Aruna Arujuna1, Rashed Karim2, Niloufar Zarinabad2, Jaspal Gill2, Kawal Rhode2, Tobias Schaeffter2, Matthew Wright3, C Aldo Rinaldi3, Michael Cooklin4, Reza Razavi3, Mark D O'Neill3, Jaswinder S Gill3.
Abstract
AIMS: To prospectively compare cardiac magnetic resonance late gadolinium enhancement (LGE) findings created by standard vs. robotically assisted catheter ablation lesions and correlate these with clinical outcomes. METHODS ANDEntities:
Keywords: Atrial fibrillation; Cardiac magnetic resonance imaging; Catheter ablation; Long-term follow-up; Pulmonary vein isolation; Robotic-assisted
Mesh:
Substances:
Year: 2015 PMID: 25687748 PMCID: PMC4535555 DOI: 10.1093/europace/euu364
Source DB: PubMed Journal: Europace ISSN: 1099-5129 Impact factor: 5.214
Baseline characteristics
| Total population ( | Robotic navigation ( | Manual catheter ablation ( | ||
|---|---|---|---|---|
| Male, | 28 (70) | 15 (75) | 13 (65) | 0.53 |
| Age (years) | 54 ± 13.8 | 54 ± 14.3 | 55 ± 11.2 | 0.86 |
| AF duration (months) | 30 ± 11.2 | 32 ± 9.1 | 30 ± 12.6 | 0.78 |
| LA size (cm) | 3.4 ± 3.5 | 3.4 ± 0.3 | 3.5 ± 0.2 | 0.92 |
| LVEF, % | 56 ± 8 | 58 ± 6 | 56 ± 8 | 0.86 |
| Hypertension | 11 | 7 | 4 | >0.10 |
| Diabetes | 1 | 0 | 1 | >0.10 |
| Coronary artery disease | 1 | 1 | 0 | >0.10 |
| Thyroid | 2 | 2 | 0 | >0.10 |
| Smoking | 6 | 4 | 2 | >0.10 |
| Fluoroscopy dose (Gycm2) | 1227 ± 667 | 1291 ± 607 | 1141 ± 755 | 0.53 |
| Procedure time (min) | 210 ± 26.1 | 220 ± 21.4 | 199 ± 27.7 | 0.06 |
| Number of applications | 134 ± 42 | 141 ± 33 | 126 ± 58 | 0.63 |
| Energy delivered (Joules) | 84201 ± 28855 | 88391 ± 28519 | 80011 ± 28338 | 0.46 |
Coefficient of variation between robotically assisted and standard ablation in acute and follow-up scans for no recurrences vs. recurrences
| Acute | Follow up | |||
|---|---|---|---|---|
| Robotically assisted | Standard | Robotically assisted | Standard | |
| No recurrences | 0.10 | 0.12 | 0.12 | 0.13 |
| Recurrences | 0.10 | 0.28 | 0.11 | 0.46 |