Tricuspid valve regurgitation following temporary or permanent endocardial lead insertion, and the impact of cardiac resynchronization therapy.

Masoud Sadreddini1, Michelle J Haroun1, Lisanne Buikema2, Carlos Morillo1, Sebastian Ribas1, Syamkumar Divakaramenon1, Stuart J Connolly1, Robby Nieuwlaat3, Eva M Lonn1, Jeff S Healey1, Hisham Dokainish1.   

Abstract

BACKGROUND: While some studies indicate that permanent pacemaker implantation is associated with development of tricuspid regurgitation (TR), other studies indicate no association.Little is known about the impact of temporary lead insertion during ablation procedures, or whether therapy (CRT) prevents TR post-device implantation. HYPOTHESIS: We hypothesized that permanent, but nottemporary endocardial leads, are associated with development of TR, and that CRT would prevent (physiologic) TR.
METHODS: We performed a retrospective study of consecutive patients who underwent first device or radiofrequency catheter ablation over a 12-month period at a single, tertiary academic center who underwent pre- and post-procedure echocardiography.
RESULTS: In the 89 patients in the device group, the degree of TR significantly increased ≥ 1 grade post-permanent lead implantation: 9 had less TR, 46 were unchanged, and 34 had more TR(p=0.005). TR increased in the 62 patients who underwent device implantation without CRT (p=0.005), but did not increase in the 27 patients with CRT (p=0.47). In the 66 patients in the ablation group, there was no significant change in TR post-ablation: 8 had less TR, 48 were unchanged, and 10 had more TR (p=0.31).
CONCLUSION: Permanent endocardial lead implantation was associated with an increase in TR; however, patients who underwent device implantation with CRT did not have an increase in TR.Temporary lead insertion during ablation was not associated with changes in the degree of TR. A large, prospective study is needed to accurately define the incidence and exact mechanisms of permanent endocardial lead-related TR.

Entities:  

Keywords:  Cardiac resynchronization; echocardiography; implantable cardiac defibrillators; pacemakers; tricuspid regurgitation.

Year:  2014        PMID: 25674163      PMCID: PMC4321203          DOI: 10.2174/1874192401408010113

Source DB:  PubMed          Journal:  Open Cardiovasc Med J        ISSN: 1874-1924


INTRODUCTION

Over the last decade, there has been a significant increase in the number of cardiac device implantations –permanent pacemakers (PPM), implantable cardiac defibrillators (ICD) and biventricular pacemakers (BiV)− worldwide [1-3]. While some small, mainly retrospective studies, case reports and reviews have suggested that permanent endocardial lead implantation can lead to tricuspid regurgitation (TR) [4-13], other studies have not demonstrated any association between device implantation and the development of (TR) [14, 15]. In addition, the impact of cardiac resynchronization therapy (CRT) on the development of TR post-device implantation has not been defined. Furthermore, patients undergoing radiofrequency catheter ablation for atrial and ventricular arrhythmias routinely undergo temporary pacemaker lead insertion, often using multiple leads that cross the tricuspid valve [16,17]. Yet, very little is known about the effects of temporary lead insertion on TR severityin these patients [18], and no studies have assessed the effects of both temporary and permanent lead insertion on tricuspid valve function. In our study, we hypothesized that patients undergoing temporary lead insertion would not have a significant increase in the degree of TR post-procedure, whereas patients undergoing permanent lead insertion via device implantation would have a significant increase in the severity of TR. In addition, we hypothesized that CRT would prevent the development of (physiologic) TR in patients undergoing permanent device implantation.

METHODS

Patient Population

We performed a retrospective study of consecutive patients who underwent first device implantation (PPM, ICD, BiV), or catheter ablation, between August 2011 and August 2012 at a single, tertiary,academic medical center. Patients were included if a two-dimensional (2D) trans-thoracic echocardiogram was performed pre-procedure and within 12 months following device implantation or catheter ablation. We excluded patients under the age of 18, patients who had prior ablation or device implantation, and patients with congenital heart disease, as many of the latter patients have other structural cardiac reasons for TR. The study was approved by our institutional Research Ethics Board.

