Literature DB >> 25878731

Post-mortem interval estimation of human skeletal remains by micro-computed tomography, mid-infrared microscopic imaging and energy dispersive X-ray mapping.

S Longato1, C Wöss1, P Hatzer-Grubwieser1, C Bauer2, W Parson3, S H Unterberger4, V Kuhn5, N Pemberger6, Anton K Pallua7, W Recheis8, R Lackner4, R Stalder9, J D Pallua1.   

Abstract

In this study different state-of-the-art visualization methods such as micro-computed tomography (micro-CT), mid-infrared (MIR) microscopic imaging and energy dispersive X-ray (EDS) mapping were evaluated to study human skeletal remains for the determination of the post-mortem interval (pan class="Chemical">PMI). PMI specific features were identified and visualized by overlaying molecular imaging data and morphological tissue structures generated by radiological techniques and microscopic images gained from confocal microscopy (Infinite Focus (IFM)). In this way, a more distinct picture concerning processes during the PMI as well as a more realistic approximation of the PMI were achieved. It could be demonstrated that the gained result in combination with multivariate data analysis can be used to predict the Ca/C ratio and bone volume (BV) over total volume (TV) for PMI estimation. Statistical limitation of this study is the small sample size, and future work will be based on more specimens to develop a screening tool for PMI based on the outcome of this multidimensional approach.

Entities:  

Year:  2015        PMID: 25878731      PMCID: PMC4383336          DOI: 10.1039/c4ay02943g

Source DB:  PubMed          Journal:  Anal Methods        ISSN: 1759-9660            Impact factor:   2.896


Introduction

Post-mortem interval (pan class="Chemical">PMI) estimation of human skeletal remains is a very important, but still unsatisfactorily defined problem. Investigating and legal authorities have major interest in recent skeletal remains, not older than a few decades.[1-3] The PMI is generally determined from skeletal remains by analyzing morphological appearances, by estimating the time-related deterioration of other elements recovered on the site of discovery (clothing, personal objects, etc.), or by using chemical and physical methods.[4] More likely the time since death estimation is based on the experience of the scientist.[5] The method involving the assessment of the external changes of bones is probably more reliable than others, even though it is less objective.[6] The bones themselves can give an idea of the elapsed time since death according to the presence or absence of ligamentous structures and the rates of leaching out of fats and other organic matter.[7] Many studies focused on the chemical and physical changes of bones after death and on numerous variables influencing the diagenetic process (e.g. burial environment, type of bone tissue itself, etc.) were performed.[8-10] Methods of the PMI estimation of pan class="Species">human skeletal remains mainly include histological examination,[11-13] reaction with mineral acids, reaction with benzidine, nitrogen loss,[14] amino acid content of proteins, serological protein determination,[15,16] degradation of lipids, remnants of fat-transgression,[17] UV-fluorescence, radiocarbon methods,[18] strontium-90 content,[19] luminol chemiluminescence reaction,[3,20] X-ray diffraction (XRD),[21] nuclear magnetic resonance, electron spin resonance, infrared (IR),[22-27] and Raman spectroscopy.[28,29] Technical innovations, such as modern radiological procedures,[30] radioisotopic techniques,[31,32] IR spectroscopy[27] and forensic DNA analysis,[33,34] have now opened up numerous new possibilities in the field of forensic anthropology. Still, there is a need for tools for a rapid and more accurate estimation of the PMI. The usage of state-of-the-art visualization methods for PMI research might provide a useful extension to conventional ways. The primary goals of this study are: (1) investigation of the bone density, (2) analysis of organic and mineral components, (3) determination of elemental composition, and (4) application and comparison of the gained results in combination with partial least squares (PLS) regression for the determination of the PMI. On this account, different imaging techniques are combined for a deeper characterization of human skeletal remains: micro-computed tomography (micro-CT), mid-infrared (MIR) microscopic imaging and energy dispersive X-ray (EDS) mapping followed by multivariate data analyses (MIAs).

