Comparison of CT-Number and Gray Scale Value of Different Dental Materials and Hard Tissues in CT and CBCT.

Naghmeh Emadi1, Yaser Safi2, Alireza Akbarzadeh Bagheban3, Saeed Asgary4.   

Abstract

INTRODUCTION: Computed tomography (CT) and cone-beam CT (CBCT) are valuable diagnostic aids for many clinical applications. This study was designed to compare the gray scale value (GSV) and Hounsfield unit (HU) of selected dental materials and various hard tissues using CT or CBCT. METHODS AND MATERIALS: Three samples of all test materials including amalgam (AM), composite resin (CR), glass ionomer (GI), zinc-oxide eugenol (ZOE), calcium-enriched mixture (CEM) cement, AH-26 root canal sealer (AH-26), gutta-percha (GP), Coltosol (Col), Dycal (DL), mineral trioxide aggregate (MTA), zinc phosphate (ZP), and polycarbonate cement (PC) were prepared and scanned together with samples of bone, dentin and enamel using two CBCT devices, Scanora 3D (S3D) and NewTom VGi (NTV) and a spiral CT (SCT) scanner (Somatom Emotion 16 multislice spiral CT);. Subsequently, the HU and GSV values were determined and evaluated. The data were analyzed by the Kruskal-Wallis and Mann-Whitney U tests. The level of significance was determined at 0.05.
RESULTS: There were significant differences among the three different scanners (P<0.05). The differences between HU/GSV values of 12 selected dental materials using NTV was significant (P<0.05) and for S3D and SCT was insignificant (P>0.05). All tested materials showed maximum values in S3D and SCT (3094 and 3071, respectively); however, bone and dentin showed low/medium values (P<0.05). In contrast, the tested materials and tissues showed a range of values in NTV (366 to15383; P<0.05).
CONCLUSION: Scanner system can influence the obtained HU/GSV of dental materials. NTV can discriminate various dental materials, in contrast to S3D/SCT scanners. NTV may be a more useful diagnostic aid for clinical practice.

Entities:  

Keywords:  Computed Tomography; Cone-Beam Computed Tomography; Dental Materials; Dentistry; Gray Scale Value; Hounsfield Unit; Spiral Computed Tomography

Year:  2014        PMID: 25386210      PMCID: PMC4224767     

Source DB:  PubMed          Journal:  Iran Endod J        ISSN: 1735-7497


Introduction

American Dental Association (ADA) Council on Dental Materials and Devices released a statement in 1981 that declared the optical density as a desirable requirement for restorative materials [1]. It is generally accepted that dental materials should have sufficient radiopacity to be detectable against enamel and dentin backgrounds in dental imaging. This characteristic is also helpful for the assessment of restorations as well as endodontic fillings, detection of secondary caries, evaluation of marginal defects, restoration contour and other defects on radiographs [2]. According to ISO standard the commonly used techniques for the assessment of radiographic density include digital radiography with occlusal film using aluminum step-wedge [3, 4]. Computed tomography (CT) was introduced in 1970 by Hounsfield [5]. For image display each pixel is assigned a CT number (Hounsfield units; HU) representing tissue density. HU ranges from -1000 to +1000; however, some newer CT scans have a range of up to 4000 HU [6]. The technology was introduced in 1978 for bone density [7]. Several studies have evaluated this object and their method/results are considered to be the gold standard for evaluation of the density of tissues [8]. Test materials and their composition Scanner settings and software applied for spiral computed tomography (SCT) and two cone-beam computed tomography (CBCT) devices Cone-beam computed tomography (CBCT) was first used for angiography in 1982 and since the late 1990s it has also been utilized for dentistry [6]. There are some controversies surrounding the reliability of density measurement by CBCT systems; however, the studies have mostly focused on bone density and imaging parameters [9, 10]. To date, no study has evaluated the density of endodontic and restorative materials using CBCT. The objective of the present laboratory study was to compare the CT number (HU) and gray scale value (GSV) of selected dental materials as well as various hard tissues using two CBCT devices and one CT scanner.

