Roll and pitch set-up errors during volumetric modulated arc delivery: can adapting gantry and collimator angles compensate?

Nienke A Hoffmans-Holtzer1, Daan Hoffmans, Max Dahele, Ben J Slotman, Wilko F A R Verbakel.   

Abstract

PURPOSE: The purpose of this work was to investigate whether adapting gantry and collimator angles can compensate for roll and pitch setup errors during volumetric modulated arc therapy (VMAT) delivery.
METHODS: Previously delivered clinical plans for locally advanced head-and-neck (H&N) cancer (n = 5), localized prostate cancer (n = 2), and whole brain with simultaneous integrated boost to 5 metastases (WB + 5M, n = 1) were used for this study. Known rigid rotations were introduced in the planning CT scans. To compensate for these, in-house software was used to adapt gantry and collimator angles in the plan. Doses to planning target volumes (PTV) and critical organs at risk (OAR) were calculated with and without compensation and compared with the original clinical plan. Measurements in the sagittal plane in a polystyrene phantom using radiochromic film were compared by gamma (γ) evaluation for 2 H&N cancer patients.
RESULTS: For H&amp;N plans, the introduction of 2°-roll and 3°-pitch rotations reduced mean PTV coverage from 98.7 to 96.3%. This improved to 98.1% with gantry and collimator compensation. For prostate plans respective figures were 98.4, 97.5, and 98.4%. For WB + 5M, compensation worked less well, especially for smaller volumes and volumes farther from the isocenter. Mean comparative γ evaluation (3%, 1 mm) between original and pitched plans resulted in 86% γ < 1. The corrected plan restored the mean comparison to 96% γ < 1.
CONCLUSION: Preliminary data suggest that adapting gantry and collimator angles is a promising way to correct roll and pitch set-up errors of < 3° during VMAT for H&amp;N and prostate cancer.

Entities:  

Mesh:

Year:  2014        PMID: 25430597     DOI: 10.1007/s00066-014-0766-2

Source DB:  PubMed          Journal:  Strahlenther Onkol        ISSN: 0179-7158            Impact factor:   3.621


  20 in total

1.  Beam-centric algorithm for pretreatment patient position correction in external beam radiation therapy.

Authors:  Supratik Bose; Himanshu Shukla; Jonathan Maltz
Journal:  Med Phys       Date:  2010-05       Impact factor: 4.071

2.  Pitch, roll, and yaw variations in patient positioning.

Authors:  Adeel Kaiser; Timothy E Schultheiss; Jeffrey Y C Wong; David D Smith; Chunhui Han; Nayana L Vora; Richard D Pezner; Yi-Jen Chen; Eric H Radany
Journal:  Int J Radiat Oncol Biol Phys       Date:  2006-09-01       Impact factor: 7.038

3.  A novel method to correct for pitch and yaw patient setup errors in helical tomotherapy.

Authors:  Sarah A Boswell; Robert Jeraj; Kenneth J Ruchala; Gustavo H Olivera; Hazim A Jaradat; Joshua A James; Alonso Gutierrez; Dave Pearson; Gary Frank; T Rock Mackie
Journal:  Med Phys       Date:  2005-06       Impact factor: 4.071

4.  Volumetric modulated arc therapy: IMRT in a single gantry arc.

Authors:  Karl Otto
Journal:  Med Phys       Date:  2008-01       Impact factor: 4.071

5.  RapidArc planning and delivery in patients with locally advanced head-and-neck cancer undergoing chemoradiotherapy.

Authors:  Patricia Doornaert; Wilko F A R Verbakel; Michael Bieker; Ben J Slotman; Suresh Senan
Journal:  Int J Radiat Oncol Biol Phys       Date:  2010-04-24       Impact factor: 7.038

6.  Residual translational and rotational errors after kV X-ray image-guided correction of prostate location using implanted fiducials.

Authors:  Reinhold Graf; Dirk Boehmer; Volker Budach; Peter Wust
Journal:  Strahlenther Onkol       Date:  2010-09-30       Impact factor: 3.621

7.  Target volume coverage and dose to organs at risk in prostate cancer patients. Dose calculation on daily cone-beam CT data sets.

Authors:  P Hüttenrauch; M Witt; D Wolff; S Bosold; R Engenhart-Cabillic; J Sparenberg; H Vorwerk; K Zink
Journal:  Strahlenther Onkol       Date:  2014-01-16       Impact factor: 3.621

8.  Volumetric intensity-modulated arc therapy vs. conventional IMRT in head-and-neck cancer: a comparative planning and dosimetric study.

Authors:  Wilko F A R Verbakel; Johan P Cuijpers; Daan Hoffmans; Michael Bieker; Ben J Slotman; Suresh Senan
Journal:  Int J Radiat Oncol Biol Phys       Date:  2009-05-01       Impact factor: 7.038

9.  Precision of image-guided radiotherapy (IGRT) in six degrees of freedom and limitations in clinical practice.

Authors:  Matthias Guckenberger; Juergen Meyer; Juergen Wilbert; Kurt Baier; Otto Sauer; Michael Flentje
Journal:  Strahlenther Onkol       Date:  2007-06       Impact factor: 3.621

10.  Clinical evaluation of a commercial surface-imaging system for patient positioning in radiotherapy.

Authors:  F Stieler; F Wenz; D Scherrer; M Bernhardt; F Lohr
Journal:  Strahlenther Onkol       Date:  2012-11-10       Impact factor: 3.621

View more
  1 in total

1.  Hypo-fractionated SBRT for localized prostate cancer: a German bi-center single treatment group feasibility trial.

Authors:  Ping Jiang; Katja Krockenberger; Reinhard Vonthein; Jane Tereszczuk; Arne Schreiber; Sebastian Liebau; Stefan Huttenlocher; Detlef Imhoff; Panagiotis Balermpas; Christian Keller; Kathrin Dellas; Rene Baumann; Claus Rödel; Guido Hildebrandt; Klaus-Peter Jünemann; Alex S Merseburger; Alan Katz; Andreas Ziegler; Oliver Blanck; Jürgen Dunst
Journal:  Radiat Oncol       Date:  2017-08-18       Impact factor: 3.481

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.