Liesbeth Eloot1, Hubert Thierens1, Yves Taeymans2, Benny Drieghe2, Jan De Pooter2, Sylvie Van Peteghem2, Dimitri Buytaert1, Thomas Gijs1, Régine Lapere1, Klaus Bacher1.
Abstract
OBJECTIVES: The purpose of this study was to quantify the reduction in patient radiation dose during coronary angiography (CA) by a new X-ray technology, and to assess its impact on diagnostic image quality.Entities:
Keywords: coronary angiography; coronary artery disease; imaging; radiation dose
Mesh:
Year: 2015 PMID: 25754018 PMCID: PMC6680340 DOI: 10.1002/ccd.25913
Source DB: PubMed Journal: Catheter Cardiovasc Interv ISSN: 1522-1946 Impact factor: 2.692
Criteria for Image Quality Assessment
| Criterion | Description | Question |
|---|---|---|
| 1 | Rating of image resolution | How would you judge the sharpness of delineation? |
| 2 | Rating of image contrast | How would you judge the contrast with the background? |
| 3 | Rating of image noise | How would you judge the noise content? |
| 4 | General image quality score | How would you judge the overall image quality? |
Patient Exposure Data for the Reference (Room A) and the Novel Imaging System (Room B)
| Room A | Room B |
| |
|---|---|---|---|
| ( | ( | ||
| KAP for total procedure, | 47.4 (33.6–66.5) | 12.0 (7.59–20.1) | <0.001 |
| KAP total CA, | 41.7 (29.6–60.3) | 8.77 (6.33–17.6) | <0.001 |
| KAP fluoro, Gycm2 | 4.62 (3.02–9.18) | 3.22 (2.00–6.62) | 0.064 |
| KAP cineangiography, Gycm2 | 41.5 (28.8–55.3) | 9.48 (5.68–14.0) | <0.001 |
| Cine runs, | 11 (10–12) | 11 (10–12) | 0.468 |
| Fluoroscopy time, s | 114 (72–196) | 172 (114–301) | 0.015 |
| CMC, | 115 (104–130) | 119 (99.7–136) | 0.492 |
| KAP LCA Lao0° Cran0°, Gycm2 | 1.57 (1.25–1.96) | 0.260 (0.169‐ 0.435) | <0.001 |
| KAP LCA Lao90° Cran0°, Gycm2 | 2.99 (1.88–5.21) | 0.617 (0.379–1.16) | <0.001 |
| KAP RCA Lao90° Cran0°, Gycm2 | 2.65 (1.92–4.48) | 0.733 (0.492–1.23) | <0.001 |
| KAP LV Rao35° Cran0°, Gycm2 | 4.37 (3.11–5.95) | 2.57 (1.55–3.31) | <0.001 |
KAP: kermfa‐area product.
CA: coronafry angiography–here without the left ventriculogram.
CMC: contrast medium consumption. Values represent the median (interquartile range).
Figure 1Distribution of total KAP values with the reference system (Room A) and the new imaging system (Room B).
Figure 2A: Posterior‐anterior view of the LCA of a patient, acquired with the reference system (Room A). B: Posterior‐anterior view of the LCA of the same patient, acquired with the new imaging technology at 30% of the radiation dose (Room B). These images were acquired when the patient returned to the department for therapeutic intervention on a different day. Original moving images are available in online Supporting Information.
Figure 3A: Lateral view of the RCA of a patient, acquired with the reference system (Room A). B: Lateral view of the RCA of the same patient, acquired with the new imaging technology at 30% of the radiation dose (Room B). These images were acquired when the patient returned to the department for therapeutic intervention on a different day. Original moving images are available in online Supporting Information.
Score for Each Question of the Image Quality Assessment for LCA and RCA, Acquired With the Reference System (Room A) or the Novel System (Room B)
| Room A | Room B |
| |
|---|---|---|---|
| LCA, | 35 | 35 | |
| Q1: image resolution | 78 ± 7 | 75 ± 10 | 0.168 |
| Q2: image contrast | 82 ± 7 | 80 ± 9 | 0.512 |
| Q3: image noise | 84 ± 7 | 64 ± 9 | <0.001 |
| Q4: general IQ score | 80 ± 8 | 76 ± 11 | 0.172 |
| RCA, | 35 | 35 | |
| Q1: image resolution | 78 ± 9 | 75 ± 12 | 0.180 |
| Q2: image contrast | 78 ± 8 | 80 ± 10 | 0.293 |
| Q3: image noise | 83 ± 7 | 64 ± 11 | <0.001 |
| Q4: general IQ score | 76 ± 11 | 76 ± 12 | 0.877 |
Scores represent the mean (± standard deviation) of four readers, expressed as a percentage of the maximum possible score.
Figure 4Distribution of the mean image quality score (%) for the images of the LCA assessment. A: Scores for the first question (image detail) for the reference X‐ray system (Room A) and the new X‐ray system (Room B); B: scores for image contrast; C: scores for image noise; D: overall image quality scores.