PTV margin definition in hypofractionated IGRT of localized prostate cancer using cone beam CT and orthogonal image pairs with fiducial markers.

Christoph Oehler1,2, Stephanie Lang3, Peter Dimmerling4, Christian Bolesch5, Stephan Kloeck6, Alessandra Tini7, Christoph Glanzmann8, Yousef Najafi9, Gabriela Studer10, Daniel R Zwahlen11,12.   

Abstract

PURPOSE: To evaluate PTV margins for hypofractionated IGRT of prostate comparing kV/kV imaging or CBCT. PATIENTS AND METHODS: Between 2009 and 2012, 20 patients with low- (LR), intermediate- (IR) and high-risk (HR) prostate cancer were treated with VMAT in supine position with fiducial markers (FM), endorectal balloon (ERB) and full bladder. CBCT's and kV/kV imaging were performed before and additional CBCT's after treatment assessing intra-fraction motion. CTVP for 5 patients with LR and CTVPSV for 5 patients with IR/HR prostate cancer were contoured independently by 3 radiation oncologists using MRI. The van Hark formula (PTV margin =2.5Σ +0.7σ) was applied to calculate PTV margins of prostate/seminal vesicles (P/PSV) using CBCT or FM.
RESULTS: 172 and 52 CBCTs before and after RT and 507 kV/kV images before RT were analysed. Differences between FM in CBCT or in planar kV image pairs were below 1 mm. Accounting for both random and systematic uncertainties anisotropic PTV margins were 5-8 mm for P (LR) and 6-11 mm for PSV (IR/HR). Random uncertainties like intra-fraction and inter-fraction (setup) uncertainties were of similar magnitude (0.9-1.4 mm). Largest uncertainty was introduced by CTV delineation (LR: 1-2 mm, IR/HR: 1.6-3.5 mm). Patient positioning using bone matching or ERB-matching resulted in larger PTV margins.
CONCLUSIONS: For IGRT CBCT or kV/kV-image pairs with FM are interchangeable in respect of accuracy. Especially for hypofractionated RT, PTV margins can be kept in the range of 5 mm or below if stringent daily IGRT, ideally including prostate tracking, is applied. MR-based CTV delineation optimization is recommended.

Entities:  

Mesh:

Year:  2014        PMID: 25384898      PMCID: PMC4229608          DOI: 10.1186/s13014-014-0229-z

Source DB:  PubMed          Journal:  Radiat Oncol        ISSN: 1748-717X            Impact factor:   3.481


Background

Image-guided radiation therapy (IGRT) is the preferred method for curative treatment of localized prostate cancer and is associated with improved outcome and reduced toxicity [1]. IGRT using cone-beam computed tomography (CBCT) or kilovoltage (kV/kV) gold fiducial marker (FM) imaging in combination with conformal radiation therapy (RT) techniques such as intensity-modulated radiotherapy (IMRT) or volumetric modulated arc therapy (VMAT) potentially allow smaller treatment margins and to escalate dose to the prostate [2]. PTV margins have been reduced from 10 mm in the pre-IGRT-era to 3-5 mm in the IGRT-era [3,4]. There is little consensus on the magnitude of PTV margin for RT of the prostate mainly due to inter-institutional technical issues and the preferred IGRT modality [5]. To compute the magnitude of a PTV-margin that allows the CTV receive the prescribed dose with a clinically acceptable and specified probability, statistics of all uncertainties in the treatment process chain should be known [6,7]. Geometrical uncertainties in RT include both treatment preparation variations and execution uncertainties. They both can be systematic such as equipment maladjustments, planning setup uncertainties, and target volume delineation or random such as treatment setup uncertainties, inter- and intra-fraction organ motion [7]. The aim of this work was to investigate PTV margin determination in the IGRT-era with endorectal balloon (ERB) calculating systematic and random treatment uncertainties. Furthermore, the IGRT modalities CBCT and kV/kV FM imaging were compared with each other in the context of PTV margin determination.

