| Literature DB >> 25207415 |
Yuichi Akino1, John P Gibbons, Daniel W Neck, Connel Chu, Indra J Das.
Abstract
Accurate beam data acquisition during commissioning is essential for modeling the treatment planning system and dose calculation in radiotherapy. Although currently several commercial scanning systems are available, there is no report that compared the differences among the systems because most institutions do not acquire several scanning systems due to the high cost, storage space, and infrequent usage. In this report, we demonstrate the intra- and intervariability of beam profiles measured with four commercial scanning systems. During a recent educational and training workshop, four different vendors of beam scanning water phantoms were invited to demonstrate the operation and data collection of their systems. Systems were set up utilizing vendor-recommended protocols and were operated with a senior physicist, who was assigned as an instructor along with vendor. During the training sessions, each group was asked to measure beam parameters, and the intravariability in percent depth dose (PDD). At the end of the day, the profile of one linear accelerator was measured with each system to evaluate intervariability. Relatively very small (SD < 0.12%) intervariability in PDD was observed among four systems at a region deeper than peak (1.5 cm). All systems showed almost identical profiles. At the area within 80% of radiation field, the average, and maximum differences were within ± 0.35% and 0.80%, respectively, compared to arbitrarily chosen IBA system as reference. In the penumbrae region, the distance to agreement (DTA) of the region where dose difference exceed ± 1% was less than 1 mm. Repeated PDD measurement showed small intravariability with SD < 0.5%, although large SD was observed in the buildup region. All four water phantom scanning systems demonstrated adequate accuracy for beam data collection (i.e., within 1% of dose difference or 1 mm of DTA among each other). It is concluded that every system is capable of acquiring accurate beam. Thus the selection of a water scanning system should be based on institutional comfort, personal preference of software and hardware, and financial consideration.Entities:
Mesh:
Substances:
Year: 2014 PMID: 25207415 PMCID: PMC5875503 DOI: 10.1120/jacmp.v15i4.4850
Source DB: PubMed Journal: J Appl Clin Med Phys ISSN: 1526-9914 Impact factor: 2.102
Details of the water tanks and chambers used for measurements
|
|
|
|
| |
|---|---|---|---|---|
| Water Tank | ||||
| Name | Blue Phantom | MP3 | DoseView 3D | 3D Scanner |
| Shape | cubic | cubic | cubic | cylindrical |
| Scan size (cm) |
|
|
|
|
| Tank setup | manual | manual | manual | automatic |
|
| shift/software | shift/software | shift/software | shift/software |
| Detector for soft‐wedge | LDA‐99 | LA‐48 | NA | water proof profiler |
| Chamber Used | ||||
| Type | CC13 | TN 31010 | Exradin A18 | CC13 |
| Sensitive volume |
|
|
|
|
; ; ; .
Figure 1Percent depth dose (PDD) (a) of one treatment unit measured with various scanning systems. The difference (b) of the PDD values compared to the data measured with IBA scanning system chosen arbitrary as a reference.
Figure 2Off‐center ratio (OCR) of the same treatment unit measured with various scanning systems at (a) 1.5 cm, (b) 5.0 cm, (c) 10.0 cm, and (d) 20.0 cm depth.
Percent difference of the off‐axis ratio between IBA and other systems inside the 80% of radiation field. (range) inside the 80% of radiation field are shown
|
|
|
|
|
|---|---|---|---|
| 15 |
|
|
|
| 50 |
|
|
|
| 100 |
|
|
|
| 200 |
|
|
|
Percent difference of the off‐axis ratio between IBA and other systems inside the 120%–130% area of half radiation field. (range) inside the 120%–130% area of half radiation field in both positive and negative directions
|
|
|
|
|
|---|---|---|---|
| 15 |
|
|
|
| 50 |
|
|
|
| 100 |
|
|
|
| 200 |
|
|
|
Figure 3Dose difference (DD) and distance to agreement (DTA) of the off‐center ratio (OCR). The IBA scanning system was chosen arbitrarily as a reference system for comparing data. DTA data are illustrated at the region between 80%–120% of the half radiation field from central axis.
Figure 4Reproducibility of percent depth dose (PDD) as measure of intrasystem variability measured with various scanning systems. Solid and dash lines represent standard deviation (SD) of three or four measurements and representative PDD profiles, respectively.
Figure 5The ratio of the percent depth dose measured with positive and negative bias for electrometer is illustrated as the polarity effects.