Methods and Materials

Holes measuring 3 mm in diameter and 2 mm in height were prepared in 36 resin molds (3 molds for each test material). All the dental materials including amalgam (AM; Cinalux, Faghihi Dental Co., Iran), Composite Resin (CR; Valux plus, 3M Dental Products, USA), Glass Ionomer (GI; Fuji II, GC Corporation, Tokyo, Japan), Zinc-Oxide Eugenol (ZOE; Kemdent; Associated Dental Products, UK), Calcium-Enriched Mixture (CEM Cement; BioniqueDent, Tehran, Iran), AH-26 sealer (AH-26; Dentsply, Tulsa Dental, Tulsa, OK, USA), gutta-percha (GP; Ariadent, Tehran, Iran), Coltosol (Col; Ariadent, Tehran, Iran), Dycal (DL; Dentsply, Tulsa Dental, Tulsa, OK), ProRoot MTA (MTA; Dentsply, Tulsa Dental, Tulsa, OK, USA), zinc-phosphate cement (ZP; Hoffmann Dental Manufaktur GmbH, Berlin, Germany) and polycarbonate cement (PC; Harvard Dental GmbH, Berlin, Germany) were prepared according to the manufactures’ instructions and placed in molds. Three extracted premolars, as well as cortical/spongy bone (from the posterior of maxilla) were scanned to evaluate the density of enamel, dentin and bone. The components of the materials are listed in Table 1.
Table 1

Test materials and their composition

Products Components
Amalgam Ag=49%, Sn=29%, Cu=22%, Hg=29%
Composite Resin Bis-GMA, TEGDMA, zirconia, silica fillers
Glass Ionomer Aluminum fluorosilicate glass, ZnO (powder), polyacrylic acid, HEMA, water, photo initiator (liquid)
Zinc-Oxide Eugenol Zinc-oxide (powder), eugenol or carboxylic acid and inert filler (liquid)
CEM Cement Calcium-silicate, calcium-phosphate, calcium-oxide, calcium-salts, barium-sulfate and zirconium
AH-26 Sealer Bismuth-oxide, calcium-hydroxide, hexamethylenetetramine, titanium-dioxide and bis-phenol
Gutta-Percha Poly-isoprene rubber, zinc-oxide, barium-sulfate, coloring agent
Coltosol Zinc-oxide, zinc-sulphate, calcium-sulphate, dibutyl-phethalate, polyvinyl-acetate chloride copolymer, peppermint
Dycal Butyene glycol disaliclate, zinc-oxide, calcium-phosphate, calcium-tungstate, iron-oxide (base), calcium-hydroxide, sulfonamide, zinc-oxide, titanium-dioxide, zinc-stearate, iron-oxide pigments (catalyst)
MTA Portland cement, bismuth-oxide and gypsum
Zinc Phosphate Cement Zinc-oxide, magnesium-oxide (powder), o-phosphoric acid (liquid)
Poly Carbonate Cement Zn, Mg and Al oxides, boric acid (powder), acrylic acid, maleic acid anhydride, distilled water (liquid)
Scans were obtained from tested materials using the NewTom VGi (NTV; QR SRL Co., Verona, Italy), Scanora 3D (S3D; Soredex, Helsinki, Finland) and multislice spiral CT scanner (SCT; Somatom Emotion; Siemens Medical Solutions, Forchheim, Germany) (Table 2). Density of various dental materials was determined using the GSV and HU obtained from CBCT and SCT scanners, respectively.
Table 2

Scanner settings and software applied for spiral computed tomography (SCT) and two cone-beam computed tomography (CBCT) devices

Equipment setting Somatom emotion 16 Scanora 3D NewTom VGi
Kilo voltage peak 130 kVp90 kVp110 kVp
Milliamperage 139 mA12 mA0.56 mA
Field of view 50×50 mm75×100 mm80×120 mm
Software Syngo FastViewOnDemand 3DNTT viewer
To overcome the probable inhomogeneous nature of the samples, each final HU/GSV of tested material was based on the average of 36 records; each sample (n=3) was divided into four parts, and three measurements of HU/GSV were recorded for each part. The data were collected and recorded by one radiologist. The softwares used for image analysis included Syngo FastView (AG 2004; Siemens, Munich, Germany), OnDemand 3D (Cybermed Inc, Irvine, CA) and NTT viewer (NTT Software Corporation, Yokohama, Japan) for SCT, S3D and NTV, respectively. The collected data were analyzed statistically with the Kruskal-Wallis test and Mann-Whitney U test (SPSS version 16.0, SPSS, Chicago, IL, USA). The level of significance was set at 0.05. Mean, standard error (SE), Minimum (Min) and Maximum (Max) of gray scale values (GSV) of experimental materials P-value obtained by the Mann-Whitney U test for pairwise comparison of groups (*) indicates significant differences (P<0.001); and (¥) indicates nonsignificant differences (P>0.05)

