Literature DB >> 10869742

The ESTRO-QUALity assurance network (EQUAL).

I H Ferreira1, A Dutreix, A Bridier, J Chavaudra, H Svensson.   

Abstract

BACKGROUND AND
PURPOSE: ESTRO has set up a Quality Assurance network (EQUAL) to check the dose delivered on axis in reference and non-reference conditions for external radiotherapy. The external audits covered by the network are based on measurements made with mailed thermoluminescent dosimeters (TLD).
MATERIAL AND METHODS: The TLD consist of LiF powder type DTL 937 read with a PCL 3 automatic TLD reader. The participating centres are instructed to deliver to the TLDs absorbed doses of 2 Gy calculated with the Treatment Planning System used in clinical routine. A maximum of three photon energies by participating centre have been checked with 10 on-axis points per beam. The quantities checked include the reference beam output, beam output variation with collimator opening, depth dose data and wedge transmission factor.
RESULTS: During the 1998 EQUAL programme 102 centres have been checked corresponding to 235 beams (28 (60)Co beams and 207 X-ray beams). About 3% of the outputs in reference conditions show deviations outside tolerance level (>+/-5%). A similar rate of deviation is noted for the percentage depth doses. A rate of deviation (6%) has been observed for the beam output variation (open and wedged beams) and the wedge transmission factor. The analysis of the results shows that for 24 out of the 102 centres, a deviation outside tolerance level is observed at least in one point, mainly for the large and rectangular field sizes and for the wedged beams.
CONCLUSIONS: The results for the EQUAL programme show the importance of a quality assurance network in Radiotherapy especially for the non reference points even if they are only located on the beam axis (In order to participate in this network, please contact EQUAL secretariat or download the attached application form ESTRO web site: Dr I.H. Ferreira or Mrs Aline Mechet, EQUAL-ESTRO, Physics Department, Institut Gustave-Roussy 39 Rue Camille Desmoulins, F-94805 Villejuif Cedex, France. e-mail:equal@igr.fr or http://www.estro.be/).

Entities:  

Mesh:

Year:  2000        PMID: 10869742     DOI: 10.1016/s0167-8140(99)00101-2

Source DB:  PubMed          Journal:  Radiother Oncol        ISSN: 0167-8140            Impact factor:   6.280


  12 in total

1.  Design and implementation of a "cheese" phantom-based Tomotherapy TLD dose intercomparison.

Authors:  Hans Schiefer; Konrad Buchauer; Simon Heinze; Guido Henke; Ludwig Plasswilm
Journal:  Strahlenther Onkol       Date:  2015-06-19       Impact factor: 3.621

Review 2.  Dosimetric audit in brachytherapy.

Authors:  A L Palmer; D A Bradley; A Nisbet
Journal:  Br J Radiol       Date:  2014-05-06       Impact factor: 3.039

Review 3.  The contribution of women to radiobiology: Marie Curie and beyond.

Authors:  Anna Gasinska
Journal:  Rep Pract Oncol Radiother       Date:  2015-12-29

4.  Dose verification in carcinoma of uterine cervix patients undergoing 3D conformal radiotherapy with Farmer type ion chamber.

Authors:  Challapalli Srinivas; P Suman Kumar; Ramamoorthy Ravichandran; S Banerjee; P U Saxena; E S Arun Kumar; Dinesh K Pai
Journal:  J Med Phys       Date:  2014-10

5.  Radiotherapy quality assurance review in a multi-center randomized trial of limited-disease small cell lung cancer: the Japan Clinical Oncology Group (JCOG) trial 0202.

Authors:  Naoko Sanuki-Fujimoto; Satoshi Ishikura; Kazushige Hayakawa; Kaoru Kubota; Yutaka Nishiwaki; Tomohide Tamura
Journal:  Radiat Oncol       Date:  2009-06-02       Impact factor: 3.481

6.  Radiotherapy dosimetry audit: three decades of improving standards and accuracy in UK clinical practice and trials.

Authors:  Catharine H Clark; Edwin G A Aird; Steve Bolton; Elizabeth A Miles; Andrew Nisbet; Julia A D Snaith; Russell A S Thomas; Karen Venables; David I Thwaites
Journal:  Br J Radiol       Date:  2015-09-02       Impact factor: 3.039

7.  Eight years of IMRT quality assurance with ionization chambers and film dosimetry: experience of the Montpellier Comprehensive Cancer Center.

Authors:  Pascal Fenoglietto; Benoit Laliberté; Norbert Aillères; Olivier Riou; Jean-Bernard Dubois; David Azria
Journal:  Radiat Oncol       Date:  2011-07-20       Impact factor: 3.481

8.  A novel simple phantom for verifying the dose of radiation therapy.

Authors:  J H Lee; L T Chang; A C Shiau; C W Chen; Y J Liao; W J Li; M S Lee; S M Hsu
Journal:  Biomed Res Int       Date:  2015-03-26       Impact factor: 3.411

Review 9.  Evolving Clinical Cancer Radiotherapy: Concerns Regarding Normal Tissue Protection and Quality Assurance.

Authors:  Won Hoon Choi; Jaeho Cho
Journal:  J Korean Med Sci       Date:  2016-01-28       Impact factor: 2.153

10.  Beam Output Audit results within the EORTC Radiation Oncology Group network.

Authors:  Coen W Hurkmans; Melissa Christiaens; Sandra Collette; Damien Charles Weber
Journal:  Radiat Oncol       Date:  2016-12-15       Impact factor: 3.481