Data Collection

The medical records of all patients undergoing catheter ablation or device implantation during the study period were reviewed for the study inclusion and exclusion criteria. Clinical variables analyzed included age, sex, hypertension, diabetes, smoking status, dyslipidemia, significant coronary artery disease (>/=50% stenosis of any major epicardial coronary at angiography or a positive functional study demonstrating ischemia), clinical heart failure, chronic kidney disease (estimated GFR <60 ml/min), and baseline medication use. Echocardiographic variables included the degree of TR, RV size and function, right atrial size, pulmonary artery systolic pressure (PASP), left ventricular ejection fraction, left atrial size, aortic valve stenosis or regurgitation and mitral valve stenosis or regurgitation [19, 20]. The degree of TR was based on the following ordinal scale: none/trivial, mild, moderate and severe [20]. A change in TR was defined as a change of at least one grade (for example, from moderate to severe or moderate to mild). RV size and function, LA size and RA size were also assessed using an ordinal scale; normal, mild, moderate and severe.

Statistical Analysis

Baseline patient characteristics were compared between the device and ablation group using the Student’s t test for continuous variables, and the Fisher’s exact test for categorical variables. The difference between ordinal echocardiographic variables (ie. pre- and post procedure TR) was tested using the McNemar-Bowker test.

RESULTS

A total of 1041 patients underwent first device implantation or catheter ablation between August 2011 and August 2012. Of these, 527 patients underwent device implantation, and 524 patients underwent radiofrequency catheter ablation. In total, 155 patients met study inclusion criteria, and were therefore analyzed. Of the 89/155 (57%) patients who underwent permanent device implantation, 37 patients received an ICD, 25 patients received a BiV-ICD, 26 patients received a PPM and one patient received a BiValone. The remaining 66/155 (43%) patients underwent radiofrequency catheter ablation for arrhythmia; 28 patients underwent atrial fibrillation ablation, 21 patients underwent atrioventricular re-entry ablation, 15 patients underwent atrial flutter ablation, and 2 patients underwent ventricular tachycardia ablation. The patient characteristics of the device and ablation groups are shown in (Table ). As expected, patients undergoing device implantation had more cardiovascular risk factors, coronary artery disease, and congestive heart failure, compared to the ablation group (Table ).The mean time to follow up echocardiographic study was 5.6 months (SD ± 3.9) in the device group and 4.7 (SD ± 3.6) months in the ablation group (p=0.39 for comparison). In patients undergoing permanent device implantation, the degree of TR increased in 34 patients, decreased in 9 patients, and was unchanged in 46 patients (p=0.005 for comparison) (Table , Fig. ).While the 42 patients who underwent device implantation without CRT had a significant increase in TR (p=0.005), the 27 patients who underwent device implantation with CRT had no significant increase in TR (p=0.47) (Table ). In the device group, there was no significant difference in right ventricular size or systolic function (p=0.68 and p=0.42 respectively), left ventricular ejection fraction (37.9%±3.7vs. 37.5%±4.5, p=0.82) or left sided valve disease observed (Table ). In patients who underwent radiofrequency catheter ablation alone, there was no significant change in the degree of TR: TR increased in 10 patients, decreased in 8 patients, and was unchanged in 48 patients (p=0.31 for comparison). (Table , Fig. ) Of note, following catheter ablation, more than 90% of patients had either trivial or mild tricuspid regurgitation. There was no significant change in RV size or RV function following catheter ablation (p=0.38 and p=0.63 respectively), nor was there a statistically significant change in left ventricular ejection fraction (59.8± 4.5 vs. 58.9± 5.3,p= 0.79) observed.

