Prenatal Foetal Non-invasive ECG instead of Doppler CTG - A Better Alternative?

N Sänger1, B R Hayes-Gill2, S Schiermeier3, W Hatzmann3, J Yuan1, E Herrmann4, F Louwen1, J Reinhard5.   

Abstract

Introduction: This study aimed to evaluate foetal signal quality obtained using an antenatal foetal ECG system (Monica 24™) and compare it with Doppler ultrasound CTG monitoring (Corometrics® 250 series). Material and
Methods: Seventy pregnant women (gestational age: between 20 + 0 weeks and 40 + 0 weeks) were examined using the Monica AN24™ system and also underwent Doppler CTG. The signal quality of both methods was compared and correlated with gestational age and pre-pregnancy body mass index (BMI).
Results: Overall, ECG had a signal quality of 77.4 % and CTG had a signal quality of 73.1 % (p > 0.05). In gestational weeks (GW) 20-26, the signal quality of ECG was significantly better compared to that obtained with CTG (75.5 vs. 45.3 %; p = 0.003), while in GW 27-36, the signal quality was better with CTG (72.3 vs. 83.0 %, p = 0.001). No difference in signal quality was found between the two methods after the 37th GW (87.7 vs. 86.1 %; p > 0.05). CTG showed a statistically significant correlation with BMI (rho 0.25, p < 0.05) while ECG showed no such correlation.
Conclusion: The use of non-invasive ECG is particularly indicated in the early weeks of pregnancy, while CTG offers superior results during the vernix period. There was no difference in signal quality after the vernix period. The signal quality with ECG was found to be independent of BMI, while the signal quality of CTG deteriorated with increasing BMI.

Entities:  

Keywords:  cardiotocogram (CTG); foetal electrocardiogram (ECG); signal quality antenatally

Year:  2012        PMID: 25278624      PMCID: PMC4168322          DOI: 10.1055/s-0032-1315012

Source DB:  PubMed          Journal:  Geburtshilfe Frauenheilkd        ISSN: 0016-5751            Impact factor:   2.915


  12 in total

Review 1.  Fetal heart rate monitoring: is it salvageable?

Authors:  J T Parer; T King
Journal:  Am J Obstet Gynecol       Date:  2000-04       Impact factor: 8.661

2.  Cardiotocography only versus cardiotocography plus PR-interval analysis in intrapartum surveillance: a randomised, multicentre trial. FECG Study Group.

Authors:  B K Strachan; W J van Wijngaarden; D Sahota; A Chang; D K James
Journal:  Lancet       Date:  2000-02-05       Impact factor: 79.321

3.  Uterine activity monitoring during labour--a multi-centre, blinded two-way trial of external tocodynamometry against electrohysterography.

Authors:  J Reinhard; B R Hayes-Gill; S Schiermeier; H Löser; L M Niedballa; E Haarmann; A Sonnwald; W Hatzmann; T M Heinrich; F Louwen
Journal:  Z Geburtshilfe Neonatol       Date:  2011-10-25       Impact factor: 0.685

4.  Obesity in pregnancy: implications for the mother and lifelong health of the child. A consensus statement.

Authors:  Lucilla Poston; Lucien F Harthoorn; Eline M Van Der Beek
Journal:  Pediatr Res       Date:  2011-02       Impact factor: 3.756

5.  [The value of Doppler cardiotocogram computer analysis system 70 minutes before delivery].

Authors:  S Schiermeier; H Hatzmann; J Reinhard
Journal:  Z Geburtshilfe Neonatol       Date:  2008-10-27       Impact factor: 0.685

6.  Signal ambiguity resulting in unexpected outcome with external fetal heart rate monitoring.

Authors:  Duncan R Neilson; Roger K Freeman; Shelora Mangan
Journal:  Am J Obstet Gynecol       Date:  2008-04-02       Impact factor: 8.661

7.  Comparison of non-invasive fetal electrocardiogram to Doppler cardiotocogram during the 1st stage of labor.

Authors:  Joscha Reinhard; Barrie R Hayes-Gill; Qian Yi; Hendrike Hatzmann; Sven Schiermeier
Journal:  J Perinat Med       Date:  2010-03       Impact factor: 1.901

8.  Computerised analysis of the fetal heart rate and relation to acidaemia at delivery.

Authors:  B K Strachan; D S Sahota; W J van Wijngaarden; D K James; A M Chang
Journal:  BJOG       Date:  2001-08       Impact factor: 6.531

Review 9.  Accuracy of frequency-related parameters of the electrohysterogram for predicting preterm delivery: a review of the literature.

Authors:  Maartje P G C Vinken; Chiara Rabotti; Massimo Mischi; S Guid Oei
Journal:  Obstet Gynecol Surv       Date:  2009-08       Impact factor: 2.347

10.  Sensitivity and specificity of intrapartum computerised FIGO criteria for cardiotocography and fetal scalp pH during labour: multicentre, observational study.

Authors:  S Schiermeier; S Pildner von Steinburg; A Thieme; J Reinhard; M Daumer; M Scholz; W Hatzmann; K T M Schneider
Journal:  BJOG       Date:  2008-08-25       Impact factor: 6.531

View more
  3 in total

1.  A systematic scoping review to identify the design and assess the performance of devices for antenatal continuous fetal monitoring.

Authors:  Kajal K Tamber; Dexter J L Hayes; Stephen J Carey; Jayawan H B Wijekoon; Alexander E P Heazell
Journal:  PLoS One       Date:  2020-12-01       Impact factor: 3.240

2.  Detection of fetal arrhythmia by adaptive single channel electrocardiogram extraction.

Authors:  M Suganthy; S Immaculate Joy; P Anandan
Journal:  Phys Eng Sci Med       Date:  2021-06-25

Review 3.  Non-Adaptive Methods for Fetal ECG Signal Processing: A Review and Appraisal.

Authors:  Rene Jaros; Radek Martinek; Radana Kahankova
Journal:  Sensors (Basel)       Date:  2018-10-27       Impact factor: 3.576

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.