A fast and robust software package was developed for automated analysis of the image data. Nine Elekta AB (Stockholm, Sweden) linacs of different models and number of years in service were investigated.
RESULTS: The average EPID sag was within 2 mm for all tested linacs. Some machines showed >1-mm gantry sag. Changes in the SDD values were within 1.3 cm. EPID skewness and tilt values were <1° in all machines. The maximum sag in multileaf collimator leaf bank assemblies was around 1 mm. A meaningful correlation was found between the age of the linacs and their mechanical performance. Conclusions and Advances in knowledge: The method and software developed in this study provide a simple tool for effective investigation of the behaviour of Elekta linac components with gantry rotation. Such a comprehensive study has been performed for the first time on Elekta machines.

Entities:  

Mesh:

Year:  2015        PMID: 25906294      PMCID: PMC4628515          DOI: 10.1259/bjr.20140581

Source DB:  PubMed          Journal:  Br J Radiol        ISSN: 0007-1285            Impact factor:   3.039


  28 in total

1.  A dosimetric comparison of various multileaf collimators.

Authors:  M Saiful Huq; Indra J Das; Todd Steinberg; James M Galvin
Journal:  Phys Med Biol       Date:  2002-06-21       Impact factor: 3.609

2.  Correction for dose-response variations in a scanning liquid ion chamber EPID as a function of linac gantry angle.

Authors:  P W Chin; D G Lewis; E Spezi
Journal:  Phys Med Biol       Date:  2004-04-21       Impact factor: 3.609

3.  Investigation of the sag in linac secondary collimator and MLC carriage during arc deliveries.

Authors:  Pejman Rowshanfarzad; Mahsheed Sabet; Daryl J O'Connor; Peter B Greer
Journal:  Phys Med Biol       Date:  2012-05-30       Impact factor: 3.609

4.  Verification of MLC based real-time tumor tracking using an electronic portal imaging device.

Authors:  Sarah Han-Oh; Byong Yong Yi; Fritz Lerma; Barry L Berman; Minzhi Gui; Cedric Yu
Journal:  Med Phys       Date:  2010-06       Impact factor: 4.071

5.  Three-dimensional portal image-based dose reconstruction in a virtual phantom for rapid evaluation of IMRT plans.

Authors:  W Ansbacher
Journal:  Med Phys       Date:  2006-09       Impact factor: 4.071

6.  Task Group 142 report: quality assurance of medical accelerators.

Authors:  Eric E Klein; Joseph Hanley; John Bayouth; Fang-Fang Yin; William Simon; Sean Dresser; Christopher Serago; Francisco Aguirre; Lijun Ma; Bijan Arjomandy; Chihray Liu; Carlos Sandin; Todd Holmes
Journal:  Med Phys       Date:  2009-09       Impact factor: 4.071

7.  Measuring the wobble of radiation field centers during gantry rotation and collimator movement on a linear accelerator.

Authors:  Weiliang Du; Song Gao
Journal:  Med Phys       Date:  2011-08       Impact factor: 4.071

8.  Gantry and isocenter displacements of a linear accelerator caused by an add-on micromultileaf collimator.

Authors:  Hans L Riis; Sune J Zimmermann; Mogens Hjelm-Hansen
Journal:  Med Phys       Date:  2013-03       Impact factor: 4.071

9.  Monitoring daily MLC positional errors using trajectory log files and EPID measurements for IMRT and VMAT deliveries.

Authors:  A Agnew; C E Agnew; M W D Grattan; A R Hounsell; C K McGarry
Journal:  Phys Med Biol       Date:  2014-04-15       Impact factor: 3.609

10.  EPID dosimetry for pretreatment quality assurance with two commercial systems.

Authors:  Daniel W Bailey; Lalith Kumaraswamy; Mohammad Bakhtiari; Harish K Malhotra; Matthew B Podgorsak
Journal:  J Appl Clin Med Phys       Date:  2012-07-05       Impact factor: 2.102

View more
  7 in total

Review 1.  Optical and x-ray technology synergies enabling diagnostic and therapeutic applications in medicine.

Authors:  Brian W Pogue; Brian C Wilson
Journal:  J Biomed Opt       Date:  2018-10       Impact factor: 3.170

2.  In Vivo Validation of Elekta's Clarity Autoscan for Ultrasound-based Intrafraction Motion Estimation of the Prostate During Radiation Therapy.

Authors:  Alexander Grimwood; Helen A McNair; Tuathan P O'Shea; Stephen Gilroy; Karen Thomas; Jeffrey C Bamber; Alison C Tree; Emma J Harris
Journal:  Int J Radiat Oncol Biol Phys       Date:  2018-04-16       Impact factor: 7.038

3.  Simultaneous Optimization of Radiation-Imaging Coincidence for a Multi-Energy Linac.

Authors:  Cory Knill; Raminder Sandhu; Robert Halford; Michael Snyder; Danielle Lack
Journal:  J Med Phys       Date:  2021-08-07

4.  Implementing corrections of isocentric shifts for the stereotactic irradiation of cerebral targets: Clinical validation.

Authors:  Sonja Wegener; Robert Schindhelm; Otto A Sauer
Journal:  J Appl Clin Med Phys       Date:  2022-03-02       Impact factor: 2.243

5.  Investigation of the mechanical performance of Siemens linacs components during arc: gantry, MLC, and electronic portal imaging device.

Authors:  Pejman Rowshanfarzad; Peter Häring; Hans L Riis; Sune J Zimmermann; Martin A Ebert
Journal:  Med Devices (Auckl)       Date:  2015-11-05

6.  The impact of isocentric shifts on delivery accuracy during the irradiation of small cerebral targets-Quantification and possible corrections.

Authors:  Linda J Wack; Florian Exner; Sonja Wegener; Otto A Sauer
Journal:  J Appl Clin Med Phys       Date:  2020-03-20       Impact factor: 2.102

7.  Dosimetric evaluation of the gantry sag effect in clinical SRS plans.

Authors:  Egor Borzov; Alex Nevelsky; Rachel Bar-Deroma; Itzhak Orion
Journal:  BJR Open       Date:  2019-02-13
  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.