Patients and methods

Study design and patient selection

From January 2009 to April 2012, 20 non-consecutive patients with histologically proven low- (LR), intermediate- (IR) or high-risk (HR) adenocarcinoma of the prostate were treated according to the CHHIP (Conventional or Hypofractionated High Dose Intensity Modulated Radiotherapy for Prostate Cancer) protocol [8], and according to randomization VMAT (RapidArc®) was used to a dose of 74Gy/37f (n = 14), 60Gy/20f (n = 1) or 57Gy/19f (n = 5) at the University Hospital Zürich, Switzerland [9]. Hypofractionation schedules were used only for patients randomized within the CHHIP trial (open between 2010 and 2011 at Zurich) and not off-study, since hypofractionation was not standard at that time. Planning CT scans, kV/kV images and CBCT’s of the 20 patients were retrospectively analysed with ethics approval (Department of Radiation Oncology, University Hospital Zurich).

Target definition and treatment planning

Two weeks prior to CT simulations, 3 FM were placed into the prostate under rectal ultrasound guidance. Non-contrasted planning CT simulation with axial slices thicknesses of 2.0 mm was performed in supine position by use of a leg holder immobilization device, empty rectum with endorectal balloon (ERB) and full bladder. Reference points were tattooed on the skin. CTV was defined as prostate (P) only for LR, prostate plus 1/3 of the seminal vesicles (PSV) for IR or the prostate plus the seminal vesicles for HR cancer patients. In order to evaluate the delineation error of prostate only or prostate/seminal vesicles, the CTVP for 5 pts with LR and CTVPSV for 5 pts with HR prostate cancer were contoured independently by 3 experienced radiation oncologists using non fused MR images with rectal coil. No formal contouring training was performed.

Treatment

Daily localization of the prostate position was performed using FM before treatment. CBCT was performed on day 1 to 3, then once every week (hypofractionation: 6 × 6 CBCT, normofractionation: 14 × 9-10 CBCT). Additional in-between CBCT were not used for evaluation. For 10 patients (57Gy/19f (n = 1), 74Gy/37f (n = 9)) an additional CBCT was performed after each fraction to assess intra-fractional motion once a week (1 × 4 CBCT, 9 × 7 CBCT). Due to patient’s distress, failure of CBCT or correlation function, not all post-treatment CBCT could be used for evaluation. On day 1 the isocenter after CBCT based corrections was marked on the patient skin. Online isocenter positioning was performed based on FM using CBCT – as indicated above - or kV image pairs. kV/kV images were performed before each fraction except when a weekly or in-between CBCT was applied. On days with post-treatment CBCT a kV/kV image pair was performed for comparison reason (14 × 37 kV/kV +5 × 19 kV/kV +1 × 20 kV/kV =633 kV/kV – 172 kV/kV (pre-CBCT days) +67 kV/kV (post-CBCT days) =528 kV/kV – 21 kV/kV (in-between CBCT) =507 kV/kV).

