| Literature DB >> 34816029 |
Kocijancic Marija1, Kraus Frank Bernhard1,2, Ludwig-Kraus Beatrice1.
Abstract
BACKGROUND: Blood-Gas Analyzers (BGAs) are commonly used in parallel with central laboratory analyzers (CLAs). Given the often-divergent results between BGAs and CLAs this study aims to: 1. Determine whether the measurements of potassium (K), sodium (Na), glucose (Glu), lactate (Lact) and total hemoglobin (ctHb) on BGAs and CLAs are interchangeable; 2. Establish reference intervals (RIs) for both analyzer systems using an indirect statistical approach.Entities:
Keywords: Blood-gas analyzers; Central-laboratory analyzers; Interchangeability; POCT; Reference intervals
Year: 2021 PMID: 34816029 PMCID: PMC8591361 DOI: 10.1016/j.heliyon.2021.e08302
Source DB: PubMed Journal: Heliyon ISSN: 2405-8440
Figure 1Passing-Bablok regression for the method comparison the ABL FLEX 90 Plus Blood gas Analyzer with the Roche Cobas c701 CLA for Na, K, Gluc and Lact and with the Sysmex XN-9000 CLA for ctHb. The grey dashed line denotes the line of equivalence and the red line the Passing-Bablok regression fit. The respective regression equations for each comparison and analyte are given in the black rimmed boxes.
Figure 2Bland-Altman plots for the method comparison the ABL FLEX 90 Plus Blood Gas Analyzer with the Roche Cobas c701 CLA for Na, K, Gluc and Lact and with the Sysmex XN-9000 CLA for ctHb. The black dashed line denotes the zero deviation between the BGA and CLA, while the red lines denote the mean relative deviation and its 95% limit of agreement (LoA).
Summary data for the ABL FLEX 90 Plus BGA in comparison with the Roche Cobas c701 CLA for Na, K, Gluc and Lact and with the Sysmex XN-9000 CLA for ctHb. Given are the medians with inter-quartile ranges (IQR), the absolute (mmol/l) and relative (%) bias between BGA and CLA, their statistical sigificane level (P), as well as the corresponding 95% Limit of Agreeement (95% LoA) and the interchangeability criteria ACL, TCL and RiliBÄK.
| Analyte | median CLA (IQR) | median BGA (IQR) | bias | 95% LoA | bias (%) | 95% LoA (%) | P | ACL (%) | TCL (%) | RiliBÄK (%) |
|---|---|---|---|---|---|---|---|---|---|---|
| Na (mmol/l) | 140 (138–142) | 140 (138–142) | 0.1 | -3.8 to 3.9 | 0.03 | -2.7 to 2.8 | 0.558 | 2.5 | 2.5 | 5.0 |
| K (mmol/l) | 4.1 (3.8–4.5) | 4.2 (3.9–4.7) | -0.1 | -0.6 to 0.3 | -2.6 | -12.9 to 7.6 | <0.001 | 3.0 | 3.8 | 8.0 |
| Gluc (mmol/l) | 6.5 (5.6–7.9) | 6.4 (5.5–7.7) | -0.1 | -0.9 to 0.6 | -1.9 | -11.6 to 7.8 | <0.001 | 3.3 | 4.0 | 15.0 |
| Lact (mmol/l) | 1.8 (1.2–2.5) | 1.6 (1.1–2.3) | -0.2 | -1.5 to 1.1 | - 8.8 | -54.4 to 36.7 | <0.001 | 5.5 | 14.7 | 18.0 |
| ctHb (mmol/l) | 7.9 (6.8–8.6) | 8.1 (7.0–8.9) | 0.3 | -0.6 to 1.1 | 3.5 | -8.2 to 15.2 | <0.001 | 1.9 | 2.4 | 6.0 |
Shown are the results of the indirect RI estimation for ABL FLEX 90 Plus BGA, the Roche Cobas c701 CLA (Na, K, Gluc and Lact) and the Sysmex XN-9000 CLA (ctHb). Given are the sample size (N), the lower and upper reference interval (LRI and URI), as well as their respective 95% confidence intervals (95% CI LIR and 95% CI URI). Also listed are the RI recommended by the manufacturers of the respective analytical test and analyzer platform (manufacturer RI).
| Analyte | N | LRI | URI | 95% CI LRI | 95% CI URI | man-fact RI | |
|---|---|---|---|---|---|---|---|
| Na (mmol/l) | CLA | 14425 | 137.0 | 145.0 | 135.1–139.0 | 143.0–147.0 | 136–145 |
| BGA | 7844 | 135.3 | 146.1 | 133.0–137.6 | 143.7–148.5 | 136–145 | |
| K (mmol/l) | CLA | 30090 | 3.43 | 4.87 | 3.30–3.60 | 4.70–5.05 | 3.4–4.5 |
| BGA | 9795 | 3.22 | 4.84 | 3.08–3.36 | 4.65–5.03 | 3.4–4.5 | |
| Gluc (mmol/l) | CLA | 37131 | 4.33 | 7.15 | 4.12–4.54 | 6.84–7.46 | 4.11–6.05 |
| BGA | 26275 | 4.37 | 7.53 | 4.15–4.59 | 7.19–7.87 | 3.89–5.83 | |
| Lac (mmol/l) | CLA | 6139 | 0.60 | n.a. | 0.53–0.66 | n.a. | 0.5–2.2 |
| BGA | 26249 | 0.42 | n.a. | 0.37–0.47 | n.a. | 0.5–1.6 | |
| ctHb (mmol/l) | CLA men | 32921 | 7.64 | 10.58 | 7.35–7.93 | 10.21–10.95 | 8.4–11.1 |
| BGA men | 14259 | 7.57 | 11.21 | 7.25–7.89 | 10.78–11.64 | 8.4–10.9 | |
| CLA women | 36142 | 6.90 | 9.54 | 6.64–7.16 | 9.2–9.88 | 7.1–9.9 | |
| BGA women | 12269 | 6.37 | 10.26 | 6.07–6.67 | 9.83–10.69 | 7.4–9.9 |