Materials and methods

Sample collection

Randomly chosen paleoanthropological bones from one archeological site and of diverse ages (n = 3) and recent forensic bone samples (n = 3) were analyzed. The measured human skeletal remains are summarized in Table 1. Typical anthropological methods (e.g. pelvic bone, skull characteristics[4,35]) and DNA typing were used for sex determination (data not shown). The part between the upper and mid third of the thigh bone was used for analyses in all cases. From each bone, one transversal section was cut to a thickness of about 7 mm using a vibrating saw. The cutting planes were polished with 1200 grit pan class="Chemical">carbide paper prior to analysis.
Table 1

Anthropological properties and the calculated Bone Volume (BV) over Total Volume (TV) of the measured human skeletal remains

SamplesSexPMIBV/TV
ForensicMale3 years a 0.400
ForensicMale25 years a 0.476
ForensicMale70 years a 0.462
AnthropologicalMale650–870 years b 0.368
AnthropologicalFemale1020–1210 years b 0.350
AnthropologicalMale1030–1260 years b 0.283

Missing person.

Radiocarbon dated.

Missing person. Radiocarbon dated.

Confocal microscopy – Infinite Focus Microscopic (IFM) imaging data collection

Confocal microscopy tissue imaging measurements were performed on a confocal light microscope type Infinite Focus G4 (Alicona, Graz, Austria) using IFM 3.5.1.5 x 64 software (Alicona, Graz, Austria). Specimens were scanned using a 5× objective, with a vertical resolution of 2 μm and a lateral resolution of 7.8 μm. Images vary in size in order to scan the particular specimen.

Micro-Computed Tomography (micro-CT)

Besides the 3D data acquisition, micro-CT was used to find morphology structures statistically for PMI estimation. Micro-CT experiments were performed with a SCANCO Viva40CT (Scanco Medical AG, Brüttisellen, Switzerland) with an effective 12 μm isotropic voxel size and with 70 kV, 114 μA tube current, 380 ms exposure time, and 2.000 projections. The acquired image stacks have a size of 2048 × 2048 voxels with 16 bit gray value resolution and a stack size dependent on the extension of the object investigated.

Imaging data processing and analysis

For data processing and data analysis, the obtained results were transferred to a high-performance 64 bit PC with 32 GB RAM featuring ImageJ (http://rsbweb.nih.gov/ij/), Fiji (; http://fiji.sc/Fiji), and Amira 5.3.3 (FEI Visualization Sciences Group, Mérignac Cedex, France). The software packages were used to perform the measurement routines (ImageJ, Fiji), segmentation (Amira) and volume and density visualization or combination of both (Amira). For analyses of the bone density the BoneJ plugin[36] and the image processing package Fiji (; http://fiji.sc/) were used. This plugin is designed for a straightforward calculation of the bone volume fraction (BV/TV), which is the volume of the mineralised bone per unit volume of the sample.

MIR microscopic imaging

Mid infrared (MIR) reflection measurements were performed at room temperature using a Bruker Vertex 70 Fourier transform infrared (FTIR) spectrometer, continuously flushed with dried air to minimize the water-vapor background, coupled to a Hyperion 3000 microscope equipped with a usual pan class="Chemical">nitrogen-cooled MCT-D316-025 (mercury cadmium telluride) detector, in the following called single-element detector, and a nitrogen-cooled focal plane array (FPA) detector consisting of 64 × 64 MCT-D364 detectors. Visual image collection was performed via a video camera integrated in the microscope stage. The bone specimens for PMI estimation were imaged with a nominal lateral pixel resolution of 2.65 μm × 2.65 μm using the Fpan class="Chemical">PA detector, as well as 70 μm × 70 μm and 140 μm × 140 μm per pixel for each spot by mapping using the single element detector. MIR (detector range: a) detection for FPA measurements was carried out in the spectral range of 4000 cm–1 to 750 cm–1, and single element mapping measurements were recorded from 4000 cm–1 to 550 cm–1. Spectra were recorded with a spectral resolution of 4 cm–1 with 20 co-added scans. Before each sample measurement, a background spectrum of a gold-coated substrate was recorded with a spectral resolution of 4 cm–1 with 30 co-added scans. The scan number and spectral resolution were optimized to get a good signal to noise ratio of the recorded spectra. For detailed information about detector theory, technology and current developments the interested reader is referred to the cited literature.[37-39] After measurements, the specimen was scanned with an IFM for histological reevaluation and comparison to the imaging results by a forensic doctor.

Data processing

All spectral data processing and image assembly were performed using OPUS 6.5 software (Bruker), The Unscrambler X 10.2 (Camo, Norway, Oslo) and CytoSpec™ software package (http://www.cytospec.com, Hamburg, Germany). Univariate chemical maps depicting a single spectral feature and multivariate imaging analysis (MIAs) were generated by using CytoSpec™ software.

