Accuracy and user performance evaluation of a blood glucose monitoring system.

Michael Caswell1, Joy Frank, Maria T Viggiani, Scott Pardo, Nancy Dunne, Mary Ellen Warchal-Windham, Robert Morin.   

Abstract

BACKGROUND: The accuracy of the Contour(®) Plus (Bayer HealthCare LLC, Diabetes Care, Whippany, NJ) blood glucose monitoring system (BGMS) was evaluated in two separate studies.
MATERIALS AND METHODS: In the laboratory study, fingerstick samples from 100 subjects were tested in duplicate using three test strip lots and assessed per International Organization for Standardization (ISO) 15197:2003, Section 7 (≥95% of results within ±15 mg/dL or ±20% of the reference result for samples with glucose concentrations <75 and ≥75 mg/dL, respectively) and ISO 15197:2013, Section 6.3 (≥95% of results within ±15 mg/dL or ±15% of the reference result for samples with glucose concentrations <100 and ≥100 mg/dL, respectively) accuracy criteria. In the clinical trial, 220 subjects with diabetes, naive to the BGMS, tested capillary glucose from fingertip and palm blood samples and completed an ease-of-use questionnaire. BGMS and YSI glucose analyzer results were compared.
RESULTS: In the laboratory study, 100% of results met ISO 15197:2003 and ISO 15197:2013 accuracy criteria. In the clinical trial, 100% and 99.1% of subject fingerstick results and 98.1% and 96.7% of subject palm results met ISO 15197:2003 and ISO 15197:2013 accuracy criteria, respectively. By Parkes Consensus Error Grid analysis, 100% of subject fingerstick results and 98.1% of subject palm results were within Zone A (remainder within Zone B). Questionnaire results showed most subjects found the BGMS easy to use.
CONCLUSIONS: The Contour Plus BGMS meets ISO 15197:2003 and ISO 15197:2013 accuracy criteria in the laboratory and when used by untrained individuals.

Entities:  

Mesh:

Substances:

Year:  2014        PMID: 25494048     DOI: 10.1089/dia.2014.0129

Source DB:  PubMed          Journal:  Diabetes Technol Ther        ISSN: 1520-9156            Impact factor:   6.118


  5 in total

1.  Assessment of System Accuracy, Intermediate Measurement Precision, and Measurement Repeatability of a Blood Glucose Monitoring System Based on ISO 15197.

Authors:  Nina Jendrike; Annette Baumstark; Stefan Pleus; Jochen Mende; Cornelia Haug; Guido Freckmann
Journal:  J Diabetes Sci Technol       Date:  2018-12-14

2.  Accuracy and User Performance of a New Blood Glucose Monitoring System.

Authors:  Leslie Klaff; Pragathi Shelat; Diana Zondorak; Amy Wayland-Smith; Phil Vernes; James M Richardson
Journal:  J Diabetes Sci Technol       Date:  2020-11-26

3.  Accuracy of capillary blood glucose for 50-g glucose challenge test for gestational diabetes screening.

Authors:  Dittakarn Boriboonhirunsarn; Amporn Robkhonburi; Maria Asad-Dehghan
Journal:  Diabetol Int       Date:  2022-02-07

Review 4.  Technologies for Diabetes Self-Monitoring: A Scoping Review and Assessment Using the REASSURED Criteria.

Authors:  Jessica Hanae Zafra-Tanaka; David Beran; Beatrice Vetter; Rangarajan Sampath; Antonio Bernabe-Ortiz
Journal:  J Diabetes Sci Technol       Date:  2021-03-09

5.  User Performance Evaluation of Four Blood Glucose Monitoring Systems Applying ISO 15197:2013 Accuracy Criteria and Calculation of Insulin Dosing Errors.

Authors:  Guido Freckmann; Nina Jendrike; Annette Baumstark; Stefan Pleus; Christina Liebing; Cornelia Haug
Journal:  Diabetes Ther       Date:  2018-03-03       Impact factor: 2.945

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.