Literature DB >> 10793428

Long-term outcome of patients with multiple (> or = 3) noninfected transvenous leads: a clinical and echocardiographic study.

C C de Cock1, M Vinkers, L C Van Campe, P M Verhorst, C A Visser.   

Abstract

To prospectively assess the incidence and clinical significance of thromboembolic complications in patients with multiple (> or = 3) noninfected transvenous leads; 48 consecutive patients were evaluated. Half of the patients had two ventricular leads and one atrial lead, 15 patients had two atrial leads and one ventricular lead, while 9 patients had two ventricular and two atrial leads. No additional care was provided except for aspirin (80 mg bid) and annually performed echo-Doppler studies. Clinical follow-up included signs and symptoms of subclavian and/or axillary vein thrombosis, the presence of right congestive heart failure, the number of hospital admissions, and death. Echo-Doppler studies assessed the presence of an enlarged right atrium or ventricle, right atrial or ventricular spontaneous contrast, and the presence of tricuspid regurgitation. During a total follow-up of 7.4 +/- 2.2 years there were no differences in the incidence of clinical variables as compared to age-matched controls with DDD pacemakers. The most common complication was transient venous thrombosis (mostly presenting as venous prominence 1-2 weeks after implantation), which was seen in 17% of the study group versus 15% in controls (NS). Cumulative mortality was not different in both groups (13% in the study group vs 15% in controls). No differences were present with respect to hospital admissions (1.1 +/- 0.27/year in the study group vs 1.2 +/- 0.30/year in the controls). In patients with multiple ventricular leads, tricuspid regurgitation on echo-Doppler studies was more frequent (24%) as compared to controls (4%); however, clinical signs of right heart failure were equally distributed. Thus, patients with multiple (> or = 3) noninfected leads have no clinical adverse outcome during long-term follow-up.

Entities:  

Mesh:

Year:  2000        PMID: 10793428     DOI: 10.1111/j.1540-8159.2000.tb00821.x

Source DB:  PubMed          Journal:  Pacing Clin Electrophysiol        ISSN: 0147-8389            Impact factor:   1.976


  19 in total

Review 1.  Pacemaker lead complications: when is extraction appropriate and what can we learn from published data?

Authors:  F A Bracke; A Meijer; L M van Gelder
Journal:  Heart       Date:  2001-03       Impact factor: 5.994

2.  Symptomatic occlusion of the access vein after pacemaker or ICD lead extraction.

Authors:  F A Bracke; A Meijer; L M Van Gelder
Journal:  Heart       Date:  2003-11       Impact factor: 5.994

Review 3.  Considerations for cardiac device lead extraction.

Authors:  Oussama Wazni; Bruce L Wilkoff
Journal:  Nat Rev Cardiol       Date:  2016-01-29       Impact factor: 32.419

4.  [2003 yearly report of the German heart pacemaker registry].

Authors:  A Markewitz
Journal:  Herzschrittmacherther Elektrophysiol       Date:  2005-09

Review 5.  Endovascular extraction techniques: Part 2: Complications and indications.

Authors:  F A Bracke; A Meijer; B van Gelder
Journal:  Neth Heart J       Date:  2001-05       Impact factor: 2.380

Review 6.  A Questionable Indication For ICD Extraction After Successful VT Ablation.

Authors:  Luca Segreti; Andrea Di Cori; Giulio Zucchelli; Ezio Soldati; Giovanni Coluccia; Stefano Viani; Luca Paperini; Maria Grazia Bongiorni
Journal:  J Atr Fibrillation       Date:  2015-04-30

7.  Predictors of severe tricuspid regurgitation in patients with permanent pacemaker or automatic implantable cardioverter-defibrillator leads.

Authors:  Mohammad Q Najib; Satya S Vittala; Suresh Challa; Amol Raizada; Fernando J Tondato; Howard R Lee; Hari P Chaliki
Journal:  Tex Heart Inst J       Date:  2013

8.  Implanted endocardial lead characteristics and risk of stroke or transient ischemic attack.

Authors:  Vaibhav R Vaidya; Christopher V DeSimone; Samuel J Asirvatham; Vishnu M Chandra; Amit Noheria; David O Hodge; Joshua P Slusser; Alejandro A Rabinstein; Paul A Friedman
Journal:  J Interv Card Electrophysiol       Date:  2014-04-27       Impact factor: 1.900

9.  Pulmonary embolism in patients with transvenous cardiac implantable electronic device leads.

Authors:  Amit Noheria; Shiva P Ponamgi; Christopher V Desimone; Vaibhav R Vaidya; Christopher A Aakre; Elisa Ebrille; Tiffany Hu; David O Hodge; Joshua P Slusser; Naser M Ammash; Charles J Bruce; Alejandro A Rabinstein; Paul A Friedman; Samuel J Asirvatham
Journal:  Europace       Date:  2015-03-12       Impact factor: 5.214

Review 10.  Pacing Without Wires: Leadless Cardiac Pacing.

Authors:  Michael L Bernard
Journal:  Ochsner J       Date:  2016
View more