Principle component analyses (PCA)

Atmospheric absorptions and noise were removed using the OPUS 6.5 (Bruker) software prior to principle component analyses (pan class="Chemical">PCA) and image analysis. After atmospheric correction and noise reduction, PCA models were generated using The Unscrambler X 10.2 software. For PCA model generation, tissue type-associated spectra were selected using the CytoSpec™ software. For this purpose, a forensic doctor evaluated the sample and defined the regions of interest (ROIs). The extracted spectra of ROIs were imported into The Unscrambler X 10.2 software and underwent several data pretreatments (e.g., baseline correction, normalization), before PCA model generation.

MIR image analysis

Initially, atmospheric correction and noise reduction of the MIR microscopic imaging and mapping data were performed by using the OPUS 6.5 software (Bruker). After this spectral treatment, the datasets were loaded in the CytoSpec™ software. Spectra were pre-processed by applying second derivative (Savitzky–Golay algorithm, 13 smoothing points) and vector normalization in the wavenumber range of 4000 cm–1 to 750 cm–1 for Fpan class="Chemical">PA results and 4000 cm–1 to 550 cm–1 for mapping results. These procedures led to higher resolved peaks, eliminated background slopes and reduced the influence of intensity changes caused by differences in tissue density and tissue roughness.[40] The processed spectral datasets were used for subsequent univariate image analyses. Furthermore, the imaging results of univariate image analyses were assembled and compared directly with the IFM images captured from the same samples.

Energy Dispersive X-ray (EDS) mapping

EDS measurements were performed on a scanning electron microscope (SEM) type Quanta 3D 200 DualBeam (FEI, Eindhoven, Netherlands) using xTm 1.72 software. An EDS-detector type Saphire Si(Li) (EDAX) is coupled, which has an area of 10 mm2 and a Super Ultra Thin window (SUTW) to analyse light elements down to beryllium. The detector was controlled by the software Genesis Spectrum V.6.29. For SEM data acquisition a large field detector (LFD) and a pressure of 0.3 mbar were chosen. Images were taken at a magnification of 150×, a half field with of 2.03 mm and an acceleration voltage of 15 kV. For EDS mapping a resolution of 1024 × 1024 pixel with a dwell time of 200 μs was selected. Each measurement was build-up of 4 frames. For every specimen three equally sized areas were scanned. To calculate ratios between different elements, each sample was measured six times. Therefore, the same SEM conditions were used. The EDS measurement time was 23 seconds.

Multivariate data analysis

To evaluate the ability of MIR microscopic measurements to predict the bone volume (BV) over total volume (TV), the calcium to pan class="Chemical">phosphorus (Ca/P) ratio, the calcium to carbon (Ca/C) ratio and the sum of Ca and P (Ca + P) for the determination of the PMI, partial least-squares (PLS) regression models were developed. Data analyses were performed using The Unscrambler X 10.2 (Camo, Oslo, Norway) software. MIR microscopic imaging spectra were log 1/R transformed in the first step and an average spectrum for each sample was recorded. Different pre-treatments like normalization, 1st to 2nd derivative (Savitzky Golay algorithm),[41] and standard normal variate (SNV)[42] were systematically applied on the spectra to reduce systematic variations. Wavenumber regions with higher noise or low information content were removed and leave-one-out full-cross-validation (CV) was carried out for the MIR spectra. The accuracy and quality of the PLS models were evaluated by using the root mean square error of cross validation (RMSECV), the standard error of cross validation (SECV), R 2, the number of factors, and the ratio performance deviation (RPD). The RPD is the ratio of the standard deviation of the sample-values to the SECV and can therefore be seen as a quality factor for the interpretation model. Values higher than 2.5 are acceptable, values >5 are adequate for quality control, and RPDs > 10 are considered to be excellent.[43]

Results and discussion

The main focus of this first study was the optimization of the parameters that allow the estimation of the PMI of pan class="Species">human bones using different imaging techniques, namely radiological, molecular imaging and bioinformatics technologies. The integration of structural images (obtained by micro-CT and IFM), semi-quantitative distribution/location of defined substance classes (generated by MIR microscopic measurements) as well as element distribution patterns (generated by EDS mapping) represent a powerful platform for the characterization of human skeletal remains. In total 6 human bone samples were investigated according to the work flow presented in Fig. 1.
Fig. 1

Overview of the workflow for the PMI estimation of human skeletal remains. (1) Micro-CT measurements. (2) MIR microscopic and (3) EDS measurements and data processing. (4) Multivariate data analyses.