Statistical analysis

The van Herk formula M =2.5Σ +0.7σ was used for PTV margin calculation, where Σ represents the systematic and σ the random uncertainty. Random uncertainties are statistical fluctuations whereas systematic uncertainties are often due to a problem which persists throughout the entire treatment. We calculated systematic uncertainty for intra-fractional motion, contouring as well as setup. Random uncertainties were calculated for intra-fractional motion and patient setup. Total Σ was then calculated: Σ = (Σ2contour + Σ2patient setup + Σ2intrafraction motion)1/2 as well as total σ: σ = (σ2patient setup + σ2intrafraction motion)1/2. Statistical analysis was performed with MATLAB, version 11 (Mathworks, MA, U.S.A.). We used a σp – where σp is the standard deviation of the normal dose penumbra - of 3 mm assuming the worst case (cranial-caudal direction) because for VMAT σp is larger in the left/right as well as anterior/posterior direction compared to the cranial/caudal direction. Furthermore, changing the σp influences only the random error and results in a small change of the margin. The online match was based on FM. Additionally offline matches based on bony anatomy and ERB (kV/kV image pairs and CBCT), as well as soft prostate tissue (CBCT) were performed to compare different matching techniques. The setup uncertainty was calculated for each of the different matching techniques in respect to the online performed FM match. To assess the difference between FM match in kV image pairs and CBCT the two imaging modalities were performed strictly consecutively and shifts were compared. To assess intra-fraction prostate motion, differences between FM position in pre-treatment and post-treatment CBCTs were established. The contouring uncertainties of the CTVP and the CTVPSV were determined by comparing intersection and union of 2 corresponding contours using a self-developed analysis script in Matlab (Mathworks): The prostate volume was divided into 6 subsections (anterior, posterior, left, right, cranial and caudal) (Figure 1). For each subsection the mean distance between the union and the intersection was calculated. The mean and standard deviation (SD) between the pairwise comparisons were calculated. Wilcoxon-Mann-Whitney-Test was used to compare contouring uncertainties for P and PSV.
Figure 1

Depiction of the mean differences in vertical, lateral and longitudinal directions between the intersections (blue) and the unions (pink) of the three CTVp as contoured by three radiation oncologists. The prostate volume was divided into 6 subsections (anterior, posterior, left, right, cranial, caudal).

Depiction of the mean differences in vertical, lateral and longitudinal directions between the intersections (blue) and the unions (pink) of the three CTVp as contoured by three radiation oncologists. The prostate volume was divided into 6 subsections (anterior, posterior, left, right, cranial, caudal).

Results

Setup (inter-fractional) uncertainty

172 and 52 CBCT before and after RT and 507 kV/kV images before RT were analyzed. The setup uncertainty was calculated for the matching on different surrogates in kV images pairs as well as CBCT using CBCT FM match as reference. The mean systematic and random setup uncertainties for each modality are shown in Tables 1 and 2. The difference in setup using FM between CBCT and kV/kV was usually below 2 mm (Figure 2). Matching on FM in kV/kV images or matching on soft tissue in CBCT lead to systematic and random uncertainties below 1 mm. Bony anatomy match as well as balloon match increased the systematic as well as the random uncertainty. This was more pronounced if kV/kV image pairs were used compared to the use of CBCT. Large systematic as well as random uncertainties in longitudinal direction were found with the use of the ERB as a surrogate for prostate position since reproducible placing of the ERB was not possible.
Table 1

Systematic Σ and random σ setup error depending on surrogate used for matching kV/kV images to the reference CT

Σ (mm) σ (mm)
Bonesvrt1.781.97
lng1.392.30
lat0.770.85
Balloonvrt1.531.36
lng3.203.15
lat1.701.35
FMvrt0.880.71
lng1.160.46
lat0.460.37
Table 2

Systematic Σ and random σ setup error depending on surrogate used for matching the CBCT to the reference CT

Σ (mm) σ (mm)
Bonesvrt1.921.87
lng1.721.69
lat0.610.90
Balloonvrt1.681.82
lng1.201.66
Soft tissuevrt0.771.09
lng0.360.17
0.490.93

Fiducial marker matching in CBCT was taken as a reference.

Figure 2

Bar diagram of differences in vertical (vrt), longitudinal (lng) and lateral (lat) direction between CBCT and kV/kV setup matching on fiducial markers (N =10).

Systematic Σ and random σ setup error depending on surrogate used for matching kV/kV images to the reference CT Systematic Σ and random σ setup error depending on surrogate used for matching the CBCT to the reference CT Fiducial marker matching in CBCT was taken as a reference. Bar diagram of differences in vertical (vrt), longitudinal (lng) and lateral (lat) direction between CBCT and kV/kV setup matching on fiducial markers (N =10).

