Correction for dose-response variations in a scanning liquid ion chamber EPID as a function of linac gantry angle.

P W Chin1, D G Lewis, E Spezi.   

Abstract

The response of electronic portal imaging devices (EPIDs) of the scanning liquid ionization chamber (SLIC) type is known to vary with linear accelerator gantry angle. This work considered several contributing factors, quantified the artefacts, monitored their reproducibility and investigated the effects of repeated gantry rotations. Unflatness of up to 5% was found. A correction technique was devised using nonlinear regression of a three-variable sinusoidal modulation. Comparison with two existing techniques found our method to be the most effective, providing a flatness well within 2%. This improved accuracy is expected to benefit more accurate dosimetric studies in particular. The post-acquisition correction process required no change in imaging protocols. Applicability of the new technique was demonstrated on images acquired on different days and with different beam sizes. Since the artefacts compromise both accurate dosimetry and image quality, their successful removal should benefit a broad range of SLIC EPID applications.

Entities:  

Mesh:

Substances:

Year:  2004        PMID: 15152695     DOI: 10.1088/0031-9155/49/8/n01

Source DB:  PubMed          Journal:  Phys Med Biol        ISSN: 0031-9155            Impact factor:   3.609


  4 in total

1.  A comprehensive study of the mechanical performance of gantry, EPID and the MLC assembly in Elekta linacs during gantry rotation.

Authors:  P Rowshanfarzad; H L Riis; S J Zimmermann; M A Ebert
Journal:  Br J Radiol       Date:  2015-04-23       Impact factor: 3.039

2.  An EPID-based method for comprehensive verification of gantry, EPID and the MLC carriage positional accuracy in Varian linacs during arc treatments.

Authors:  Pejman Rowshanfarzad; Conor K McGarry; Michael P Barnes; Mahsheed Sabet; Martin A Ebert
Journal:  Radiat Oncol       Date:  2014-11-26       Impact factor: 3.481

3.  Investigation of the mechanical performance of Siemens linacs components during arc: gantry, MLC, and electronic portal imaging device.

Authors:  Pejman Rowshanfarzad; Peter Häring; Hans L Riis; Sune J Zimmermann; Martin A Ebert
Journal:  Med Devices (Auckl)       Date:  2015-11-05

4.  Comparison of measured and computed portal dose for IMRT treatment.

Authors:  Savino Cilla; Pietro Viola; Luigi Azario; Luca Grimaldi; Maurizio Craus; Guido D'Onofrio; Andrea Fidanzio; Alessio Giuseppe Morganti; Angelo Piermattei
Journal:  J Appl Clin Med Phys       Date:  2006-08-24       Impact factor: 2.102

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.