Luciano Marcondes Machado Nardozza1, Ana Paula Passos2, Edward Araujo Júnior1, Rafael Frederico Bruns3, Antonio Fernandes Moron4.
Abstract
OBJECTIVE: To establish reference intervals for the fetal cisterna magna volume by means of two-dimensional (2D) method using the multiplanar mode of three-dimensional ultrasonography.Entities:
Keywords: Cisterna magna; Fetus; Reference intervals; Three-dimensional ultrasonography; Two-dimensional ultrasonography
Year: 2014 PMID: 25741085 PMCID: PMC4337116 DOI: 10.1590/0100-3984.2013.1852
Source DB: PubMed Journal: Radiol Bras ISSN: 0100-3984
Figure 1Calculation of the fetal cisterna magna volume by the 2D method, utilizing the multiplanar mode of three-dimensional ultrasonography. Width (D1 = 2.69 cm) and length (D2 = 0.95 cm) will be marked in the axial plane; and height (D3 = 2.17 cm), in the sagittal plane. Automatically, the apparatus calculates the volume by means of the product of the three distances, multiplied by the constant 0.52 constant
Figure 2Scatter plot of correlation between volume of the fetal cisterna magna and gestational age, by means of the 2D method
Percentiles 5th, 50th and 95th of the fetal cisterna magna (cm3) by means of the 2D method, obtained in the multiplanar mode of 3DUS
| Gestational age (weeks) |
| 5th | 50th | 95th |
|---|---|---|---|---|
| 17 | 10 | 0.31 | 0.44 | 0.67 |
| 18 | 18 | 0.40 | 0.70 | 1.03 |
| 19 | 10 | 0.50 | 0.97 | 1.47 |
| 20 | 18 | 0.62 | 1.25 | 1.91 |
| 21 | 21 | 0.76 | 1.54 | 2.37 |
| 22 | 24 | 0.91 | 1.85 | 2.84 |
| 23 | 21 | 1.08 | 2.17 | 3.32 |
| 24 | 9 | 1.26 | 2.51 | 3.81 |
| 25 | 13 | 1.46 | 2.85 | 4.31 |
| 26 | 19 | 1.67 | 3.21 | 4.82 |
| 27 | 25 | 1.90 | 3.59 | 5.35 |
| 28 | 21 | 2.14 | 3.97 | 5.88 |
| 29 | 15 | 2.40 | 4.37 | 6.43 |
N, number of patients at each gestational age.
Figure 3Bland-Altman graph showing the average of the relative difference between two measurements performed by a single investigator (intraobserver) plotted against the difference of his mean values for the 2D volume of the fetal cisterna magna
Figure 4Bland-Altman graph showing the average of the relative difference of two measurements performed by two investigators (interobserver) plotted against the difference of their mean values for the 2D volume of the fetal cisterna magna