| Literature DB >> 25883980 |
J H Lee1, L T Chang1, A C Shiau2, C W Chen3, Y J Liao4, W J Li5, M S Lee6, S M Hsu7.
Abstract
A standard protocol of dosimetric measurements is used by the organizations responsible for verifying that the doses delivered in radiation-therapy institutions are within authorized limits. This study evaluated a self-designed simple auditing phantom for use in verifying the dose of radiation therapy; the phantom design, dose audit system, and clinical tests are described. Thermoluminescent dosimeters (TLDs) were used as postal dosimeters, and mailable phantoms were produced for use in postal audits. Correction factors are important for converting TLD readout values from phantoms into the absorbed dose in water. The phantom scatter correction factor was used to quantify the difference in the scattered dose between a solid water phantom and homemade phantoms; its value ranged from 1.084 to 1.031. The energy-dependence correction factor was used to compare the TLD readout of the unit dose irradiated by audit beam energies with (60)Co in the solid water phantom; its value was 0.99 to 1.01. The setup-condition factor was used to correct for differences in dose-output calibration conditions. Clinical tests of the device calibrating the dose output revealed that the dose deviation was within 3%. Therefore, our homemade phantoms and dosimetric system can be applied for accurately verifying the doses applied in radiation-therapy institutions.Entities:
Mesh:
Year: 2015 PMID: 25883980 PMCID: PMC4391650 DOI: 10.1155/2015/934387
Source DB: PubMed Journal: Biomed Res Int Impact factor: 3.411
Figure 1(a) Self-designed simple phantom. (b) Dosimeter hole in homemade phantoms.
Figure 2The schematic of setup condition.
Figure 3The dose linearity of TLD (calculated dose versus irradiated dose).
Figure 4The energy-dependence correction factor of TLD.
Figure 5Scatter correction factor of phantom.
The dose comparison between ionization chamber and TLDs.
|
|
| Difference (%) | |
|---|---|---|---|
| 6 MV photon | 194.48 ± 2.32 | 199.50 ± 1.92 | −2.519% |
| 10 MV photon | 197.21 ± 3.93 | 197.97 ± 1.87 | −0.386% |
| 6 MeV electron | 203.71 ± 2.04 | 199.32 ± 1.56 | 2.205% |
| 9 MeV electron | 194.84 ± 3.62 | 198.45 ± 2.03 | −1.819% |
| 12 MeV electron | 198.99 ± 3.13 | 198.04 ± 1.58 | 0.478% |
D TLD: dose measured from TLDs; D IC: dose measured from the ionization chamber.