DISCUSSION

This study is the first to analyze the impact of both temporary and permanent lead insertion on TR degree in patients undergoing permanent device implantation and ablation, respectively. The findings show that patients who underwent temporary lead insertion during ablation procedures did not have a significant increase in the degree of TR, whereas patients who underwent permanent lead insertion via device implantation demonstrated a significant increase in TR. Importantly, and a novel finding, patients who underwent permanent device implantation with CRT had no significant change in TR, in contrast to patients who had a permanent device without CRT, who did have a significant increase in TR. Previous studies—primarily case reports, small retrospective studies and reviews [6-13] have described an association between permanent endocardial lead implantation and post-procedure TR. Paniagua, et al. [4] reported a higher prevalence of moderate and severe TR in patients with 374 PPM leads compared to age and sex matched controls (25% vs. 12%, odds ratio 4.75,P<0.001); however, the study did not compare the degree of TR pre- and post-procedure. Klutstein et al. [6] retrospectively analyzed 545 patients before and after PPM placement and showed an increase in TR severity by two or more grades in 18% of patients. Kim et al. [7] retrospectively studied 248 patients undergoing device implantation with pre- and post-implantation echocardiograms, findings that TR increased by at least one grade in 24.2% patients (p=0.048). A recent retrospective study in 791 patients undergoing tricuspid valve surgery over 21 years at a single center showedthat a transvenous pacemaker implanted preoperatively or ≤30 days of surgery was an independent predictor of moderate to severe TR [9]. Another study of 125 patients reported an increase in significant TR from 8.7% to 31.6% (P<0.001) following device implantation [10]. On the other hand, two small studies demonstrated no significant effect of device implantation on TR severity [14, 15]. Kucukarslan et al. found that, in 61 patients referred for PPM or ICD, there was no change in TR severity at 6 months follow-up [14]. In a small prospective study of 35 patients undergoing echocardio-graphy pre- and post-PPM or ICD, there also was no change seen in the degree of TR post-device implantation. [15]. The findings of our study indicate that permanent endocardial leads are associated with an increase in TR. While there were no differences in RV size and function in patients with permanent devices at follow-up echocardiography, and the study was not designed to adjudicate clinical endpoints such as development of heart failure, the results remain important, as it is known that chronic moderate or severe TR can lead to RV failure and possibly death, although this may take many years to develop [25, 26]. Significantly, patients undergoing device implantation demonstrated that cardiac resynchronization therapy (CRT) was not associated with an increase in post procedure TR. This finding an been seen as analogous to findings that CRT may reduce mitral valve regurgitation as left ventricular resynchronization decreases ventricular remodeling and subsequent atrioventricular valve functional regurgitation,although this has only been demonstrated previously in patients with mitral regurgitation [22, 23]. While mechanical mechanisms of TR post-permanent device implantation such as leaflet perforation, leaflet impingement or mechanical interference of tricuspid leaflet coaptation are important and have been previously described and verified in 41 patients undergoing TV surgery for PPM-associated TR [5], physiologic mechanisms of TR, by dyssynchronous ventricular activation and resultant induction of leaflet malcoaptation that may be reversible with de-activation of pacing and restoration of ventricular synchrony, are also important and have been described [24]. It is possible that the CRT patients in our study had less right ventricular dyssynchrony and less resultant functional (physiologic) TR and therefore did not have an increase in TR post-BIV device, as opposed to the significant increase seen in patients in our study who underwent PPM or ICD implantation without CRT. However, these putative mechanisms of TR, and the effects of CRT, would need detailed evaluation in a prospective, adequately powered study. Only one retrospective study published in 1996 with echocardiography prior to, and after radiofrequency catheter ablation identified no significant changes in echocardiographic findings post-ablation, mentioning that valvular regurgitation did not increase. [18] The first, however, to specifically assess TR prior to and post-radiofrequency ablation and post-permanent lead insertion performed during the same period, at the same centre and by the same operators, finding that there was no significant change in TR after ablation. The differences in TR following catheter ablation and permanent lead implantation are not likely due to chance alone. Many of the proposed mechanisms of device related TR such as chronic lead impingement/restriction, scarring of leaflets or RV pacing would be important in patients with permanent, but not temporary, endocardial leads [4-16,18]. Indeed, one retrospective study suggested that in the 48 patients who received ≥3 permanent endocardial leads, there was more TR seen than in the 48 patients who only underwent implantation of 2 permanent leads (DDD pacing), suggesting that interference of TV function by (multiple) permanent leads crossing the TV may be an important mechanism of TR. [27] However, the limited power of studies in this field, and the lack of clear delineation of the true incidence and exact mechanisms of post-device TR point to the clear need for adequately powered, prospective studies of patients undergoing device implantation, especially given the global increase in PPM, ICD and BIV devices [1-3]. If future prospective studies do indeed show that permanent device implantation is a cause of clinically important TR, the current development of leadless pacemakers would be arguably even more important.
Table 1.

Baseline characteristics of patients undergoing first device implantation or catheter ablation.

Patient CharacteristicsDevice group  N=89 (%)Ablation group  N=66 (%)P-value
Mean Age, years (SD)69.6 (13.1)58.1 (16.5)<0.001
Female58 (65.2)35 (53.0)0.273
Diabetes Mellitus17 (19.1)12 (18.2)0.888
Hypertension64 (71.6)24 (36.4)<0.001
Any Smoking History45 (50.1)15 (22.7)<0.001
Dyslipidemia62 (69.7)23 (34.3)<0.001
Coronary Artery Disease54 (60.7)8 (12.1)<0.001
Clinical Heart Failure42 (47.2)1(1.5)<0.001
*Chronic Kidney Disease 22 (23.6)3 (4.5)0.001
Medications    
Aspirin 41 (46.1)20 (30.3)0.047
Oral Anticoagulant 38 (42.7)36 (54.5)0.144
Beta Blockers 77 (86.5)33 (50)<0.001
Calcium Channel Blockers 10 (11.2)28 (42.4)<0.001
Digoxin 12 (13.5)3 (4.5)0.063
Class III Anti-arrhythmic 12 (13.5)16 (24.2)0.085
ACE-inhibitors or Angiotensin Receptor Blockers 67 (75.3)22 (33.3)<0.001
Statins 59 (66.3)21 (31.8)<0.001
Aldosterone Antagonists 35 (39.3)7 (10.8)<0.001
Diuretic48 (53.9) 5 (7.7)<0.001

Chronic Kidney Disease defined as glomerular filtration rate < 60 mls/min.