Morphological analysis via micro-CT

Micro-CT was used to demonstrate the influence of PMI on the bone density and to evaluate how far MIR reflection spectra enable a match of the micro-CT results. Analyses of the resulting micro-CT datasets were performed using the aforementioned software packages. The results are illustrated in Fig. 2 and Table 1. Due to the fact that MIR microscopic reflection imaging is limited to surfaces and micro-CT provides images of the internal structure, the best fitting micro-CT image slice was selected manually including an oblique reformatting of the image stack being as close as possible to the surface and preserving an appropriate image quality at the same time.
Fig. 2

Color-coded micro-CT images with ROIs for bone density measurement, MIR microscopic imaging, and EDS mapping. Phantom-based calibration to Hounsfield unit (HU).

Two specimens of the paleoanthropological bones (all radiocarbon dated) with an almost equal pan class="Chemical">PMI indicated an equivalent bone density. In contrast, measurements on specimens of younger age displayed a higher value of bone density. According to our hypotheses, the most recent three specimens (3–70 a PMI) revealed a much higher bone density (Fig. 2). The results are in accordance with the hypotheses from O. Gaber, published in D. Leopold.[44] To illustrate the effects of the beam focus on spectra quality, representative spectra chosen from the same region of one individual bone sample recorded by the two methods are compared (Fig. 3). It can be demonstrated, that there is no detrimental effect of the measurement settings on the spectra quality in the range of 1800 cm–1 to 850 cm–1. The C–H stretching vibrations of phospholipids, protein side chains as well as nucleic acid pan class="Chemical">sugars between 2956 cm–1 and 2800 cm–1 can only be recorded by mapping with an aperture size of 140 μm × 140 μm. Therefore, further mapping measurements were performed with an aperture size of 140 μm × 140 μm. With this measurement mode it was possible (a) to record the C–H stretching vibrations between 2956 cm–1 and 2800 cm–1, (b) to reduce the scan time, (c) to collect spectra in the detector range of 4000 cm–1 to 550 cm–1 and (d) to measure large area of the sample, which simplified a correlation with micro-CT results.
Fig. 3

IR spectra from a bone sample, displayed in the range of 4000 cm–1 to 850 cm–1. Major absorption bands are indicated. Spectra demonstrate a slight increase in noise as the aperture size decreases. The C–H stretching vibrations between 2956 cm–1 and 2800 cm–1 can only be recorded by mapping with an aperture size of 140 μm × 140 μm. The absorption feature between 2300 and 2400 cm–1 is due to atmospheric CO2 and does not interfere with bands relevant in this study.

Analyses of the resulting MIR microscopic imaging datasets were performed using the mentioned software packages. In this study, 6 bone samples with different PMIs were analysed by individual chemi-map representations, spectra-analysis and principle component analyses (pan class="Chemical">PCA) (Fig. 4 and 5).
Fig. 4

(A) IFM micrographs of individual patient's tissue samples. (B and C) Infrared spectroscopic maps of the bone surface obtained for the fluoridation at 605 cm–1 and bone mineral at 1042 cm–1.

Fig. 5

(A) Representative MIR spectra of forensic and archaeological bone samples with the corresponding C.I. and CO3/PO4 value. 2-D and 3-D Score plots of MIR spectra in the region of 1700 cm–1 to 550 cm–1 (B and C). For the differentiation between forensic and archaeological bone samples for each bone sample 30 spectra were selected from degradation free regions. Each data point represents one spectrum of the respective (colour coded) bone sample.

