Tiago da Silva Jornada1, Teógenes Augusto da Silva2. 1. Master, PhD student in Clinical Radiology, Department of Imaging Diagnosis - Escola Paulista de Medicina da Universidade Federal de São Paulo (EPM-Unifesp), São Paulo, SP, Brazil. 2. PhD, Senior Researcher, Comissão Nacional de Energia Nuclear (CNEN), Teacher of the Post-graduation Course at Centro de Desenvolvimento da Tecnologia Nuclear (CDTN/CNEN), Belo Horizonte, MG, Brazil.
Abstract
OBJECTIVE: Aiming at contributing to the knowledge on doses in computed tomography (CT), this study has the objective of determining dosimetric quantities associated with pediatric abdominal CT scans, comparing the data with diagnostic reference levels (DRL). MATERIALS AND METHODS: The study was developed with a Toshiba Asteion single-slice CT scanner and a GE BrightSpeed multi-slice CT unit in two hospitals. Measurements were performed with a pencil-type ionization chamber and a 16 cm-diameter polymethylmethacrylate trunk phantom. RESULTS: No significant difference was observed in the values for weighted air kerma index (CW), but the differences were relevant in values for volumetric air kerma index (CVOL), air kerma-length product (PKL,CT) and effective dose. CONCLUSION: Only the CW values were lower than the DRL, suggesting that dose optimization might not be necessary. However, PKL,CT and effective dose values stressed that there still is room for reducing pediatric radiation doses. The present study emphasizes the importance of determining all dosimetric quantities associated with CT scans.
OBJECTIVE: Aiming at contributing to the knowledge on doses in computed tomography (CT), this study has the objective of determining dosimetric quantities associated with pediatric abdominal CT scans, comparing the data with diagnostic reference levels (DRL). MATERIALS AND METHODS: The study was developed with a Toshiba Asteion single-slice CT scanner and a GE BrightSpeed multi-slice CT unit in two hospitals. Measurements were performed with a pencil-type ionization chamber and a 16 cm-diameter polymethylmethacrylate trunk phantom. RESULTS: No significant difference was observed in the values for weighted air kerma index (CW), but the differences were relevant in values for volumetric air kerma index (CVOL), air kerma-length product (PKL,CT) and effective dose. CONCLUSION: Only the CW values were lower than the DRL, suggesting that dose optimization might not be necessary. However, PKL,CT and effective dose values stressed that there still is room for reducing pediatric radiation doses. The present study emphasizes the importance of determining all dosimetric quantities associated with CT scans.
Entities:
Keywords:
Air kermalength product; Dosimetric quantities; Pediatric computed tomography scans; Volumetric air kerma index; Weighted dose index
Developments in computed tomography (CT) as a medical imaging method, and its constant
technological advances over the years have expanded its application; currently, CT
became one of the most relevant radiological techniques easily accessible to the
greatest part of the population(.