Intra-fractional motion uncertainty

For intra-fractional motion uncertainty calculation of the prostate FM shift of pre- and post-treatment CBCT’s were compared with each other (time between 3 to 6 min). Maximum FM shifts reached 5 mm in vertical and 8 mm in longitudinal direction (Figure 3). The systematic intra-fractional uncertainty was 1.39 mm (vrt), 1.36 mm (lng), 0.92 mm (lat) and the day to day random uncertainty was 1.62 mm (vrt), 1.39 mm (lng), 0.97 mm (lat), respectively.
Figure 3

Bar diagram of intra-fraction motion in vertical (vrt), longitudinal (lng) and lateral (lat) direction during treatment (N =10). Intra-fraction motion was determined by fiducial marker difference between pre- and post- CBCT.

Bar diagram of intra-fraction motion in vertical (vrt), longitudinal (lng) and lateral (lat) direction during treatment (N =10). Intra-fraction motion was determined by fiducial marker difference between pre- and post- CBCT.

Contouring error

Systematic contouring uncertainties were larger for PSV than for P (p =0.012). The mean systematic uncertainty was largest in the longitudinal axis as shown in Table 3. In vertical direction the uncertainty was found to be larger in anterior (1.98 mm (P), 3.49 mm (PSV)) than posterior (1.04 mm (P), 1.58 mm (PSV) direction.
Table 3

Average contouring errors for prostate (P) or prostate with seminal vesicles (PSV) in anterior (vrt/anterior), posterior (vrt/posterior), longitudinal (long) and lateral (lat) direction

P (mm) PSV (mm)
vrt1.522.57
vrt/anterior1.983.49
vrt/posterior1.041.58
Long2.033.20
Lat1.682.52
Average contouring errors for prostate (P) or prostate with seminal vesicles (PSV) in anterior (vrt/anterior), posterior (vrt/posterior), longitudinal (long) and lateral (lat) direction

PTV margin (according to van Herk formula)

The calculated PTV margins can be found in Tables 4 and 5. The margin for P only was calculated to be 7.2 mm in anterior, 5.5 mm posterior, 7.1 mm longitudinal, and 5.5 mm in lateral direction. PTV margins were larger for PSV volumes than for P only volumes due to the contouring differences. The difference was largest in longitudinal direction (2.5 mm). The contouring uncertainty contributed most to the PTV margin compared with the intra-fractional and setup uncertainties. Matching on soft tissue in the CBCT or on FM in kV/kV image pairs increased the margin by less than 0.7 mm. Matching on the ERB required an additional margin of up to 5.4 mm.
Table 4

Margins needed to be applied for patients with low risk prostate cancer when matching on fiducial markers (FM), soft tissue (ST), bones (B) or balloon (ERB)

M (mm) CBCT ST (mm) CBCT B (mm) CBCT Ba (mm) CBCT M (mm) kV B (mm) kV Ba (mm) kV
Anterior7.187.699.459.077.678.648.64
Posterior5.476.098.207.756.107.277.27
Long7.097.169.008.327.7912.4812.48
Lat5.475.885.956.105.657.587.58
Table 5

Margins needed to be applied for patients with high risk prostate cancer when matching on fiducial markers (FM), soft tissue (ST), bones (B) or balloon (ERB)

M (mm) CBCT ST (mm) CBCT B (mm) CBCT Ba (mm) CBCT M (mm) kV B (mm) kV Ba (mm) kV
Anterior10.5210.9412.2811.9910.8812.1811.62
Posterior6.396.948.858.446.948.687.99
Long9.679.7211.2310.7110.1911.2414.23
Lat7.397.767.807.957.537.889.11

For low risk prostate cancer patients the prostate only was treated. For high risk prostate cancer patients the prostate and the seminal vesicles were treated.