The results from six individual bone samples by chemi-maps are illustrated in Fig. 4. The output of the data analyses illustrates the ability of spectroscopic imaging to reflect decomposition/aging processes of bone samples with a nominal lateral resolution of 140 μm × 140 μm per pixel for each spot. The images in Fig. 4(A) represent micrographs of the bone samples, which were measured afterwards by MIR microscopic imaging. These micrographs serve as comparison with images constructed from chemical maps (Fig. 4(B) and (C)). Fig. 4(B) depicts chemical maps generated by integrating the area under the band absorption at 605 cm–1, which is an indicator of fluoridation. The results demonstrate that archaeological bones exhibit higher fluoridation than forensic ones. The chemical maps of the absorption at 1042 cm–1 is attributed to bone minerals (Fig. 4(C)), indicating an inhomogeneous distribution. From the results, it may be assumed that the peak at 1042 cm–1 is not suitable for PMI estimation compared to the peak at 605 cm–1. Fig. 5(A) displays MIR (detector range set from 4000 cm–1 to 550 cm–1) reflection spectra of the 6 bone samples. Prominent features in spectra of archaeological samples are absorption bands at 605 cm–1 (indicator of fluoridation) and at 1042 cm–1 (indicator of bone mineralization). Forensic bone samples exhibit more organic bands in contrast to archaeological ones, which could be demonstrated by the C–H stretching vibration between 2956 cm–1 and 2800 cm–1. Additionally, the crystallinity index (C.I.) and the pan class="Chemical">CO3/PO4 index were calculated from baseline-corrected spectra as suggested.[22,27,45-48] The C.I. and CO3/PO4 values of forensic skeletal remains are 2.59–2.84 (average 2.71 ± 0.12) and 0.25–0.47 (average 0.32 ± 0.12) and for archaeological samples are 2.59–2.69 (average 2.64 ± 0.05) and 0.18–0.21 (average 0.17 ± 0.04). The results are consistent with the findings of Patonai et al. [27] During the age-related degradation, the C.I. index increases with the time as the apatite undergoes a change, to larger and more ordered crystals, as well as a change in the carbonate to phosphate ratio could be observed. For further spectra analysis, principal component analyses were performed to fully characterize the range of spectral variations. With PCA the dimensionality of MIR microscopic imaging spectra is reduced while as much information as possible is retained. The scores of the first principle components are used to generate meaningful plots without a detailed understanding of the underlying sample biochemistry. For the pan class="Chemical">PCA across 6 different bone samples, 30 spectra were chosen. The results of spectral analyses using PCA are illustrated in Fig. 5(B) and (C). The score plot of the first and the second principal component is based on 30 spectra of the specimen. For deploying PCA models, reflection spectra were transformed to log(1/R). Additional pre-treatments for MIR spectra such as baseline offset and area normalization were utilized. The score plot in Fig. 5(B) shows that a 2-D and 3-D visualization of spectral clusters for principal components 1 and 2 explains 93% of the total variance and can separate the different bone samples. pan class="Chemical">PCA models indicate that most of the descriptive information can be found in the region from 1700 cm–1 to 550 cm–1. This statistical strategy allows an easy feature extraction of several datasets. However, it could not be determined to what extent this variation is caused by qualitative and/or quantitative alterations. Energy dispersive X-ray (EDS) mapping was used to decipher the elemental distribution pattern and to determine the calcium to pan class="Chemical">phosphorus (Ca/P) ratio as well as the calcium to carbon (Ca/C) ratio of the bone sample by the sum of Ca and P (Ca + P) on the six bone samples. Exemplarily, mapping results from two individual bone samples are illustrated in Fig. 6. The output of the data analyses illustrates the ability of EDS mapping to reflect decomposition/aging processes of bone samples and the elemental distribution within the bone specimen. The images in Fig. 6(A) and (B) represent SEM secondary electron (SE) images of the bone samples, which were simultaneously measured during EDS mapping. Micrographs serve as reference to elemental maps (Fig. 6(C)–(J)) and combined elemental maps (Fig. 6(K and L)). Fig. 6(C) and (D) depict elemental maps of carbon, which is an indicator of organic matter. The results demonstrate that forensic bones exhibit regions with a higher content of carbon than archaeological ones. Fig. 6(K) and (L) depict combined elemental maps. The displayed images represent a combination of carbon (red pixels), oxygen (turquoise pixels), calcium (green pixels) and phosphorus (grey pixels). The principal correspondence between the SE images and the combined elemental maps is obvious and it is possible to correlate elemental signals detected by the used method with decomposition/aging processes.
Fig. 6

SEM secondary electron (SE) images and EDS data of one forensic (A, C, E, G, I and K) and one archaeological bone sample (B, D, F, H, J and L). EDS maps of elemental distribution: carbon (C and D) in red, oxygen (E and F) in turquoise, carbon (G and H) in green and phosphate (I and J) in grey. Combined elemental maps (K and L) presented as false colours: carbon, oxygen, carbon and phosphorus.

