| Literature DB >> 34566290 |
Cory Knill1, Raminder Sandhu1, Robert Halford1, Michael Snyder1, Danielle Lack2.
Abstract
INTRODUCTION: Medical physics guidelines stress the importance of radiation-imaging coincidence, especially for stereotactic treatments. However, multi-energy linear accelerators may only allow a single imaging isocenter. A procedure was developed to simultaneously optimize radiation-imaging isocenter coincidence for all linac photon energies on a Versa HD.Entities:
Keywords: Image-guided radiotherapy; isocentricity; linac; stereotactic body radiation therapy; stereotactic radiosurgery
Year: 2021 PMID: 34566290 PMCID: PMC8415251 DOI: 10.4103/jmp.JMP_7_21
Source DB: PubMed Journal: J Med Phys ISSN: 0971-6203
Figure 1Half beam block setup for steering beam centers. (a) Farmer chamber is taped on a block tray at the central axis in the gun-target direction, (b) the collimator is closed to the central axis to block half the field
Figure 2Correlating half beam block measurements with BB-beam center deviation results in the gun-target direction for Versa HD photon energies (6X, 10X, 15X, 6FFF, and 10FFF). The 0.0% half beam block result corresponding to 0.87 mm is the desired beam block-beam center gun-target deviation
Figure 3Versa HD couch schematic showing the location of adjustment bolts A and B. The gun (G) and target (T) directions are also labeled, while the up (U) and down (D) directions are not shown
Figure 4Iterative adjustment of bending fine to achieve the ideal BB-beam center gun-target deviation (0.87 mm). The 10X energy did not require adjustment
Versa HD multi-energy isocentricity
| Versa HD multi-energy isocentricity | ||||||
|---|---|---|---|---|---|---|
|
| ||||||
| Linac (mm) | Couch (mm) | Overall radius (mm) | ||||
|
|
| |||||
| AB | GT | UD | Table X | Table Y | ||
| 6X | 0.04 | 0.00 | 0.08 | 0.15 | −0.08 | 0.8 |
| 10X | 0.02 | −0.06 | −0.01 | 0.1 | 0.12 | 0.88 |
| 15X | 0.02 | −0.01 | −0.11 | 0.04 | 0.20 | 0.91 |
| 6FFF | −0.06 | 0.04 | 0.07 | −0.15 | −0.08 | 0.98 |
| 10FFF | −0.03 | 0.01 | −0.01 | −0.18 | −0.06 | 0.87 |
| Average | 0.00 | 0.00 | 0.00 | −0.01 | 0.02 | 0.89 |
WL results showing the linac isocenter coincidence across all energies, coincidence between the couch rotation axis and linac isocenter, and overall radiation-imaging coincidence for all gantry, collimator, and couch angles. AB: couch left-right, GT: linac gun-target, UD: couch up-down
Versa HD isocentricity tolerances
| Versa HD isocentricity tolerances | ||
|---|---|---|
|
| ||
| Tolerance A (mm) | Tolerance B (mm) | |
| Beam center versus collimator rotation axis coincidence | 0.20 | 0.1 |
| Cross-energy linac isocenter coincidence | 0.50 | 0.20 |
| Linac isocenter versus couch rotation axis coincidence | 0.50 | 0.25 |
Tolerance A: Easily achievable on all Versa HDs. Tolerance B: lower expected limit of what is achievable