Table 2.

Tricuspid regurgitation (tr) degree before and after device implantation and ablation.

 Device (n=89)Ablation (n=66)
TR DegreeTR Before TR Afterp-valueTR Before TR Afterp-value
None/trivial48 (54%)33 (37%)0.00550 (76%)48 (73%)0.30
Mild26 (29%)31 (35%)13 (19%)11 (17%)
Moderate13 (15%)20 (22%)1 (2%)6 (9%)
Severe2 (2%)5 (6%)2 (3%)1 (2%)
Table 3.

TR degree before and after device implantation with or without cardiac resynchronization therapy (crt).

 CRT(n=27)No CRT(n=62)
TR DegreeTR BeforeTR Afterp-valueTR BeforeTR Afterp-value
None/Trivial11 (41%)11 (41%)   0.4737 (60%)22 (35%)   0.005
Mild10 (37%)9 (33%)17 (27%)22 (35%)
Moderate6 (22%)6 (22%)7 (11%)14 (23%)
Severe0 (0%)1 (4%)2 (3%)4 (6%)
Table 4.

Echocardiographic measurements at baseline and follow up.

  VariableDevice GroupAblation Group
Pre (%)Post (%)P valuePre (%)Post (%)P value
Right Ventricular Size
Normal79.871.90.6892.4970.38
Mild11.212.4 6.13 
Moderate7.910.1 1.50 
Severe02.2 00 
Right Ventricular Function
Normal66.562.90.4295.5970.63
Mild 14.612.4 4.51.5 
Moderate 12.412.4 01.5 
Severe3.47.9 00 
Right Atrial Size
Normal43.833.70.1562.165.20.15
Mild 14.614.6 18.29.1 
Moderate 15.713.5 6.19.1 
Severe5.614.6 6.14.5 
Pulmonary Hypertension
None49.448.30.3251.556.10.71
Mild 16.918 7.64.5 
Moderate 14.618 01.5 
Severe2.214.6 00 
Left Atrial Size
None20.213.50.2045.543.90.80
Mild 2724.7 19.719.7 
Moderate 24.729.2 19.712.1 
Severe22.529.2 12.119.7 
Mitral Regurgitation
None33.731.50.8174.269.70.34
Mild 34.834.8 18.219.7 
Moderate 24.724.7 6.19.1 
Severe4.57.9 1.51.5 
Aortic Stenosis
None9189.91.009798.51.00
Mild 7.97.9 1.50 
Moderate 01.1 1.51.5 
Severe00 00 
Aortic Regurgitation
None85.485.41.0092.493.90.50
Mild 12.412.4 7.63 
Moderate 1.11.1 03 
Severe00 00 
  26 in total

1.  Transvenous pacemaker leads do not worsen tricuspid regurgitation: a prospective echocardiographic study.

Authors:  D W Leibowitz; S Rosenheck; A Pollak; M Geist; D Gilon
Journal:  Cardiology       Date:  2000       Impact factor: 1.869

2.  Recommendations for chamber quantification: a report from the American Society of Echocardiography's Guidelines and Standards Committee and the Chamber Quantification Writing Group, developed in conjunction with the European Association of Echocardiography, a branch of the European Society of Cardiology.

Authors:  Roberto M Lang; Michelle Bierig; Richard B Devereux; Frank A Flachskampf; Elyse Foster; Patricia A Pellikka; Michael H Picard; Mary J Roman; James Seward; Jack S Shanewise; Scott D Solomon; Kirk T Spencer; Martin St John Sutton; William J Stewart
Journal:  J Am Soc Echocardiogr       Date:  2005-12       Impact factor: 5.251

3.  Effects of cardiac resynchronization therapy on the mechanisms underlying functional mitral regurgitation in congestive heart failure.

