Herbert Nägele1, Jolana Lipoldová2, Hanno Oswald3, Gunnar Klein3, Arif Elvan4, Ernst Vester5, Wolfgang Bauer6, Hansjürgen Bondke7, Sebastian Reif8, Claudia Daub9, Frank Menzel10, Jürgen Schrader11, Göran Zach12.
Abstract
AIMS: Intracardiac electrograms (IEGMs) are essential for the assessment of implantable cardioverter-defibrillator (ICD) function. The Biotronik Home Monitoring systems transmit an 'IEGM Online' that is shorter than the full-length programmer IEGM due to technical constrains. The aim of this study was to evaluate the accuracy of the physician's classification of the underlying rhythm based on the second-generation IEGM Online. METHODS ANDEntities:
Keywords: Arrhythmia detection; Home monitoring; Implantable cardioverter-defibrillator; Intracardiac electrogram; Remote monitoring; Telemedicine
Mesh:
Year: 2015 PMID: 25567067 PMCID: PMC4381835 DOI: 10.1093/europace/euu349
Source DB: PubMed Journal: Europace ISSN: 1099-5129 Impact factor: 5.214
Baseline patient characteristics (n = 1537)
| Age (years) | 64 ± 12 |
| Male | 1263 (82) |
| Ejection fractiona (%) | 32 ± 12 |
| NYHA functional class | |
| I | 141 (9) |
| II | 507 (33) |
| III | 481 (31) |
| IV | 18 (1) |
| No heart failure | 99 (6) |
| Not evaluated | 287 (19) |
| Underlying disease | |
| Ischaemic heart disease | 900 (59) |
| Cardiomyopathy | 1006 (66) |
| Diabetes | 452 (30) |
| Renal insufficiency | 353 (23) |
| History of AFIB | 422 (28) |
| Medication | |
| Beta-blocker | 1344 (88) |
| ACE-inhibitor or angiotensin-antagonist | 1270 (83) |
| Diuretic | 1174 (77) |
| Class I or III antiarrhythmic drug | 335 (22) |
| Ca-antagonist | 133 (9) |
| Digitalis | 305 (20) |
| Anticoagulant | 667 (44) |
| Platelet aggregation inhibitor | 816 (53) |
| Secondary prevention ICD indication | 598 (39) |
| Implanted deviceb | |
| Single-chamber ICD | 717 (47) |
| Dual-chamber ICD | 361 (24) |
| CRT-D | 452 (30) |
ACE, angiotensin-converting enzyme; AFIB, atrial fibrillation; CRT-D, cardiac resynchronization therapy defibrillator; ICD, implantable cardioverter-defibrillator; NYHA, New York Heart Association; SD, standard deviation.
aReported in 1236 patients.
bReported in 1530 patients.
Percentage of correct rhythm classifications based on IEGM Online HD
| The underlying rhythm (Investigators' judgement) | No. of episodes based on programmer IEGM | No. (%) of correct classifications based on IEGM online HD |
|---|---|---|
| Total | 1953 | 1803 (92.3) |
| Classification into 6 categories | ||
| Ventricular fibrillation | 106 | 68 (64.2) |
| Ventricular tachycardia | 472 | 443 (93.9) |
| Atrial fibrillation | 816 | 772 (94.6) |
| Other supraventricular tachyarrhythmia | 426 | 395 (92.7) |
| Lead-related oversensing | 40 | 37 (92.5) |
| T-wave oversensing | 93 | 88 (94.6) |
| Device types | ||
| Single-chamber ICDs | 614 | 571 (93.0) |
| Dual-chamber ICDs | 705 | 651 (92.3) |
| CRT-Ds | 628 | 575 (91.6) |
| Rhythm groups as used in RIONI[ | ||
| Ventricular tachyarrhythmia | 578 | 555 (96.0) |
| Supraventricular tachyarrhythmia | 1242 | 1219 (98.1) |
| Oversensing | 133 | 125 (94.0) |
| Total | 1953 | 1899 (97.2) |
CRT-D, cardiac resynchronization therapy defibrillator; HD, high definition; ICD, implantable cardioverter-defibrillator; IEGM, intracardiac electrogram; RIONI, the Reliability of IEGM Online Interpretation study.
Details of rhythm classification by investigators
| Programmer IEGM arrhythmia … | IEGM online HD | |||||
|---|---|---|---|---|---|---|
| VF | VT | AF | Other SVT | Leada | T-wavea | |
| VF | 20 | 17 | 1 | 0 | 0 | |
| VT | 24 | 2 | 1 | 1 | 1 | |
| AF | 3 | 5 | 35 | 1 | 0 | |
| Other SVT | 1 | 13 | 17 | 0 | 0 | |
| Lead-related oversensing | 2 | 0 | 0 | 1 | 0 | |
| T-wave oversensing | 0 | 5 | 0 | 0 | 0 | |
The values in bold indicate equal classifications based on IEGM Online HD and Programmer IEGM.
AF, atrial fibrillation; HD, high definition; IEGM, intracardiac electrogram; SVT, supraventricular tachyarrhythmia; VF, ventricular fibrillation; VT, ventricular tachycardia.
aLead-related or T-wave oversensing.