Hanna Tolonen1, Päivikki Koponen2, Androniki Naska3, Satu Männistö4, Grazyna Broda5, Tarja Palosaari6, Kari Kuulasmaa7.
Abstract
BACKGROUND: Accurate blood pressure measurements are needed in clinical practice, intervention studies and health examination surveys. Blood pressure measurements are sensitive: their accuracy can be affected by measurement environment, behaviour of the subject, measurement procedures, devices used for the measurement and the observer. To minimize errors in blood pressure measurement, a standardized measurement protocol is needed.Entities:
Mesh:
Year: 2015 PMID: 25880766 PMCID: PMC4403933 DOI: 10.1186/s12874-015-0020-3
Source DB: PubMed Journal: BMC Med Res Methodol ISSN: 1471-2288 Impact factor: 4.615
Reported effect of different factors of observed blood pressure levels
|
|
|
|
|
|---|---|---|---|
| Environment in which measurement is conducted | Cold room vs. comfortable room temperature [ | ⇑ 14 mmHg | ⇑ 15 mmHg |
| Person being measured | Full bladder [ | ⇑ 10–15 mmHg, in case of uncomfortably distended bladder | ⇑ 10 mmHg, in case of uncomfortably distended bladder ⇑ 40 mmHg |
| ⇑ 50 mmHg | |||
| Heavy physical exercise before measurement [ | ⇓ 18–20 mmHg | ⇓ 7–9 mmHg | |
| Heavy meal before measurement [ | ⇓ 20 mmHg | ⇓ 20 mmHg | |
| Smoking before measurement [ | ⇑ 10 mmHg | ⇑ 8 mmHg | |
| Measurement procedure | Not resting at least 5 min before measurement [ | ⇑ 10–20 mmHg | ⇑ 14 mmHg |
| Left arm vs. right arm [ | ⇓ 1–3 mmHg | ⇑ 1 mmHg | |
| Supine vs. sitting [ | ⇑ 3–10 mmHg | ⇑ 1–5 mmHg | |
| Back/feet unsupported vs. supported [ | ⇑ 5–15 mmHg | ⇑ 6 mmHg | |
| Arm unsupported during the measurement vs. arm supported [ | ⇑ 1–7 mmHg | ⇑ 5–11 mmHg | |
| Legs crossed vs. not crossed [ | ⇑ 5–8 mmHg | ⇑ 3–5 mmHg | |
| Talking during the measurement vs. silent [ | ⇑ 17 mmHg | ⇑ 13 mmHg | |
| Arm below heart level vs. arm at the heart level [ | ⇑ 10 mmHg | ⇑ 10 mmHg | |
| Cuff over clothing vs. cuff on bare arm [ | ⇑ 5 mmHg | not reported | |
| Device | Cuff too large [ | ⇓ 10–30 mmHg | ⇓ 10–30 mmHg |
| Cuff too small [ | ⇑ 3–12 mmHg, in obese persons | ⇑ 2–8 mmHg, in obese persons | |
| ⇑ 30 mmHg | ⇑ 30 mmHg | ||
| Diaphragm of stethoscope vs. bell (auscultation method used) [ | ⇓ 0–2 mmHg | ⇓ 0–2 mmHg | |
| Calibration error [ | 0–5 mmHg | 0–5 mmH |
⇑ Arrow upwards: increases observed blood pressure level; ⇓ arrow down wards: decreases observed blood pressure level.
Blood pressure measurement protocols used in EHES pilot surveys
|
|
|
|
|
|
|
|
|
|
|
|---|---|---|---|---|---|---|---|---|---|
| A | 123 | May-Jun 2011 | 22.3 (21,24) | Right | Sitting | Mercury sphygmomanometer | Riester diplomat presometer | 22–32 cm and 33–41 cm | Bell |
| B | 111 | Apr-Jul 2011 | 25.0 (18,33) | Right1,2 | Sitting3,4 | Mercury sphygmomanometer | Riester diplomat presometer | 17–26 cm, 24–32 cm, 32–48 cm | Diaphragm |
| C | 393 | Oct 2010-Jan 2011 | 22.8 (21,26) | Right | Sitting | Mercury sphygmomanometer | Riva Rocci | ≤34 cm, > 34 cm | Bell |
| D | 183 | Nov 2010-Jan 2011 | 23.3 (22,24) | Right1,2 | Sitting3,4 | Oscillometric device | Omron i-c10 | 22–42 cm | not relevant |
| E | 305 | May-Jun 2011 | 23.0 (16,29) | Right1 | Sitting | Oscillometric device | Omron i-c10 | 22–42 cm | not relevant |
| F | 131 | Nov 2010-Feb 2011 | 23.2 (22,25) | Right1,2 | Sitting3,4 | Oscillometric device | Omron i-c10 | 22–42 cm, > 42 cm | not relevant |
| G | 1302 | Oct-Dec 2010 | # | Right1 | Sitting | Oscillometric device | Omron M6 | 17–22 cm, 22–32 cm, 32–42 cm | not relevant |
| H | 137 | May-Jul 2010 | # | Right1,2 | Sitting3 | Oscillometric device | Omron M6 | 22–32 cm | not relevant |
| I | 168 | Jan-Mar 2011 | 21.3 (16,25) | Right1,2 | Sitting3,4 | Oscillometric device | Omron 705IT | ≤21 cm, 22–31 cm, ≥ 32 cm | not relevant |
| J | 922 | Jan-May 2011 | 19.4 (12,25) | Right1 | Sitting | Oscillometric device | Omron HEM-907 | 17–22 cm, 22–32 cm, 32–42 cm | not relevant |
| K | 190 | Jun-Jul 2011 | 23.4 (19,32) | Right1 | Sitting3 | Oscillometric device | Datascope Accutorr Plus | 13–20 cm, 21–27 cm, 28–35 cm, 36–46 cm | not relevant |
| L | 162 | Nov-Dec 2010 | # | Right1 | Sitting | Oscillometric device | Citizen CH-432B | 20–26 cm, 25–34 cm, 32–43 cm | not relevant |
# Room temperature not measured, 1If the measurement was done on the left arm that was recorded, 2The reason for use of left arm was also recorded, 3If the measurement was done in the supine posture, that was recorded, 4The reason why the measurement was done in the supine posture was also recorded.