Margins needed to be applied for patients with low risk prostate cancer when matching on fiducial markers (FM), soft tissue (ST), bones (B) or balloon (ERB) Margins needed to be applied for patients with high risk prostate cancer when matching on fiducial markers (FM), soft tissue (ST), bones (B) or balloon (ERB) For low risk prostate cancer patients the prostate only was treated. For high risk prostate cancer patients the prostate and the seminal vesicles were treated.

Discussion

In the IMRT-IGRT-era, PTV margin determination has become of crucial importance [3,10]. We found a PTV margin between 5-8 mm for the P (LR) and between 6-11 mm for the PSV (IR/HR). A similar prostate PTV margin of 7 mm was determined by van Herk and colleagues though intra-fractional motion was not considered in their calculation and no ERB was used [11]. PTV margins used at the institution based on the CHHIP protocol and were 1 cm in all directions for PTV1 (PSV), 1 cm/5 mm (posterior) for PTV2 (P) and 5 mm/0 mm (posterior) for PTV3 (P) managing the risk of rectal toxicity [8]. However use of IGRT was not required to compensate for organ motion. We found that random errors produced by intra-fraction (organ motion) and inter-fraction (setup) uncertainties were of similar magnitude. Interestingly, the largest uncertainty was introduced by CTV delineation (LR PC: 1 - 2 mm, IR/HR PC: 1.6 - 3.5 mm). Patient positioning like bone or ERB matching resulted in considerably larger PTV margins [12]. Assuming an optimal delineation error of 1 mm our recommendations for an anisotropic PTV margin are as follows: PTVP 6.2 mm (ant), 5.43 mm (post), 6.06 mm (lng) and 4.79 mm (lat), PTVPSV 8.03 mm (ant), 5.81 mm (post), 7.47 mm (lng) and 6.87 mm (lat). Application of organ tracking would allow for further PTV lowering. Image-guided patient positioning considerably reduced PTV margin by about 2- 4.7 mm compared with bone or ERB matching which is known from the literature [10]. IGRT modalities such as CBCT and kV/kV-image pairs with FM or CBCT soft tissue matching were interchangeable in respect of accuracy. While some studies have shown the same, others have found that identification of FM on volumetric or planar images was not equivalent (±3 mm) [5]. However, one needs to keep in mind intra-fractional prostate motion between imaging modalities. In this study, the intra-fractional prostate movement was generally below 2 mm (Σ was 0.36 – 0.75 mm) for treatment duration of around 90 seconds, similar to results from the literature (Σ 1-2 mm) [13]. Prolonged RT duration can increase the intra-fractional prostate movement up to 3-6 mm [14]. Using an ERB might decrease intra-fractional prostate motion. Importantly, the SV move independently from the prostate gland with an uncertainty for PTV between 1.4 – 4.5 mm and should be considered in future margin protocols [15]. Remarkably, the largest uncertainties for PTV margins were produced by the radiation oncologist’s CTV delineation, particularly in the longitudinal axis. In the literature, the inter-observer variation for contouring has been reported to be largest in regions near the seminal vesicles and the apex [16]. Several geometrical uncertainties are involved in the delineation process including imaging resolution [6]. CTV delineation variation can be significantly reduced using MRI, teaching/protocols or automated model-based organ segmentation [17]. In practice, for protocols such as RTOG 0938 trial, a randomized phase II trial on hypofractionated radiation therapy, PTV was defined as the CTVp plus 5 mm in all dimensions except 3 mm posteriorly or anteriorly if necessary [18]. However, reducing treatment margins based on improved treatment accuracy may lead to geographical miss and serious underdosage of the CTV [7]. In order to prevent a geographical miss adaptive IGRT (-2.5-4.5 mm) might be recommended with daily imaging, especially in hypofractionation protocols [19]. The preference of IGRT modality is subject to various considerations. In our study we have chosen FM match in CBCT as reference since the stability of implanted markers was shown by several studies with an average seed migration of 1.2 +/- 0.2 mm (seeds) and 0.8 mm +/- 0.6 mm (coils) [20]. Advantages of kV/kV image pairs compared to CBCT are reduced acquisition/matching time, imaging dose and lower cost. However, it is an invasive procedure with the risk for prostate infection. In our opinion, FM and CBCT complement each other, especially in view of adaptive and real-time tumor tracking [21].