Authors:  Maria Cristina Porciani; Roberto Macioce; Gabriele Demarchi; Marco Chiostri; Nicola Musilli; Francesco Cappelli; Alessio Lilli; Giuseppe Ricciardi; Luigi Padeletti
Journal:  Eur J Echocardiogr       Date:  2006-01

4.  Increased prevalence of significant tricuspid regurgitation in patients with transvenous pacemakers leads.

Authors:  D Paniagua; H R Aldrich; E H Lieberman; G A Lamas; A S Agatston
Journal:  Am J Cardiol       Date:  1998-11-01       Impact factor: 2.778

5.  The importance of tricuspid regurgitation and right ventricular overload in ICD/CRT recipients: beside the left, beyond the left.

Authors:  Serafino Fazio; Guido Carlomagno
Journal:  Pacing Clin Electrophysiol       Date:  2011-08-07       Impact factor: 1.976

6.  Tricuspid insufficiency does not increase early after permanent implantation of pacemaker leads.

Authors:  Nezihi Kucukarslan; Ata Kirilmaz; Eralp Ulusoy; Mehmet Yokusoglu; Nikola Gramatnikovski; Ertugrul Ozal; Harun Tatar
Journal:  J Card Surg       Date:  2006 Jul-Aug       Impact factor: 1.620

Review 7.  Tricuspid regurgitation in patients with pacemakers and implantable cardiac defibrillators: a comprehensive review.

Authors:  Rasha Al-Bawardy; Amar Krishnaswamy; Mandeep Bhargava; Justin Dunn; Oussama Wazni; E Murat Tuzcu; William Stewart; Samir R Kapadia
Journal:  Clin Cardiol       Date:  2013-03-25       Impact factor: 2.882

Review 8.  Meta-analysis of ablation of atrial flutter and supraventricular tachycardia.

Authors:  Peter Spector; Matthew R Reynolds; Hugh Calkins; Manu Sondhi; Yingxin Xu; Amber Martin; Catherine J Williams; Isabella Sledge
Journal:  Am J Cardiol       Date:  2009-09-01       Impact factor: 2.778

9.  Impact of tricuspid regurgitation on long-term survival.

Authors:  Jayant Nath; Elyse Foster; Paul A Heidenreich
Journal:  J Am Coll Cardiol       Date:  2004-02-04       Impact factor: 24.094

10.  The effect of transvenous pacemaker and implantable cardioverter defibrillator lead placement on tricuspid valve function: an observational study.

Authors:  Juyong B Kim; Daniel M Spevack; Paul A Tunick; John R Bullinga; Itzhak Kronzon; Larry A Chinitz; Harmony R Reynolds
Journal:  J Am Soc Echocardiogr       Date:  2007-07-02       Impact factor: 5.251

View more
  4 in total

1.  Multisite cardiac resynchronization therapy for traditional and non-traditional indications.

Authors:  Avishag Laish-Farkash; Sharon Bruoha; Vladimir Khalameizer; Chaim Yosefy; Yoav Michowitz; Mahmoud Suleiman; Amos Katz
Journal:  J Interv Card Electrophysiol       Date:  2018-02-13       Impact factor: 1.900

2.  Comparison of Tricuspid Regurgitation Severity Between Cardiac Resynchronization Therapy Versus Right Ventricular Pacing in Patients With Chronic Obstructive Pulmonary Disease.

Authors:  James Livesay; Benjamin Fogelson; Hassan Tahir; Raj Baljepally
Journal:  Cardiol Res       Date:  2022-06-16

3.  Prospective study of tricuspid valve regurgitation associated with permanent leads in patients undergoing cardiac rhythm device implantation: Background, rationale, and design.

Authors:  Hisham Dokainish; Esam Elbarasi; Simona Masiero; Caroline Van de Heyning; Michela Brambatti; Sami Ghazal; Said Al-Maashani; Alessandro Capucci; Lisanne Buikema; Darryl Leong; Bharati Shivalkar; Johan Saenen; Hielko Miljoen; Carlos Morillo; Syam Divarakarmenon; Guy Amit; Sebastian Ribas; Aaron Brautigam; Erika Baiocco; Alessandro Maolo; Andrea Romandini; Simone Maffei; Stuart Connolly; Jeff Healey
Journal:  Glob Cardiol Sci Pract       Date:  2015-10-17

4.  Lead-Specific Features Predisposing to the Development of Tricuspid Regurgitation After Endocardial Lead Implantation.

Authors:  Kalilur Anvardeen; Rajeev Rao; Samir Hazra; Karen Hay; Hongyan Dai; Nik Stoyanov; David Birnie; Girish Dwivedi; Kwan Leung Chan
Journal:  CJC Open       Date:  2019-10-31
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.