Literature DB >> 29766058

Prospective observational study of point-of-care creatinine in trauma.

Anthony J Carden1, Edgardo S Salcedo1, Nam K Tran2, Eric Gross3, Jennifer Mattice3, Jan Shepard3, Joseph M Galante1.   

Abstract

BACKGROUND: Patients with trauma are at risk for renal dysfunction from hypovolemia or urological injury. In austere environments, creatinine values are not available to guide resuscitation. A new portable device, the Stat Sensor Point-of-care (POC) Whole Blood Creatinine Analyzer, provides accurate results in <30 s and requires minimal training. This device has not been evaluated in trauma despite the theoretical benefit it provides. The purpose of this study is to determine the clinical impact of the POC device in trauma.
METHODS: 40 patients with trauma were enrolled in a prospective observational study. One drop of blood was used for creatinine determination on the Statsensor POC device. POC creatinine results were compared to the laboratory. Turnaround time (TAT) for POC and laboratory methods was calculated as well as time elapsed to CT scan if applicable.
RESULTS: Patients (n=40) were enrolled between December 2014 and March 2015. POC creatinine values were similar to laboratory methods with a mean bias of 0.075±0.27 (p=0.08). Mean analytical TATs for the POC measurements were significantly faster than the laboratory method (11.6±10.0 min vs 78.1±27.9 min, n=40, p<0.0001). Mean elapsed time before arrival at the CT scanner was 52.9±34.2 min.
CONCLUSIONS: The POC device reported similar creatinine values to the laboratory and provided significantly faster results. POC creatinine testing is a promising development for trauma practice in austere environments and workup of a subset of stable patients with trauma. Further study is warranted to determine clinical impact, both in hospital-based trauma and austere environments.

Entities:  

Keywords:  Creatinine; Point-of-care testing; laboratory; renal function

Year:  2016        PMID: 29766058      PMCID: PMC5891719          DOI: 10.1136/tsaco-2016-000014

Source DB:  PubMed          Journal:  Trauma Surg Acute Care Open        ISSN: 2397-5776


Introduction

Trauma patients are at risk for renal dysfunction from hypovolemia or urological injury. In austere environments, however, creatinine values are typically not available to guide resuscitation. A new portable device, the Stat Sensor Point-of-care (POC) Whole Blood Creatinine Analyzer (Nova Biomedical, Waltham, Massachusetts, USA), has been evaluated in other fields, including monitoring resuscitation in burn surgery,1 2 trending post-operative renal function in patients with renal transplant,3 and in radiology departments to screen for renal failure prior to contrast administration.4 It provides accurate results4–6 in <30 s and requires minimal training. This device has not been evaluated in the trauma setting despite the theoretical benefit it provides. This device can potentially be used for rapid creatinine results in austere environments7 that historically do not have access to renal function tests. The purpose of this study is to determine the potential clinical impact of this POC creatinine analyzer in the trauma setting. Our hypothesis is that the POC creatinine device can be integrated into the existing trauma system and will result creatinine values similar to the hospital laboratory, but significantly faster.

Methods

The present study was approved by the Institutional Review Board. We performed a prospective observational study of adult patients who came into the emergency room at a single Level I trauma center coded as major traumas, based on prehospital triage criteria. Prisoners, pediatric patients (<18 years old), and pregnant women were excluded. Patients were also excluded if their planned workup did not include creatinine testing. Enrollment occurred between 6:00 and 18:00, on weekdays between January and April of 2015, in a population of convenience. The primary outcome was POC creatinine level, compared to the laboratory value. The secondary outcomes were turnaround time (TAT) of the POC and laboratory creatinine levels and the time elapsed for the patient to arrive at the CT scanner, when applicable. For each enrolled patient, a single drop of blood was taken from the same initial sample as the patient's other laboratory tests. The blood sample was tested using the POC whole blood creatinine meter. The creatinine value, the time it was obtained, and the patient's arrival time were recorded. The creatinine and time information was not provided to the managing trauma team and did not impact clinical decisions. Informed consent was obtained within 24 hours of arrival, prior to further data abstraction from the medical record. The electronic medical record was used to obtain the laboratory creatinine value, the time it became available, and the time the patient arrived at the CT scanner, when applicable. POC creatinine results were compared directly to the laboratory value. TATs for POC and laboratory methods were calculated as well as time elapsed to CT scan when applicable. Creatinine values and TATs were calculated using a standard t-test.