Consent

Written informed consent was obtained from the patient for data evaluation and publication; in case of participation in the CHIPP trial an additional informed consent was obtained.
  19 in total

1.  The probability of correct target dosage: dose-population histograms for deriving treatment margins in radiotherapy.

Authors:  M van Herk; P Remeijer; C Rasch; J V Lebesque
Journal:  Int J Radiat Oncol Biol Phys       Date:  2000-07-01       Impact factor: 7.038

Review 2.  Errors and margins in radiotherapy.

Authors:  Marcel van Herk
Journal:  Semin Radiat Oncol       Date:  2004-01       Impact factor: 5.934

3.  First clinical release of an online, adaptive, aperture-based image-guided radiotherapy strategy in intensity-modulated radiotherapy to correct for inter- and intrafractional rotations of the prostate.

Authors:  Heinz Deutschmann; Gerhard Kametriser; Philipp Steininger; Philipp Scherer; Helmut Schöller; Christoph Gaisberger; Michaela Mooslechner; Bernhard Mitterlechner; Harald Weichenberger; Gert Fastner; Karl Wurstbauer; Stephan Jeschke; Rosemarie Forstner; Felix Sedlmayer
Journal:  Int J Radiat Oncol Biol Phys       Date:  2011-12-30       Impact factor: 7.038

4.  Adaptive off-line protocol for prostate external radiotherapy with cone beam computer tomography.

Authors:  M Piziorska; P Kukołowicz; A Zawadzka; M Pilichowska; P Pęczkowski
Journal:  Strahlenther Onkol       Date:  2012-10-10       Impact factor: 3.621

5.  A general methodology for three-dimensional analysis of variation in target volume delineation.

Authors:  P Remeijer; C Rasch; J V Lebesque; M van Herk
Journal:  Med Phys       Date:  1999-06       Impact factor: 4.071

Review 6.  Target definition in prostate, head, and neck.

Authors:  Coen Rasch; Roel Steenbakkers; Marcel van Herk
Journal:  Semin Radiat Oncol       Date:  2005-07       Impact factor: 5.934

7.  Impact of inter- and intrafraction deviations and residual set-up errors on PTV margins. Different alignment techniques in 3D conformal prostate cancer radiotherapy.

Authors:  T Langsenlehner; C Döller; P Winkler; G Gallé; K S Kapp
Journal:  Strahlenther Onkol       Date:  2013-02-28       Impact factor: 3.621

8.  A strategy for the use of image-guided radiotherapy (IGRT) on linear accelerators and its impact on treatment margins for prostate cancer patients.

Authors:  Olaf Nairz; Florian Merz; Heinz Deutschmann; Peter Kopp; Helmut Schöller; Franz Zehentmayr; Karl Wurstbauer; Gerhard Kametriser; Felix Sedlmayer
Journal:  Strahlenther Onkol       Date:  2008-12-24       Impact factor: 3.621

9.  Improved clinical outcomes with high-dose image guided radiotherapy compared with non-IGRT for the treatment of clinically localized prostate cancer.

Authors:  Michael J Zelefsky; Marisa Kollmeier; Brett Cox; Anthony Fidaleo; Dahlia Sperling; Xin Pei; Brett Carver; Jonathan Coleman; Michael Lovelock; Margie Hunt
Journal:  Int J Radiat Oncol Biol Phys       Date:  2012-02-11       Impact factor: 7.038

10.  Escalated-dose versus standard-dose conformal radiotherapy in prostate cancer: first results from the MRC RT01 randomised controlled trial.

