Application of Sigma Metrics and Performance Comparison Between Two Biochemistry Analyser and a Blood Gas Analyser for the Determination of Electrolytes.

Yasemin Ustundag-Budak1, Kagan Huysal2.   

Abstract

INTRODUCTION: Electrolytes have a narrow range of biological variation and small changes are clinically significant. It is important to select the best method for clinical decision making and patient monitoring in the emergency room. The sigma metrics model provides an objective method to evaluate the performance of a method. AIM: To calculate sigma metrics for electrolytes measured with one arterial blood gas analyser including two auto-analysers that use different technologies. To identify the best approach for electrolyte monitoring in an emergency setting and the context of routine emergency room workflow.
MATERIALS AND METHODS: The Coefficient of Variation (CV) was determined from Internal Quality Control (IQC). Data was measured from July 2015 to January 2016 for all three analysers. The records of KBUD external quality data (Association of Clinical Biochemists, Istanbul, Turkey) for both Mindray BS-2000M analyser (Mindray, Shenzhen, China) and Architect C16000 (Abbott Diagnostics, Abbott Park, IL) and MLE clinical laboratory evaluation program (Washington, DC, USA) for Radiometer ABL 700 (Radiometer Trading, Copenhagen, Denmark) during the study period were used to determine the bias.
RESULTS: The calculated average sigma values for sodium (-1.1), potassium (3.3), and chloride (0.06) were with the Radiometer ABL700. All calculated sigma values were better than the auto-analysers.
CONCLUSION: The sigma values obtained from all analysers suggest that running more controls and increasing the calibration frequency for electrolytes is necessary for quality assurance.

Entities:  

Keywords:  Chloride; Potassium; Quality control; Quality improvement; Sodium

Year:  2017        PMID: 28384850      PMCID: PMC5376911          DOI: 10.7860/JCDR/2017/23486.9259

Source DB:  PubMed          Journal:  J Clin Diagn Res        ISSN: 0973-709X


  16 in total

1.  Current databases on biological variation: pros, cons and progress.

Authors:  C Ricós; V Alvarez; F Cava; J V García-Lario; A Hernández; C V Jiménez; J Minchinela; C Perich; M Simón
Journal:  Scand J Clin Lab Invest       Date:  1999-11       Impact factor: 1.713

2.  Practical application of Sigma Metrics QC procedures in clinical chemistry.

Authors:  Christian H H Schoenmakers; Andre J M Naus; Henricus J Vermeer; Douwe van Loon; Gerard Steen
Journal:  Clin Chem Lab Med       Date:  2011-07-29       Impact factor: 3.694

3.  Assessing quality on the Sigma scale from proficiency testing and external quality assessment surveys.

Authors:  James O Westgard; Sten A Westgard
Journal:  Clin Chem Lab Med       Date:  2015-09-01       Impact factor: 3.694

4.  Utilizing global data to estimate analytical performance on the Sigma scale: A global comparative analysis of methods, instruments, and manufacturers through external quality assurance and proficiency testing programs.

Authors:  Sten A Westgard
Journal:  Clin Biochem       Date:  2016-03-04       Impact factor: 3.281

5.  Comparison of blood gas, electrolyte and metabolite results measured with two different blood gas analyzers and a core laboratory analyzer.

Authors:  Metin Uyanik; Erdim Sertoglu; Huseyin Kayadibi; Serkan Tapan; Muhittin A Serdar; Cumhur Bilgi; Ismail Kurt
Journal:  Scand J Clin Lab Invest       Date:  2014-11-28       Impact factor: 1.713

6.  An Evaluation of Laboratory Efficiency in Shanghai Emergency by Turn Around Times Level.

Authors:  Yiming Lu; Waiian Leong; Bohua Wei; Ping Yu; Cuicui Wang; Yilin Ying; Tingsong Wang; Jianjing Tong; Dingliang Zhu; Jing Ye
Journal:  J Clin Lab Anal       Date:  2014-08-17       Impact factor: 2.352

7.  Use of a blood gas analyzer and a laboratory autoanalyzer in routine practice to measure electrolytes in intensive care unit patients.

Authors:  Yasemin U Budak; Kagan Huysal; Murat Polat
Journal:  BMC Anesthesiol       Date:  2012-08-03       Impact factor: 2.217

8.  Impact of diuretic therapy-associated electrolyte disorders present on admission to the emergency department: a cross-sectional analysis.

Authors:  Spyridon Arampatzis; Georg-Christian Funk; Alexander Benedikt Leichtle; Georg-Martin Fiedler; Christoph Schwarz; Heinz Zimmermann; Aristomenis Konstantinos Exadaktylos; Gregor Lindner
Journal:  BMC Med       Date:  2013-03-27       Impact factor: 8.775

9.  General characteristics of patients with electrolyte imbalance admitted to emergency department.

Authors:  Arif Kadri Balcı; Ozlem Koksal; Ataman Kose; Erol Armagan; Fatma Ozdemir; Taylan Inal; Nuran Oner
Journal:  World J Emerg Med       Date:  2013

10.  Application of sigma metrics for the assessment of quality assurance using the MQ-2000 PT HbA1c analyzer.

Authors:  Kağan Huysal; Yasemin U Budak
Journal:  Biochem Med (Zagreb)       Date:  2015-10-15       Impact factor: 2.313

View more
  2 in total

1.  Application of Sigma metrics in assessing the clinical performance of verified versus non-verified reagents for routine biochemical analytes.

Authors:  Shuang Cao; Xiaosong Qin
Journal:  Biochem Med (Zagreb)       Date:  2018-06-15       Impact factor: 2.313

2.  Interchangeability of Sodium and Potassium Result Values of Arterial Blood Gas with Laboratory Analyzer: Narrative Review.

Authors:  Yasemin Ustundağ; Kağan Huysal; Şeyda E Ozgunay; Ali R Turkoğlu
Journal:  Indian J Crit Care Med       Date:  2019-01
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.