Jason Callahan1,2, Tomas Kron3,4, Shankar Siva5,6, Nathalie Simoens7,8, Amanda Edgar9, Sarah Everitt10,11,12, Michal E Schneider13, Rodney J Hicks14,15.
Abstract
BACKGROUND: PET/CT scans acquired in the radiotherapy treatment position are typically performed without compensating for respiratory motion. The purpose of this study was to investigate geographic miss of lung tumours due to respiratory motion for target volumes defined on a standard 3D-PET/CT.Entities:
Mesh:
Year: 2014 PMID: 25511904 PMCID: PMC4278238 DOI: 10.1186/s13014-014-0291-6
Source DB: PubMed Journal: Radiat Oncol ISSN: 1748-717X Impact factor: 3.481
Figure 1The top row shows the four different margin expansions applied to the GTV to create four PTV contours. The bottom row shows the types of geographic miss as a result of tumour motion due to respiration investigated.
A comparison of mean 3D-PET/CT defined volumes to 4D-PET/CT defined volumes (total n = 29)
|
|
| |||
|---|---|---|---|---|
|
|
|
|
| |
| GTV | 15.7 | 23.5 | 50% | 0.0013 |
| PTV 5 mm | 38.5 | 53.9 | 40% | 0.0001 |
| PTV 10 mm | 72.8 | 96.1 | 32% | <0.0001 |
| PTV 15 mm | 119.2 | 155.9 | 31% | <0.0001 |
| Asym | 85.5 | 112.0 | 31% | <0.0001 |
| 15 mm vs 10 mm | 119.2 | 96.1 | 19% | <0.0001 |
Figure 2Summary of results for the three types of geographic miss. a: The proportions of patients with a type 1 miss in the four motion groups using the different PTV margins. b: The mean % of the 4D-GTV that was missed by the 3D-PTV when there was a type 1 miss. c: The mean % of the 4D-PTV that was missed by the 3D-PTV. d: The mean coverage of the 4D-PTV with 95% of the prescribed dose.
Type 3 miss: the severity of the Type 3 geographic miss in each of the motion groups based on the mean [±SD] coverage of the 4D-PTV with 95% of the prescribed dose (total n = 29)
|
|
|
|
|
|
|
|
| |
|---|---|---|---|---|---|---|---|---|
|
| Minor | 97 [2.9] | Minor | 97 [1.7] | Minor | 98 [3.3] | Minor | 98 [4.2] |
|
| Moderate | 93 [5.9] | Moderate | 95 [7.9] | Moderate | 95 [4.8] | Minor | 95 [5.4] |
|
| Significant | 86 [8.8] | Moderate | 92 [5.3] | Moderate | 92 [4.4] | Moderate | 92 [3.9] |
|
| Significant | 67 [16.7] | Significant | 76 [10.9] | Significant | 79 [12.0] | Significant | 81 [14.0] |
Figure 3CT scan of a patient with a small lower lobe tumour. The 4D-GTV is shown in red and the 3D-PTV is shown in green. The highlighted red area is the part of the 4D-GTV that was not covered by the 3D-PTV for the margin applied.
Figure 4A CT scan of a patient with a small lower lobe tumour. The 4D-PTV is shown in red and the green area show the 95% isodose line. The solid red area is the part of the 4D-PTV that would not receive the prescribed dose and constitutes a Type 3 geographic miss.