Authors:  David P Dearnaley; Matthew R Sydes; John D Graham; Edwin G Aird; David Bottomley; Richard A Cowan; Robert A Huddart; Chakiath C Jose; John Hl Matthews; Jeremy Millar; A Rollo Moore; Rachel C Morgan; J Martin Russell; Christopher D Scrase; Richard J Stephens; Isabel Syndikus; Mahesh K B Parmar
Journal:  Lancet Oncol       Date:  2007-06       Impact factor: 41.316

View more
  20 in total

1.  Evaluating deviations in prostatectomy patients treated with IMRT.

Authors:  Ana Cravo Sá; Ana Peres; Mónica Pereira; Carina Marques Coelho; Fátima Monsanto; Ana Macedo; Adrian Lamas
Journal:  Rep Pract Oncol Radiother       Date:  2015-12-29

2.  Shorter treatment times reduce the impact of intra-fractional motion : A real-time 4DUS study comparing VMAT vs. step-and-shoot IMRT for prostate cancer.

Authors:  Hendrik Ballhausen; Minglun Li; Ute Ganswindt; Claus Belka
Journal:  Strahlenther Onkol       Date:  2018-03-09       Impact factor: 3.621

3.  Penile bulb sparing in prostate cancer radiotherapy : Dose analysis of an in-house MRI system to improve contouring.

Authors:  F Böckelmann; M Hammon; S Lettmaier; R Fietkau; C Bert; F Putz
Journal:  Strahlenther Onkol       Date:  2018-10-12       Impact factor: 3.621

4.  Reducing ExacTrac intrafraction imaging uncertainty for prostate stereotactic body radiotherapy using a pre-treatment CBCT.

Authors:  Barry Jordan; Luis Muñoz; Christopher Colyer
Journal:  Phys Eng Sci Med       Date:  2022-04-19

5.  Technical note: improved positioning protocol for patient setup accuracy in conventional radiotherapy for lung cancer.

Authors:  Hongbo Chai; Yuichiro Narita; Masafumi Takagi; Mikiko Kudo; Tomomi Kimura; Keiichi Kattou
Journal:  Radiol Phys Technol       Date:  2019-09-23

6.  Long-term results of a study using individualized planning target volumes for hypofractionated intensity-modulated radiotherapy boost for prostate cancer.

Authors:  William Chu; D Andrew Loblaw; Kelvin Chan; Gerard Morton; Richard Choo; Ewa Szumacher; Cyril Danjoux; Jean-Philippe Pignol; Patrick Cheung
Journal:  Radiat Oncol       Date:  2015-04-18       Impact factor: 3.481

7.  Influence of daily imaging on plan quality and normal tissue toxicity for prostate cancer radiotherapy.

Authors:  Katharina Bell; Marina Heitfeld; Norbert Licht; Christian Rübe; Yvonne Dzierma
Journal:  Radiat Oncol       Date:  2017-01-10       Impact factor: 3.481

8.  A Dosimetric Comparison between Conventional Fractionated and Hypofractionated Image-guided Radiation Therapies for Localized Prostate Cancer.

Authors:  Ming Li; Gao-Feng Li; Xiu-Yu Hou; Hong Gao; Yong-Gang Xu; Ting Zhao
Journal:  Chin Med J (Engl)       Date:  2016-06-20       Impact factor: 2.628

9.  Editorial: Image-Guided Radiotherapy for Effective Radiotherapy Delivery.

Authors:  Nam P Nguyen; Ulf Lennart Karlsson
Journal:  Front Oncol       Date:  2015-11-18       Impact factor: 6.244

10.  A novel CBCT-based method for derivation of CTV-PTV margins for prostate and pelvic lymph nodes treated with stereotactic ablative radiotherapy.

Authors:  Ciara A Lyons; Raymond B King; Sarah O S Osman; Stephen J McMahon; Joe M O'Sullivan; Alan R Hounsell; Suneil Jain; Conor K McGarry
Journal:  Radiat Oncol       Date:  2017-08-04       Impact factor: 3.481

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.