The pediatric patient group has increased in this population over the years; for
example, in 1980, 3 million CT scans were performed in the United States of America
(USA) and in 1996 it reached 62 million with 4 million in children(. It is estimated that 10% of all CT
scans performed in the world involve pediatric patients(.The consequence of the dissemination of this technique in the medical practice is an
increase in the collective dose due to patient exposures, since the doses are much
higher than those from the exposures related to any other conventional radiology
technique(. According to many
epidemiological studies, the significant increase in the number of CT scans associated
with the increase in dose should presumably lead to a higher probability of development
of harmful effects, particularly in children(.The current scenario raises the necessity of a radiological protection policy aiming at
the knowledge and control of radiation doses involved in pediatric CT procedures. The
first international discussion focused on the radiological protection of patients
occurred in 2001 during the Conference of Malaga, approaching radiotherapy and
radiodiagnosis procedures as well as those related to nuclear medicine(.The most frequent method used to estimate doses in CT consists in measurements with
either a ionization chamber positioned free-in-air or inserted into a head or neck
phantom; the computed tomography dose index (CDTI) is determined and adopted as
reference(. The International
Atomic Energy Agency (IAEA) suggests the use of the term "air kerma índex"
(Ca) replacing CTDI, but the two dosimetric quantities are obtained by a
same procedure and have a same numeric value(.The purpose of knowing the dosimetric quantity values is to allow the comparison between
such values and the diagnostic reference levels (DRL). DRLs are utilized as reference
tools for quality control of the technique, but should not be used as exact values
adopted with the purpose of dose restriction. The DRL objective is to avoid radiation
dose to the patient that does not contribute to the clinical purpose of a medical
imaging procedure, indicating the necessity of an optimization process(.The concern with radiation levels in pediatric CT scans has stimulated actions aimed at
radiological protection of children, among them the Image Gently campaign in the
USA(. The California state
has sanctioned a regulation establishing the inclusion of dosimetric quantity values
involved in tomographic procedures in the patients' medical records(. In Brazil, no similar action has been
adopted by the authorities. Furthermore, studies approaching dosimetric quantity values
in tomographic procedures practically do not exist(.The present study is aimed at quantifying dosimetric quantity values specific for CT,
focusing on pediatric patients undergoing abdominal CT scans at two institutions in the
city of Belo Horizonte, MG, Brazil. The objective is to study the application of
dosimetric quantity values in the process of dose optimization, besides evaluating the
conformity of such values with the diagnostic reference levels.
MATERIALS AND METHODS
The IAEA definitions( were adopted
for the dosimetric quantities expressed in terms of kerma (kinetic energy released per
unit mass). The weighted air kerma index (CW) (equation 1) has the objective of measuring the air kerma index
within the phantoms; the volumetric air kerma index (CVOL) (equation 2) provides the estimate of the
dose in a single section; the air kerma-length product (PKL,CT) (equation 3) provides the air kerma in the
whole irradiated area during the acquisition of the tomographic image. The authors have
also adopted the concept of effective dose estimated from PKL,CT as a
function of a conversion factor k (equation 4; Table 1),
depending only on the irradiated body region(.
Table 1
Conversion factor k values for calculation of effective
dose(.
Body region
k (mSv mGy-1 cm-1)
0 year-old
1-year-old
5-year-old
10-year-old
Adult
Head and neck
0.013
0.0085
0.0057
0.0042
0.0031
Head
0.011
0.0067
0.0040
0.0032
0.0021
Neck
0.017
0.012
0.011
0.0079
0.0059
Chest
0.039
0.026
0.018
0.013
0.014
Abdomen
0.049
0.030
0.020
0.015
0.015
Trunk
0.044
0.028
0.019
0.014
0.015
Conversion factor k values for calculation of effective
dose(.Air kerma rates were obtained within a cylindrical polymethyl metacrylate (PMMA) phantom
with density 1.19 ± 0.01 g.cm-3, 16 cm in diameter and 15 cm in length, which
is a pediatric trunk phantom indicated for patients in the age range from 1 to 15
years). The phantom was positioned and carefully aligned with the laser beam of the CT
apparatus within the gantry. With the ionization chamber inserted between the peripheral
and central holes of the phantom, three measurements were performed in terms of
PKL, and the readings were duly corrected by the calibration factor
NPKL,Q = 9.97 × 103 Gy.cm.unit-1(15), by the energy
dependence factor (kQ = 1) and by the pressure and temperature factor
(kTP = 0.9). The measurements of the dosimetric quantity were carried out
in two devices of two hospitals where abdominal CT scans were performed in pediatric
patients, as follows: a Toshiba Asteion single-slice and a GE BrightSpeed multi-slice
machines. A 10X4-CT pencil-type Radcal® ionization chamber (100 mm in length
and radius of approximately 3 mm) coupled with a 9060 model Radcal®
electrometer was utilized to determine the air kerma rates in the center
(CPMMA,100,C) and extremity holes of the phantom (CPMMA,100,p)
in order to calculate the CW (equation 1).