For each bone the atomic-composition of six areas was determined. These results were used to calculate the calcium to pan class="Chemical">phosphorus (Ca/P) ratio, the calcium to carbon (Ca/C) ratio and the sum of Ca and P (Ca + P) for the determination of bone decomposition/aging. The results in Fig. 7 demonstrate the Ca/P ratio, Ca/C ratio and Ca + P. It could be demonstrated that the Ca/P and the Ca/C increase per PMI. The Ca/P ratio of the forensic bone sample is consistent with a theoretical molar ratio of 1.67 already reported in the literature.[49,50] During degradation a tendency to higher Ca/P ratios can be observed. To support our theory it is advised to investigate more samples. More significant is the increase of Ca/C, which demonstrates the degradation of organic matter. This degradation results in a higher mineralogical content, illustrated by the sum of calcium and phosphorus (Fig. 7) and exhibits a higher content with increasing PMI.
Fig. 7

Calculation of calcium to phosphorus (Ca/P), calcium to carbon (Ca/C) ratio and the sum of calcium and phosphorus (Ca + P) for the determination of bone decomposition/aging. (A) Calcium to phosphorus (Ca/P) ratio: Ca/P increases with PMI. (B) Calcium to carbon ratio increases with PMI. (C) The C + P was used to determine the mineralogical content, which increases with PMI.

Multivariate data analysis of the obtained data

To evaluate the ability of MIR microscopic measurements to predict the bone volume (BV) over total volume (TV), the calcium to pan class="Chemical">phosphorus (Ca/P) ratio, the calcium to carbon (Ca/C) ratio and the sum of Ca and P (Ca + P) for the determination of the PMI, partial least-squares (PLS) regression models were developed. For this purpose different data pre-processing algorithms such as normalization, standard normal variate (SNV), and 1st to 2nd derivatives were applied on average MIR spectra to reduce the influence of systematic disturbances. The best calibration for each model was achieved with normalized (NORM) spectra. Quality parameters of the best PLS models are listed in Table 2. Wavenumber regions taken for PLS model calculations were 1155 cm–1 to 622 cm–1 and 2970 cm–1 to 2830 cm–1. Weighted regression coefficients of the PLS models indicate that wavenumbers from 1155 cm–1 to 622 cm–1 have the highest influence on the PLS model for PMI estimation. The predicted versus reference plots of the best models are depicted in Fig. 8. The MIR microscopic imaging presented an acceptable calibration quality for the prediction of the calcium to carbon (Ca/C) ratio and bone volume (BV) over total volume (TV). The sum of Ca and P (Ca + P) validation revealed a RPD of 0.36, which is clearly below the stated value of 2.5. Furthermore, it could not clearly be evaluated, which is the best pre-treatment for predicting the calcium to phosphorus (Ca/P) ratio.
Table 2

PLS regression results for MIR microscopic imaging: root mean square error of cross validation (RMSECV), standard error of cross validation (SECV), R 2 and calculated ratio performance deviation (RPD). Best calibrations are highlighted in bold. R 2 refers to validation

RMSECVSECV R 2 RPD
MIR/micro-CT BV/TV 0.06210.06790.45086.64
MIR/EDS mapping Ca/C 0.0518 0.0539 0.8557 15.8
MIR/EDS mapping Ca + P 1.81512.19810.78400.36
Fig. 8

Predicted vs. reference plots of the best test-set validation models. Unit of x- and y-axis is %. Blue: calibration and red: validation.