Results

Forty patients were enrolled between December 2014 and March 2015. The mean age was 71.2 years. Thirty per cent were women. Ninety-five per cent had sustained blunt trauma and 20% required operative management. The median hospital length of stay was 2 days. The median injury severity score was 4 (range 0–36). Please see table 1 for demographics.
Table 1

Demographics

Mean age71.2 years
Gender30% women
Type of trauma95% blunt, 5% penetrating
Median hospital length of stay2 days
Patients who underwent operative management20%
Median injury severity score4
Demographics Mean POC creatinine values were 1.01±0.33, compared to 1.08±0.34 for laboratory creatinine. The difference between these values was not statistically significant, with a mean bias or difference between POC and laboratory values, of 0.075±0.27 (p=0.08). Figure 1 is a Bland-Altman Plot, which shows the laboratory creatinine value on the x axis and the bias on the y axis. Figure 2 plots laboratory creatinine on the x-axis and POC creatinine on the y-axis.
Figure 1

Bland-Altman plot. POC, point-of care.

Figure 2

POC versus laboratory creatinine. POC, point-of care.

Bland-Altman plot. POC, point-of care. POC versus laboratory creatinine. POC, point-of care. Mean analytical TATs for the POC measurements were significantly faster than the laboratory method (11.6±10.0 min vs 78.1±27.9 min, p<0.0001). The mean elapsed time before arrival at the CT scanner for the 30 patients who underwent CT imaging was 52.9±34.2 min. At the time of CT scanning, 29 of 30 had POC creatinine values available, while only 5 of 30 had laboratory creatinine values available. Please see table 2 for these results.
Table 2

Creatinine data

POC deviceLaboratoryDifferencep Value
Mean creatinine value (mg/dL)1.01±0.331.08±0.34−0.07±0.27 (bias)0.08
Mean time to availability (min)11.6±10.078.1±27.966.4±27.4<0.0001

POC, point-of-care.

Creatinine data POC, point-of-care.

Discussion

The purpose of this prospective observational study is to determine the potential clinical impact of this POC creatinine analyzer in the trauma setting. While its accuracy was known in other settings, it was important to know if the device could be used in a practical manner in the trauma setting. We had three key findings: the device provided accurate creatinine values, the information was available rapidly, typically before contrast was administered for imaging, and it was successfully implemented in our trauma workflow. The study supported the hypothesis that the POC creatinine device can be integrated into the existing trauma system and will result in creatinine values similar to the hospital laboratory, but significantly faster. The strength of this study is that it demonstrates prospectively that the device is functional in the trauma setting. The study design was chosen since the use of this device has not been established in trauma. This study has several limitations. As a small, prospective observational study, it does not evaluate a specific intervention; neither does it influence the situations in which this device should be used. Further study is needed to determine if obtaining creatinine values truly impacts management in the practical setting, as well as to evaluate the short-term and long-term outcomes of using this information to tailor workup and management of individual patients. On initial observation, the p value of 0.08 for the difference between POC and laboratory creatinine values is close to our predefined cutoff of 0.05, suggesting that there is a chance there would be a statistically significant difference if a larger sample was evaluated. Despite this, the absolute difference, or bias, of 0.07 is low and is the more critical value to evaluate in laboratory medicine. This tells us that even if there is a systematic difference between the values that is due to more than chance, it is small enough to not be clinically significant. This could be assessed further with a larger study. The difference in values may be explained by the difference in biochemical reactions as the POC device uses electrochemistry to test whole blood and the laboratory uses the kinetic Jaffe reaction on plasma. The vast majority of patients in this study had normal creatinine values. Therefore, while the data support the accuracy of the creatinine values for patients with normal creatinine, the accuracy outside of the normal range was not demonstrated. A larger study or a different patient population would be necessary to look at the device in patients with renal failure. This device may affect practice in austere environments, where determination of creatinine values with a small, inexpensive device could prove beneficial. There is a precedent for POC creatinine determination in an austere environment;7 more research would be needed to determine its role in the care of trauma and surgical patients. Potential uses include resuscitation and management during prolonged transport and management of perioperative patients without access to a full laboratory. Forward surgical teams in role II settings may benefit by having an easily portable device that can be transported to each destination. Additionally, development is underway for similar devices that measure other important values, including hemoglobin. This study also has the potential to affect a subset of trauma patients who have a low probability of critical findings on CT scanning, but have an increased likelihood of developing contrast-induced nephropathy. Examples of patients who may benefit are elderly patients who sustain a ground level fall requiring CT imaging and patients who are found down and at risk for rhabdomyolysis. The elderly portion of the trauma population is steadily growing, and even low energy mechanisms can cause significant injury in these patients. If found to have elevated creatinine, elderly patients could be resuscitated prior to a delayed CT scan with contrast, or an algorithm could be used to initially scan without contrast. Notably, the fact that creatinine values were available prior to contrast administration in this study means that this algorithm could potentially be implemented with minimal delay of care. Determination of creatinine values with a POC device is feasible in a major trauma center. More research would be necessary to determine its feasibility in an austere environment. The device is quite portable; however, the test strips must be refrigerated in the long term. Our outcome of obtaining similar creatinine results sooner than laboratory values is likely reproducible in further clinical practice or research. It required no change to the structure of our trauma service. Further research will be necessary to determine if obtaining creatinine values faster leads to improvements in outcomes such as rate of contrast-induced nephropathy or time markers such as time to CT scanner, time to operating room, or time to admission. A randomized controlled trial would be beneficial to determine if POC creatinine determination, coupled with a protocol for managing select patients with a low energy trauma mechanism, can achieve better outcomes. Please see figure 3 for a sample algorithm for use in elderly ground level fall patients.
Figure 3

Example algorithm for elderly (age>60) ground-level fall patients.