RESULTS
With the single-slice CT equipment, the CW value was 18.73 ± 0.26 mGy, which, for the
pitch of 1.2 adopted by the hospital, corresponds to the CVOL value of 14.61 mGy. The
scan length for a pediatric abdominal CT is 22 cm over the patient's body, resulting in
PKL,CT of 343.51 mGy.cm; this corresponds to an effective dose of 6.87
mSv, considering a conversion factor k = 0.020 (Table 1) for the abdominal region of a patient at the age of five
years(.With the multi-slice CT equipment, for a pitch of 0.9 and the same 22 cm scan length,
the CW was 18.81 ± 0.22 mGy; the CVOL, 20.07 mGy; the
PKL,CT, 441.64 mGy.cm; and the effective dose 8.83 mSv.
DISCUSSION
A comparison of the results obtained in the two CT devices showed that the difference
between the CW values was not significant, but for CVOL,
PKL,CT and effective dose, the increases of, respectively, 37%, 29% and
29% observed with the multi-slice equipment may be considered significant. Such a fact
is explained by the pitch of 0.9 adopted by the hospital, which causes tomographic
sections overlapping and, consequently, greater patient exposure to radiation.DRLs for pediatric abdominal CT scans in five-year-old patients are recommended as 25
mGy for CW, and 360 mGy.cm for PKL,CT
(. Both hospitals indicated values
lower than the DRLs for such dosimetric quantities, except for the value of
PKL,CT in the hospital with the multi-slice equipment, that remained 23%
higher than the DRL, as shown on Figure 1.
Considering that there is no tolerance range for DRL values, the result suggests that
studies should be done in the hospital with the multi-slice equipment, in order to
verify if an appropriate level of imaging quality could be achieved with lower levels of
radiation doses.
Figure 1
Comparison between values for CW and PKL,CT and DRL in
pediatric abdominal CT scans.
Comparison between values for CW and PKL,CT and DRL in
pediatric abdominal CT scans.The increase of 2 mSv in the effective dose during the scan with the multi-slice
equipment, as compared with the scan performed with the single-slice equipment, does not
indicate the necessity of optimization, even considering that the first equipment offers
better technological resources to provide patients with lower radiation doses without
loss in the imaging quality. Until the present moment, none DRL was established in terms
of effective dose for pediatric scans. The values obtained with surveys of the effective
doses for abdominal scans in adult patients in some countries range from 5.3 to 13.2
mSv, with a median of 9.5 mSv(. Such
data suggest that the effective dose of 8.83 mSv observed in the multi-slice equipment
is not appropriate for pediatric scans; the results suggest that there is room for
improving the optimization in the hospital.
CONCLUSIONS
If only the dosimetric quantity CW is considered as a comparative tool, both
hospitals would not need to implement an optimization process. However, as the
PKL,CT values in the hospital with the multi-slice equipment are analyzed,
one may conclude that an optimization of the technical procedures should be considered
in order to reduce the radiation doses in pediatric abdominal CT scans.The effective dose value might be adopted as a criterion to be taken into account and
for decision-making about the implementation of an optimization process; it should be
considered that a tomographic image is to be obtained without loss in the diagnostic
quality, but the radiation dose delivered to the patient should be as low as
possible.CT scan dosimetry has at least four dosimetric quantities with specific objectives. The
medical physicist responsible for determining their values should be aware of the
differences among them and able to select the one that best fits his/her purpose. The
consideration about the necessity of optimization based on a single dosimetric quantity
may be limited.
Authors: Mark S Pearce; Jane A Salotti; Mark P Little; Kieran McHugh; Choonsik Lee; Kwang Pyo Kim; Nicola L Howe; Cecile M Ronckers; Preetha Rajaraman; Alan W Sir Craft; Louise Parker; Amy Berrington de González Journal: Lancet Date: 2012-06-07 Impact factor: 79.321