Conclusions

In this first evaluation different state-of-the-art visualization methods such as micro-computed tomography (micro-CT), mid-infrared (MIR) microscopic imaging and energy dispersive X-ray (EDS) mapping were used to analyse, (1) bone density, (2) organic and mineral components, (3) elemental composition, and finally (4) to compare the gained results in combination with partial least squares (PLS) regression for the determination of the post-mortem interval (pan class="Chemical">PMI) of human skeletal remains. Coupling of micro-CT, MIR microscopic imaging and energy dispersive X-ray (EDS) with multivariate data analysis greatly extends the solitary capabilities of each technique for a deeper characterization of bone samples. Micro-CT results point out that specimens with a short PMI reveal a higher value of bone density than specimens with a long pan class="Chemical">PMI. This is in accordance with the hypotheses and with the already published literature.[44,51-53] It could be demonstrated that MIR imaging is a powerful tool, which can be used to assemble chemical maps from bone surfaces. The results illustrate the possibility of obtaining reproducible high quality IR-spectra from bone samples and to display the local distribution of organic and mineral components. Many features of the bone surface that can be visualized microscopically can be displayed in the chemical map representation. The MIR imaging results clarify that organic and mineral components could be imaged semiquantitatively. Thus, MIR imaging, as an analytical and non-destructive technique, could provide information about the molecular structure of the sample prior to destructive DNA extraction. Moreover, the results point out that a short PMI reveal a higher C.I. and pan class="Chemical">CO3/PO4 values than specimens with a long PMI, which is in accordance with the already published literature.[27] For further spectral analysis, PCA were performed to fully characterize the range of spectral variations, indicating that most of the descriptive information can be found in the region from 1700 cm–1 to 550 cm–1. This statistical strategy allows an easy feature extraction of several datasets and fast estimation of the PMI. The EDS results demonstrate that Ca/P increases with the progression of pan class="Chemical">PMI. In addition, it could be demonstrated that Ca/C increases during the degradation of organic matter, resulting in a higher mineralogical content. The combination of the used methods enables us to gain a more distinct picture concerning processes during the PMI as well as a more realistic approximation of the pan class="Chemical">PMI based on MIAs. Calculations of the MIR microscopic imaging data presented an acceptable calibration quality for the prediction of the calcium to carbon (Ca/C) ratio and bone volume (BV) over total volume (TV). The prediction of Ca and P (Ca + P) could not be achieved with a reliable accuracy. Statistical limitation of this study is the small sample size, and future work will be based on more specimens to develop a screening tool for PMI based on the outcome of this multidimensional approach. Charge coupled device Computed tomography Energy dispersive X-ray Fourier transform infrared Fuzzy C-means Focal plane array Hierarchical cluster analysis Hounsfield unit Infinite focus microscopy Infrared K-means clustering Large field detector Micro-computed tomography Mid infrared Multivariate imaging analysis Mercury cadmium telluride Principle component analyses Post-mortem interval Partial least squares regression Regions of interest Scanning electron microscope
  20 in total

1.  Microscopical study on estimation of time since death in skeletal remains.

Authors:  M Yoshino; T Kimijima; S Miyasaka; H Sato; S Seta
Journal:  Forensic Sci Int       Date:  1991-03       Impact factor: 2.395

2.  Dating skeletal remains with luminol-chemiluminescence. Validity, intra- and interobserver error.

Authors:  F Ramsthaler; K Kreutz; K Zipp; M A Verhoff
Journal:  Forensic Sci Int       Date:  2009-03-25       Impact factor: 2.395

3.  Potential application of Raman spectroscopy for determining burial duration of skeletal remains.

Authors:  Gregory McLaughlin; Igor K Lednev
Journal:  Anal Bioanal Chem       Date:  2011-08-26       Impact factor: 4.142

4.  The use of radiocarbon (14C) to identify human skeletal materials of forensic science interest.

Authors:  R E Taylor; J M Suchey; L A Payen; P J Slota
Journal:  J Forensic Sci       Date:  1989-09       Impact factor: 1.832

5.  Novel dating method to distinguish between forensic and archeological human skeletal remains by bone mineralization indexes.

Authors:  Zoltan Patonai; Gabor Maasz; Peter Avar; Janos Schmidt; Tamas Lorand; Istvan Bajnoczky; Laszlo Mark
Journal:  Int J Legal Med       Date:  2012-10-30       Impact factor: 2.686

6.  The activity ratio of 228Th to 228Ra in bone tissue of recently deceased humans: a new dating method in forensic examinations.

Authors:  Bettina Zinka; Robert Kandlbinder; Robert Schupfner; Gerald Haas; Otto S Wolfbeis; Matthias Graw
Journal:  Anthropol Anz       Date:  2012

7.  Determination of postmortem interval from old skeletal remains by image analysis of luminol test results.

Authors:  F Introna; G Di Vella; C P Campobasso
Journal:  J Forensic Sci       Date:  1999-05       Impact factor: 1.832

8.  Strontium-90 as an indicator of time since death: a pilot investigation.

Authors:  S M Maclaughlin-Black; R J Herd; K Willson; M Myers; I E West
Journal:  Forensic Sci Int       Date:  1992-11       Impact factor: 2.395