Recording of cuff size used and measured arm circumference, mean arm circumference, proportion of miss-cuffed subject, and proportion of identical readings
|
|
|
|
|
|
|
| ||
|---|---|---|---|---|---|---|---|---|
|
|
|
|
| |||||
| A | No | B | 30.8 cm (24.5,43.5) | # | 8% | 29% | 16% | 28% |
| B | Yes | B | 31.6 cm (23.0,48.0) | 1% | 10% | 29% | 31% | 41% |
| C | Yes | B | 28.3 cm (20.0, 40.0) | 0% | 10% | 13% | 19% | 19% |
| D | Yes | N | § | # | 4% | 10% | 7% | 12% |
| E | Yes | B | 29.9 cm (22.0,41.5) | 0% | 5% | 9% | 6% | 11% |
| F | Yes | B | 28.9 cm (21.0,41.5) | 2% | 3% | 9% | 9% | 11% |
| G | Yes | M | § | # | 7% | 12% | 8% | 13% |
| H | Yes | B | 30.1 cm (21.0,48.0) | 20% | 3% | 8% | 5% | 9% |
| I | Yes | B | 30.9 cm (22.0,46.0) | 0% | 8% | 10% | 8% | 10% |
| J | Yes | N | § | # | 6% | 14% | 9% | 15% |
| K | Yes | B | 30.4 cm (21.5,43.3) | 5% | 4% | 10% | 7% | 10% |
| L | Yes | B | 30.4 cm (24.0,40.0) | 1% | 5% | 9% | 5% | 11% |
§ Arm circumference not measured, # Not possible to calculate.
Proportion of terminal digits for systolic blood pressure
|
|
|
|
| |||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|
|
|
|
|
|
|
|
|
|
|
| |||
| Mercury sphygmomanometer | A | 22 | 0 | 23 | 0 | 21 | 0 | 18 | 0 | 17 | 0 | 0.4488 |
| B | 33 | 6 | 9 | 3 | 5 | 22 | 5 | 4 | 9 | 4 | <0.0001 | |
| C | 18 | 0 | 20 | 0 | 20 | 0 | 19 | 0 | 23 | 0 | 0.1352 | |
| Omron i-C10 | D | 9 | 11 | 11 | 9 | 10 | 11 | 11 | 9 | 11 | 9 | 0.9826 |
| E | 11 | 10 | 10 | 9 | 10 | 10 | 9 | 11 | 10 | 9 | 0.7936 | |
| F | 12 | 9 | 11 | 7 | 11 | 7 | 6 | 13 | 15 | 9 | 0.0007 | |
| Omron M6 | G | 11 | 10 | 9 | 10 | 11 | 10 | 10 | 10 | 9 | 10 | 0.0464 |
| H | 10 | 8 | 12 | 8 | 9 | 11 | 9 | 13 | 12 | 8 | 0.2126 | |
| Other Omron models | I | 10 | 8 | 12 | 9 | 11 | 8 | 9 | 9 | 11 | 12 | 0.4071 |
| J | 11 | 10 | 12 | 10 | 9 | 9 | 9 | 10 | 10 | 10 | 0.1308 | |
| Other brands | K | 9 | 11 | 9 | 12 | 11 | 10 | 13 | 7 | 9 | 9 | 0.1753 |
| L | 10 | 10 | 13 | 10 | 9 | 10 | 8 | 9 | 9 | 11 | 0.5258 | |
Proportion of terminal digits for diastolic blood pressure
|
|
|
|
| |||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|
|
|
|
|
|
|
|
|
|
|
| |||
| Mercury sphygmomanometer | A | 21 | 0 | 19 | 0 | 26 | 0 | 18 | 0 | 16 | 0 | 0.0392 |
| B | 35 | 8 | 8 | 5 | 7 | 17 | 6 | 5 | 6 | 4 | <0.0001 | |
| C | 22 | 0 | 17 | 0 | 19 | 0 | 19 | 0 | 23 | 0 | 0.0317 | |
| Omron i-C10 | D | 15 | 12 | 9 | 9 | 8 | 9 | 10 | 10 | 8 | 9 | 0.0122 |
| E | 11 | 9 | 9 | 9 | 11 | 10 | 9 | 12 | 10 | 11 | 0.2596 | |
| F | 11 | 10 | 7 | 10 | 9 | 12 | 11 | 10 | 9 | 10 | 0.4936 | |
| Omron M6 | G | 10 | 9 | 10 | 10 | 10 | 10 | 10 | 10 | 10 | 10 | 0.7859 |
| H | 9 | 9 | 10 | 12 | 10 | 10 | 10 | 11 | 10 | 9 | 0.9850 | |
| Other Omron models | I | 12 | 12 | 10 | 11 | 12 | 9 | 9 | 9 | 9 | 8 | 0.1123 |
| J | 11 | 9 | 10 | 11 | 9 | 10 | 10 | 11 | 10 | 9 | 0.4148 | |
| Other brands | K | 11 | 11 | 10 | 12 | 9 | 10 | 8 | 8 | 9 | 12 | 0.2285 |
| L | 11 | 9 | 11 | 9 | 11 | 9 | 9 | 11 | 11 | 9 | 0.9030 | |