Results

AM and AH-26 had the highest GSV among the tested dental materials. ZP, GP and ZOE stood in the second rank, followed by PC and MTA. Except for cortical bone, spongy bone and dentin, HU/GSV of tested materials in SCT scan and S3D was equal to the maximum value demonstrated by the devices (3071 and 3094, respectively). In NTV, AM and AH-26 had the highest GSV, which was equivalent to 15383. The obtained values of other tested materials were all below this limit. In samples with HU/GSV below the maximum threshold of the device, S3D and NTV revealed no significant differences in comparison with SCT (P>0.05). The result of the tested groups showed inequality of variances, which was largely due to the absence of disparity in AM and AH-26 groups, the nonparametric Kruskal-Wallis test were used for the inter-group comparisons. Significant differences were observed between the groups scanned with the NTV device in GSV (P<0.05). Table 3 demonstrates the GSV of the 16 experimental materials.
Table 3

Mean, standard error (SE), Minimum (Min) and Maximum (Max) of gray scale values (GSV) of experimental materials

Materials Mean SE Min Max
Amalgam 15383.00.015383.015383.0
AH-26 Sealer 15383.00.015383.015383.0
Zinc Phosphate Cement 13987.342.812789.014322.0
Gutta-Percha 13964.447.213452.014424.0
Zinc-Oxide Eugenol 13860.867.813156.014632.0
Poly Carbonate Cement 12661.944.412024.013045.0
Mineral Trioxide Aggregate 11270.336.510981.011956.0
Coltosol 8616.831.48142.08971.0
Calcium-Enriched Mixture 8506.522.98307.08876.0
Composite Resin 6896.014.56719.07076.0
Glass Ionomer 6391.314.76207.06543.0
Dycal 4635.018.24423.04906.0
Enamel 4358.911.34187.04490.0
Dentin 2583.717.22403.02764.0
Cortical bone 1766.234.61470.02134.0
Spongy bone 541.617.8366.0745.0
Pairwise comparison of groups with nonparametric Mann-Whitney U test detected statistically significant differences between majorities of tested materials (Table 4). Due to the absence of variation in the AM and AH-26 groups, these two materials were not compared with others.
Table 4

P-value obtained by the Mann-Whitney U test for pairwise comparison of groups

Materials GI ZOE CEM GP CR DL MTA ZP PC
Composite Resin (CR) *********
Glass Ionomer (GI) -********
Zinc-Oxide Eugenol (ZOE) *-*¥***¥*
Calcium-Enriched Mixture (CEM) **-******
Gutta-Percha (GP) ***-***¥*
Coltosol (Col) ****-****
Dycal (DL) *****-***
Mineral Trioxide Aggregate (MTA) ******-**
Zinc Phosphate Cement (ZP) *******-*
Poly Carbonate Cement (PC) ********-

(*) indicates significant differences (P<0.001); and (¥) indicates nonsignificant differences (P>0.05)

Discussion

In the present study, we used one CT device (SCT) and two CBCT systems (NTV and S3D) for determination of HU/GSV of various dental materials. Our results clearly demonstrated that the choice of imaging device could affect the measured HU/GSV of a dental material. Currently, conventional intraoral radiography is the most commonly used system for the evaluation of the technical quality of the orthograde as well as retrograde filling materials; however, CBCT technology has recently gained popularity in endodontic profession [11]. Contrary to S3D and SCT scanners, the NTV device could precisely discriminate various dental materials. According to ANSI/ISO standards, radiopacity of endodontic materials using conventional radiography should be higher than tooth structure and comparable with that of the determined thickness of aluminum. For example, the radiopacity of 1 mm of dentin and aluminum is equal but an endodontic sealer should present radiopacity correspond to at least 3 mm aluminum; however, differences in the step-wedge of aluminum alloy as well as device/test setting (i.e. exposure time, kVp, mA, focal film distance, film speed, imaging technique, and developing process) can affect the outcome and lead to inhomogeneous results [12]. On the other hand, there are no specific standards for HU/GSV of dental materials using CT/CBCT up to the present moment. Several studies have evaluated the GSV and bone density measurements by CBCT and have yielded conflicting results [8, 13, 14]. Some reported a higher bone density value obtained by CBCT than by SCT and found it to be unreliable [13]. However, other investigations achieved a high level of correlation (r=0.965) of density between CBCT and SCT [15]. Furthermore, evidence suggested that HU value of the respective tissues could be obtained from the GSV by using linear attenuation coefficient and its application in an equation and reported small differences with actual HU values [16]. Our results demonstrated that using NTV, AM and AH-26 root canal sealer had the highest GSV among the tested dental materials. ZP cement, GP and ZOE took the second rank, followed by PC cement and MTA. Similar results have been reported from some recent studies by the step-wedge method [17].