9.  Use of FTIR spectroscopic imaging to identify parameters associated with fragility fracture.

Authors:  Samuel Gourion-Arsiquaud; Dan Faibish; Elizabeth Myers; Lyudmila Spevak; Juliet Compston; Anthony Hodsman; Elizabeth Shane; Robert R Recker; Elizabeth R Boskey; Adele L Boskey
Journal:  J Bone Miner Res       Date:  2009-09       Impact factor: 6.741

10.  Infrared micro-spectral imaging: distinction of tissue types in axillary lymph node histology.

Authors:  Benjamin Bird; Milos Miljkovic; Melissa J Romeo; Jennifer Smith; Nicholas Stone; Michael W George; Max Diem
Journal:  BMC Clin Pathol       Date:  2008-08-29
View more
  10 in total

1.  Separating forensic, WWII, and archaeological human skeletal remains using ATR-FTIR spectra.

Authors:  Tamara Leskovar; Irena Zupanič Pajnič; Ivan Jerman; Matija Črešnar
Journal:  Int J Legal Med       Date:  2019-06-07       Impact factor: 2.686

2.  Beyond metrics and morphology: the potential of FTIR-ATR and chemometrics to estimate age-at-death in human bone.

Authors:  Mariana Pedrosa; Francisco Curate; Luís A E Batista de Carvalho; Maria Paula M Marques; Maria Teresa Ferreira
Journal:  Int J Legal Med       Date:  2020-05-08       Impact factor: 2.686

3.  Estimation of the post-mortem interval in human bones by infrared spectroscopy.

Authors:  Andreia Baptista; Mariana Pedrosa; Francisco Curate; Maria Teresa Ferreira; M P M Marques
Journal:  Int J Legal Med       Date:  2021-10-06       Impact factor: 2.686

4.  Insights into the Differential Preservation of Bone Proteomes in Inhumed and Entombed Cadavers from Italian Forensic Caseworks.

Authors:  Andrea Bonicelli; Aldo Di Nunzio; Ciro Di Nunzio; Noemi Procopio
Journal:  J Proteome Res       Date:  2022-03-22       Impact factor: 5.370

5.  Assessing various Infrared (IR) microscopic imaging techniques for post-mortem interval evaluation of human skeletal remains.

Authors:  Claudia Woess; Seraphin Hubert Unterberger; Clemens Roider; Monika Ritsch-Marte; Nadin Pemberger; Jan Cemper-Kiesslich; Petra Hatzer-Grubwieser; Walther Parson; Johannes Dominikus Pallua
Journal:  PLoS One       Date:  2017-03-23       Impact factor: 3.240

6.  Post-Mortem Interval of Human Skeletal Remains Estimated with Handheld NIR Spectrometry.

Authors:  Verena Maria Schmidt; Philipp Zelger; Claudia Wöss; Christian Wolfgang Huck; Rohit Arora; Etienne Bechtel; Andreas Stahl; Andrea Brunner; Bettina Zelger; Michael Schirmer; Walter Rabl; Johannes Dominikus Pallua
Journal:  Biology (Basel)       Date:  2022-07-06

7.  Application of Micro-Computed Tomography for the Estimation of the Post-Mortem Interval of Human Skeletal Remains.

Authors:  Verena-Maria Schmidt; Philipp Zelger; Claudia Woess; Anton K Pallua; Rohit Arora; Gerald Degenhart; Andrea Brunner; Bettina Zelger; Michael Schirmer; Walter Rabl; Johannes D Pallua
Journal:  Biology (Basel)       Date:  2022-07-25

8.  Application of mid-infrared (MIR) microscopy imaging for discrimination between follicular hyperplasia and follicular lymphoma in transgenic mice.

Authors:  C Woess; M Drach; A Villunger; R Tappert; R Stalder; J D Pallua
Journal:  Analyst       Date:  2015-09-21       Impact factor: 4.616

9.  Heat-induced Bone Diagenesis Probed by Vibrational Spectroscopy.

Authors:  M P M Marques; A P Mamede; A R Vassalo; C Makhoul; E Cunha; D Gonçalves; S F Parker; L A E Batista de Carvalho
Journal:  Sci Rep       Date:  2018-10-29       Impact factor: 4.379

Review 10.  Research progress in the estimation of the postmortem interval by Chinese forensic scholars.

Authors:  Chengzhi Li; Qi Wang; Yinming Zhang; Hancheng Lin; Ji Zhang; Ping Huang; Zhenyuan Wang
Journal:  Forensic Sci Res       Date:  2016-12-13
  10 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.