Example algorithm for elderly (age>60) ground-level fall patients. This device also has the potential to be used in austere environments due to its small size and inexpensive cost. Rapid determination of renal function can tailor resuscitation efforts during extended transport, or in the field. A rigorous study to establish feasibility can be followed by determination of the patient populations likely to benefit and the algorithms that can incorporate rapid creatinine values. Results of the POC testing also need to be tested in patients with profound shock and high ISS. POC creatinine testing is a promising development for trauma practice in austere environments and workup of a subset of stable trauma patients with trauma. Further study is warranted to determine the clinical impact, both hospital-based trauma and austere environments.
  7 in total

1.  Application of a point of care creatinine device for trend monitoring in kidney transplant patients: fit for purpose?

Authors:  Céline L van Lint; Paul J M van der Boog; Fred P H T M Romijn; Paul W Schenk; Sandra van Dijk; Ton J M Rövekamp; Anja Kessler; Lothar Siekmann; Ton J Rabelink; Christa M Cobbaert
Journal:  Clin Chem Lab Med       Date:  2015-09-01       Impact factor: 3.694

2.  Evaluation of a handheld creatinine measurement device for real-time determination of serum creatinine in radiology departments.

Authors:  Stefan Haneder; Alexandra Gutfleisch; Claudia Meier; Joachim Brade; Dieter Hannak; Stefan O Schoenberg; Christoph R Becker; Henrik J Michaely
Journal:  World J Radiol       Date:  2012-07-28

3.  Whole blood neutrophil gelatinase-associated lipocalin predicts acute kidney injury in burn patients.

Authors:  Soman Sen; Zack R Godwin; Tina Palmieri; David Greenhalgh; Amanda N Steele; Nam K Tran
Journal:  J Surg Res       Date:  2015-03-20       Impact factor: 2.192

4.  Point-of-care B-type natriuretic peptide and neutrophil gelatinase-associated lipocalin measurements for acute resuscitation: a pilot study.

Authors:  Erin Howell; Soman Sen; Tina Palmieri; Zack Godwin; Jennifer Bockhold; David Greenhalgh; Nam K Tran
Journal:  J Burn Care Res       Date:  2015 Mar-Apr       Impact factor: 1.845

5.  Evaluation of the analytical performance of the nova StatSensor creatinine meter for blood testing.

Authors:  Chayanit Srihong; Kavinyanee Pangsapa; Kanyaorn Chuaboonmee; Yongyuth Kotipan; Navapun Charuruks
Journal:  J Med Assoc Thai       Date:  2012-09

6.  Chronic kidney disease prevalence in Rivas, Nicaragua: use of a field device for creatinine measurement.

Authors:  Kailey Minnings; Eric Kerns; Melissa Fiore; Madeline Fiore; Rulan S Parekh; Jeff DuBois; T Scott Isbell; Ryan Ferguson; Louis Fiore
Journal:  Clin Biochem       Date:  2015-01-21       Impact factor: 3.281

7.  Evaluation of the Nova StatSensor® Xpress(TM) Creatinine point-of-care handheld analyzer.

Authors:  Cara Simone Kosack; Wim de Kieviet; Kubra Bayrak; Anastacija Milovic; Anne Laure Page
Journal:  PLoS One       Date:  2015-04-17       Impact factor: 3.240

  7 in total
  2 in total

1.  Point of care creatinine derived eGFR measurement in capillary blood for identifying patients at risk.

Authors:  Sandrine Lemoine; Anne-Cécile Rouveure; Laurence Dubourg; Solenne Pelletier; Christelle Marolho; Evelyne Decullier; Maurice Laville
Journal:  Pract Lab Med       Date:  2022-07-14

2.  Undisclosed Antiretroviral Therapy Use at Primary Health Care Clinics in Rural KwaZulu Natal South Africa: A DO-ART Trial Sub-study.

Authors:  Nsika Sithole; Resign Gunda; Olivier Koole; Meighan Krows; Torin Schaafsma; Mosa Moshabela; Mark J Siedner; Connie Celum; Ruanne V Barnabas
Journal:  AIDS Behav       Date:  2021-06-07
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.