Conclusion

Under the conditions of this study, the results demonstrated that the tested dental materials had different GSVs with the NewTom VGi CBCT device; compared to Scanora 3D CBCT and Somatom Emotion spiral CT scanner. It appears that NewTom VGi is a useful diagnostic imaging system for detecting the various dental materials.
  14 in total

1.  Deriving Hounsfield units using grey levels in cone beam computed tomography.

Authors:  P Mah; T E Reeves; W D McDavid
Journal:  Dentomaxillofac Radiol       Date:  2010-09       Impact factor: 2.419

2.  Evaluation of voxel values in mandibular cancellous bone: relationship between cone-beam computed tomography and multislice helical computed tomography.

Authors:  Munetaka Naitoh; Akiko Hirukawa; Akitoshi Katsumata; Eiichiro Ariji
Journal:  Clin Oral Implants Res       Date:  2009-02-25       Impact factor: 5.977

3.  Bone density: comparative evaluation of Hounsfield units in multislice and cone-beam computed tomography.

Authors:  Isabela Maria de Carvalho Crusoé Silva; Deborah Queiroz de Freitas; Glaucia Maria Bovi Ambrosano; Frab Norberto Bóscolo; Solange Maria Almeida
Journal:  Braz Oral Res       Date:  2012 Nov-Dec

4.  Influence of cone beam CT scanning parameters on grey value measurements at an implant site.

Authors:  A Parsa; N Ibrahim; B Hassan; A Motroni; P van der Stelt; D Wismeijer
Journal:  Dentomaxillofac Radiol       Date:  2012-08-29       Impact factor: 2.419

5.  How accurate is CBCT in measuring bone density? A comparative CBCT-CT in vitro study.

Authors:  Michele Cassetta; Luigi Vito Stefanelli; Andrea Pacifici; Luciano Pacifici; Ersilia Barbato
Journal:  Clin Implant Dent Relat Res       Date:  2013-01-07       Impact factor: 3.932

6.  Evaluation of the radiopacity of luting cements by digital radiography.

Authors:  Subutay Han Altintas; Tahsin Yildirim; Saadettin Kayipmaz; Aslihan Usumez
Journal:  J Prosthodont       Date:  2012-10-25       Impact factor: 2.752

Review 7.  Cone beam computed tomography in endodontics.

Authors:  Conor Durack; Shanon Patel
Journal:  Braz Dent J       Date:  2012

8.  Bone quality evaluation: comparison of cone beam computed tomography and subjective surgical assessment.

Authors:  Jayasankar V Valiyaparambil; Ibrahim Yamany; Denise Ortiz; David M Shafer; David Pendrys; Martin Freilich; Sanjay M Mallya
Journal:  Int J Oral Maxillofac Implants       Date:  2012 Sep-Oct       Impact factor: 2.804

9.  The desirability of using radiopaque plastics in dentistry: a status report. Council on Dental Materials, Instruments, and Equipment.

Authors: 
Journal:  J Am Dent Assoc       Date:  1981-03       Impact factor: 3.634

10.  Variability of dental cone beam CT grey values for density estimations.

Authors:  R Pauwels; O Nackaerts; N Bellaiche; H Stamatakis; K Tsiklakis; A Walker; H Bosmans; R Bogaerts; R Jacobs; K Horner
Journal:  Br J Radiol       Date:  2013-01       Impact factor: 3.039

View more
  2 in total

1.  Methodology to Produce Specimen-Specific Models of Vertebrae: Application to Different Species.

Authors:  Fernando Y Zapata-Cornelio; Gavin A Day; Ruth H Coe; Sebastien N F Sikora; Vithanage N Wijayathunga; Sami M Tarsuslugil; Marlène Mengoni; Ruth K Wilcox
Journal:  Ann Biomed Eng       Date:  2017-07-25       Impact factor: 3.934

2.  Diagnostic accuracy of cone-beam computed tomography in detecting secondary caries under composite fillings: an in vitro study.

Authors:  Elif Yildizer Keris; Oguzhan Demirel; Melih Ozdede; Bulent Altunkaynak; Ilkay Peker
Journal:  J Istanb Univ Fac Dent       Date:  2017-01-02
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.