Repeatability and comparison of clinical techniques for anterior chamber angle assessment.

Peter Campbell1, Tony Redmond, Rishi Agarwal, Lewis R Marshall, Bruce J W Evans.   

Abstract

PURPOSE: To assess the repeatability of gonioscopy, van Herick method and anterior segment Optical Coherence Tomography (AS-OCT) and determine the agreement between these techniques within a community optometry setting.
METHODS: Gonioscopy, van Herick method and AS-OCT imaging were performed by an optometrist on two occasions, 1 month apart, on 80 subjects aged over 40 years recruited from community optometry practices. Anterior segment images were captured with a spectral domain OCT (Topcon 3D OCT-2000; wavelength 840 nm) set to the Anterior Segment (AS) mode. Eyes were graded as open or occludable for each method. AS-OCT images from both visits were graded by a second optometrist masked to the gonioscopy and van Herick method results, and the visit on which the images were acquired. Cohen's kappa (κ) was used to describe the intra-observer repeatability. Likelihood ratios, sensitivity and specificity of van Herick method and AS-OCT were calculated, using gonioscopy as the reference standard.
RESULTS: Measurements were obtained from 80 eyes of 80 subjects. In four cases, AS-OCT images were un-gradable due to difficulty in locating the scleral spur. The repeatability of gonioscopy was fair κ = 0.29, while that of the van Herick method (κ = 0.54) and AS-OCT (κ = 0.47) were somewhat better. The van Herick method showed good sensitivity (visit 1: 75%, visit 2: 69%) and high specificity (visit 1: 88%, visit 2: 96%). The sensitivity of AS-OCT was fair (visit 1: 46%, visit 2: 25%), but specificity was high (visit 1: 87%. visit 2: 89%).
CONCLUSION: Intra-observer repeatability was better for van Herick method and for AS-OCT than for gonioscopy, despite the latter being considered the gold standard method. The van Herick method appeared to be more sensitive than AS-OCT when identifying eyes at risk of angle closure. A standalone anterior segment OCT with a longer wavelength laser could afford better visualisation of the angle, and might therefore be expected to enable the examiner to make more precise classifications. These instruments are not widely used by optometrists in clinical practice in the UK at present.
© 2015 The Authors Ophthalmic & Physiological Optics © 2015 The College of Optometrists.

Entities:  

Keywords:  Gonioscopy; Van Herick; anterior chamber; anterior segment; anterior segment optical coherence tomography; imaging

Mesh:

Year:  2015        PMID: 25761580     DOI: 10.1111/opo.12200

Source DB:  PubMed          Journal:  Ophthalmic Physiol Opt        ISSN: 0275-5408            Impact factor:   3.117


  9 in total

1.  A technician-delivered 'virtual clinic' for triaging low-risk glaucoma referrals.

Authors:  A Kotecha; J Brookes; P J Foster
Journal:  Eye (Lond)       Date:  2017-02-17       Impact factor: 3.775

2.  Anterior Segment Optical Coherence Tomography: Applications for Clinical Care and Scientific Research.

Authors:  Jing Shan; Charles DeBoer; Benjamin Y Xu
Journal:  Asia Pac J Ophthalmol (Phila)       Date:  2019-04-25

3.  Anterior segment biometric measurements explain misclassifications by a deep learning classifier for detecting gonioscopic angle closure.

Authors:  Alice Shen; Michael Chiang; Anmol A Pardeshi; Roberta McKean-Cowdin; Rohit Varma; Benjamin Y Xu
Journal:  Br J Ophthalmol       Date:  2021-10-06       Impact factor: 4.638

4.  Non-contact tests for identifying people at risk of primary angle closure glaucoma.

Authors:  Anish Jindal; Irene Ctori; Gianni Virgili; Ersilia Lucenteforte; John G Lawrenson
Journal:  Cochrane Database Syst Rev       Date:  2020-05-28

5.  Agreement between Gonioscopic Examination and Swept Source Fourier Domain Anterior Segment Optical Coherence Tomography Imaging.

Authors:  Mohammed Rigi; Nicholas P Bell; David A Lee; Laura A Baker; Alice Z Chuang; Donna Nguyen; Vandana R Minnal; Robert M Feldman; Lauren S Blieden
Journal:  J Ophthalmol       Date:  2016-11-20       Impact factor: 1.909

Review 6.  Anterior Chamber Angle Assessment Techniques: A Review.

Authors:  Ivano Riva; Eleonora Micheletti; Francesco Oddone; Carlo Bruttini; Silvia Montescani; Giovanni De Angelis; Luigi Rovati; Robert N Weinreb; Luciano Quaranta
Journal:  J Clin Med       Date:  2020-11-25       Impact factor: 4.241

7.  Anterior segment optical coherence tomography imaging and ocular biometry in cataract patients with open angle glaucoma comorbidity.

Authors:  Giedre Pakuliene; Kirilas Zimarinas; Irena Nedzelskiene; Brent Siesky; Loreta Kuzmiene; Alon Harris; Ingrida Januleviciene
Journal:  BMC Ophthalmol       Date:  2021-03-08       Impact factor: 2.209

8.  Optical coherence tomography analysis of anterior segment parameters before and after laser peripheral iridotomy in primary angle-closure suspects by using CASIA2.

Authors:  Xiaoxiao Chen; Xiaolei Wang; Xinghuai Sun; Yuhong Chen; Yizhen Tang
Journal:  BMC Ophthalmol       Date:  2022-03-28       Impact factor: 2.209

9.  Remote Grading of the Anterior Chamber Angle Using Goniophotographs and Optical Coherence Tomography: Implications for Telemedicine or Virtual Clinics.

Authors:  Jack Phu; Henrietta Wang; Vincent Khou; Sophia Zhang; Michael Kalloniatis
Journal:  Transl Vis Sci Technol       Date:  2019-09-25       Impact factor: 3.283

  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.