| Literature DB >> 25741119 |
Bruno Beraldo Oliveira1, Marcio Alves de Oliveira2, Lucas Paixão1, Maria Helena Araújo Teixeira3, Maria do Socorro Nogueira4.
Abstract
OBJECTIVE: To evaluate the mean glandular dose with a solid state detector and the image quality in a direct radiography system, utilizing phantoms.Entities:
Keywords: Direct radiography; Image; Mean glandular dose
Year: 2014 PMID: 25741119 PMCID: PMC4341379 DOI: 10.1590/0100-3984.2013.1876
Source DB: PubMed Journal: Radiol Bras ISSN: 0100-3984
Figure 1Positioning of materials in the determination of exposure parameters. A: Breast compression paddle. B: PMMA spacers. C: PMMA plates.
Figure 2Solid-state detector positioning for measurement of Ki and HVL.
Figure 3CDMAM phantom positioning for evaluation of image quality.
Figure 4Phantom Mama positioning for evaluation of image quality.
Voltage, load, anode/filter combination for different thicknesses.
| PMMA plate thickness (mm) | Equivalent breast thickness (mm) | Fibroglandular tissue proportion of the breast equivalent (%) | Voltage (kV) | Load (mA.sCAE) | Anode/filter combination |
|---|---|---|---|---|---|
| 20 | 21 | 97 | 25 | 46 | W/Rh |
| 30 | 32 | 67 | 26 | 70 | W/Rh |
| 40 | 45 | 41 | 28 | 93 | W/Rh |
| 50 | 60 | 20 | 31 | 133 | W/Rh |
| 60 | 75 | 9 | 31 | 165 | W/Ag |
| 70 | 90 | 4 | 33 | 184 | W/Ag |
K and HVL values for different thicknesses.
| CSR (mmAl) | ||||
|---|---|---|---|---|
| PMMA plate thickness (mm) | Equivalent breast thickness (mm) | Calculated | Tolerance | |
| 20 | 21 | 1.078 | 0.50 | 0.28–0.55 |
| 30 | 32 | 1.900 | 0.51 | 0.29–0.56 |
| 40 | 45 | 3.150 | 0.54 | 0.31–0.58 |
| 50 | 60 | 5.996 | 0.56 | 0.34–0.61 |
| 60 | 75 | 7.675 | 0.62 | 0.34–0.63 |
| 70 | 90 | 9.532 | 0.65 | 0.36–0.65 |
Figure 5Mean glandular dose for different thicknesses.
Visual and automatic evaluation for different gold diameters utilizing the CDMAM.
| Gold thickness evaluations (μm) | |||||
|---|---|---|---|---|---|
| Diameter (mm) | Visual in the present study | Visual by Oliveira et al.( | Automatic – present study | Automatic by Oliveira et al.( | Reference value |
| 0.10 | 2.00 | 2.00 | 0.995 | 2.00 | 1.680 |
| 0.25 | 0.36 | 0.36 | 0.142 | 0.16 | 0.352 |
| 0.50 | 0.16 | 0.16 | 0.064 | 0.10 | 0.150 |
| 1.00 | 0.10 | 0.08 | 0.037 | 0.03 | 0.091 |
| 2.00 | 0.06 | 0.06 | 0.040 | 0.05 | 0.069 |
Number of visualized structures utilizing the Phantom Mama.
| Structures (total) | Visualized structures | Tolerance |
|---|---|---|
| Fibers (6) | 4 | ≥ 4 |
| Discs (8) | 7 | ≥ 7 |
| Tumor masses (5) | 4 | ≥ 4 |
| Microcalcifications (5) | 4 | ≥ 4 |
| Metal grids (4) | 3 | 3 |